Facet Screw Systems - Performance Criteria for Safety and Performance Based Pathway: Guidance for Industry and Food and Drug Administration Staff

us-fda-guidance-4da49eb9baaa8e255f2f71c6c97cc338Facet Screw Systems - Performance Criteria for Safety and Performance Based Pathway: Guidance for Industry and Food and Drug Administration StaffGuidance

Administrative guidance, not enacted legislation. This captured publisher edition preserves the listed document and attachments; later developments are not independently consolidated.

Docket Number:
FDA-2021-D-0604
Issued by:
Guidance Issuing Office
Center for Devices and Radiological Health

This guidance document provides the FDA's recommendations on performance criteria to support premarket submissions for facet screw systems in the Safety and Performance Based Pathway. Under this framework, submitters planning to submit a 510(k) using the Safety and Performance Based Pathway for facet screw systems will have the option to use the performance criteria proposed in this final guidance to support substantial equivalence, rather than a direct comparison of the performance of the subject device to that of a predicate device. The recommendations are intended to promote consistency and facilitate efficient review of facet screw systems submissions.


Submit Comments

You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))

If unable to submit comments online, please mail written comments to:

Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852

All written comments should be identified with this document's docket number: FDA-2021-D-0604.

Contains Nonbinding Recommendations
Facet Screw Systems – Performance
Criteria for Safety and Performance
Based Pathway
Guidance for Industry and
Food and Drug Administration Staff
Document issued on April 13, 2022.
The draft of this document was issued on August 30, 2021.
For questions about this document, contact OHT6: Office of Orthopedic Devices/DHT6B:
Division of Spinal Devices at 301-796-5650.
U.S. Department of Health and Human Services
Food and Drug Administration
Center for Devices and Radiological Health
Contains Nonbinding Recommendations
Preface
Public Comment
You may submit electronic comments and suggestions at any time for Agency consideration to
https://www.regulations.gov. Submit written comments to the Dockets Management Staff, Food
and Drug Administration, 5630 Fishers Lane, Room 1061, (HFA-305), Rockville, MD 20852.
Identify all comments with the docket number FDA-2021-D-0604. Comments may not be acted
upon by the Agency until the document is next revised or updated.
Additional Copies
Additional copies are available from the Internet. You may also send an email request to CDRH-
Guidance@fda.hhs.gov to receive a copy of the guidance. Please include the document number
21001 and complete title of the guidance in the request.
Contains Nonbinding Recommendations
Facet Screw Systems – Performance
Criteria for Safety and Performance
Based Pathway
Guidance for Industry and
Food and Drug Administration Staff
This guidance represents the current thinking of the Food and Drug Administration (FDA or
Agency) on this topic. It does not establish any rights for any person and is not binding on
FDA or the public. You can use an alternative approach if it satisfies the requirements of the
applicable statutes and regulations. To discuss an alternative approach, contact the FDA staff
or Office responsible for this guidance as listed on the title page.
I. 	Introduction
This guidance provides performance criteria for facet screw systems in support of the Safety and
Performance Based Pathway.1 Under this framework, submitters (you) planning to submit a
510(k) using the Safety and Performance Based Pathway for facet screw systems will have the
option to use the performance criteria proposed in this guidance to support substantial
equivalence, rather than a direct comparison of the performance of the subject device to that of a
predicate device.
For the current edition of the FDA-recognized consensus standard(s) referenced in this
document, see the FDA Recognized Consensus Standards Database.2 For more information
regarding use of consensus standards in regulatory submissions, please refer to the FDA
guidance titled Appropriate Use of Voluntary Consensus Standards in Premarket Submissions
for Medical Devices.3
The contents of this document do not have the force and effect of law and are not meant to bind
the public in any way, unless specifically incorporated into a contract. This document is intended
only to provide clarity to the public regarding existing requirements under the law. FDA
guidance documents, including this guidance, should be viewed only as recommendations, unless
1 Available at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/safety-and-performance-
based-pathway
2 Available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/search.cfm
3 Available at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/appropriate-use-
voluntary-consensus-standards-premarket-submissions-medical-devices
1
Contains Nonbinding Recommendations
specific regulatory or statutory requirements are cited. The use of the word should in Agency
guidance means that something is suggested or recommended, but not required.
II. 	Scope/Device Description
The facet screw systems that are the subject of this guidance consist of metallic bone screws and
optional washer components. These devices are unclassified and are identified with the product
code MRW (system, facet screw spinal device).
Intended Use/Indications for Use:
The facet screw systems that fall within the scope of this guidance document are intended for
bilateral immobilization of facet joints to stabilize the spine as an aid to fusion. The optional
washer components are intended for use with the facet screw to aid in load distribution at the
screw head/bone interface.
Device Design Characteristics:
The facet screw systems that fall within the scope of this guidance document consist of solid or
cannulated screws with fully or partially threaded screw shafts, and optional washer components,
constructed solely from the following material in conformance with the associated FDA-
recognized consensus standard:
• 	ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI
(Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401)
A dimensional comparison of the subject device should be performed, and the dimensions should
fall within the dimensional ranges listed in Table 1. Washer components should have an inner
diameter that is larger than the thread diameter (major diameter) of the compatible screw and less
than the diameter of the screw head.
Note: Based on review of historical submissions, screws below 4.5 mm in diameter and screws
4.5 mm and above were often indicated for different anatomical regions and have different
design characteristics and different performance characteristics. Therefore, screw design
characteristics and performance criteria are stratified in this document based on these diameter
ranges.
Table 1 – Dimensional ranges for facet screws*
Facet Screws Parameters 	< 4.5 mm Diameter 	≥ 4.5 mm Diameter
Nominal Major Diameter
Range
3.5-4.3 mm 	4.5-6.0 mm
Minimum Total Screw
Length**
6.0 mm 	15 mm
Minimum Threaded Length 	6.0 mm 	12 mm
* The dimensional ranges listed were derived from historical data submitted to FDA in 510(k) submissions for
devices previously found substantially equivalent.
2
Contains Nonbinding Recommendations
** The maximum facet screw length reported in the historical data was 60 mm. However, maximum screw
length can be justified based on the anatomic region into which the subject facet screws are intended to be
implanted.
Facet screw systems that fall within the following categories are outside the scope of this
guidance for the Safety and Performance Based Pathway:
• 	Combination products
• 	Resorbable devices
• 	Device with coatings
• 	Additively manufactured devices
• 	Devices that utilize surgical techniques or associated instruments outside the standard of
care
• 	Devices with complex geometries, or unique technological characteristics (e.g., unique
screw thread, modularity, fenestrations)
• 	Devices sterilized using novel sterilization methods as described in FDA’s guidance
Submission and Review of Sterility Information in Premarket Notification (510(k))
Submissions for Devices Labeled as Sterile.4
FDA may determine, on a case-by-case basis, that additional data are necessary to evaluate
whether the device is appropriate for the Safety and Performance Based Pathway. In situations
where you determine that additional testing outside of those identified in this guidance are
necessary to make a determination regarding eligibility into the Safety and Performance Based
Pathway, we would encourage sponsors to submit a Pre-Submission5 to engage in discussion
with FDA prior to submission of the 510(k).
III. 	Testing Performance Criteria
If your device is appropriate for submission through the Safety and Performance Based Pathway,
and you choose to use that option, we do not expect you to provide direct comparison testing
against a legally marketed predicate to demonstrate substantially equivalent performance
characteristics. To ensure that the performance criteria outlined in this guidance remain
contemporary and take into account relevant data from recent clearances, FDA recommends that
you provide a results summary for all tests evaluated in addition to the other submission
information (e.g., Declaration of Conformity (DoC)) recommended below for each test or
evaluation. Consistent with FDA policy for all 510(k) submissions, for all 510(k) submissions
under the Safety and Performance Based Pathway, FDA may request and review underlying data
demonstrating that a new device meets the FDA-identified performance criteria and testing
methodology, as necessary. Unless otherwise identified in the submission information sections
below, test information such as results summary, test protocols, or complete test reports should
be submitted as part of the 510(k) as described in FDA’s guidance: Safety and Performance
4 Available at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/submission-and-review-
sterility-information-premarket-notification-510k-submissions-devices-labeled
5 Available at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/requests-feedback-and-
meetings-medical-device-submissions-q-submission-program
3
Contains Nonbinding Recommendations
Based Pathway.6 For additional information regarding the submission of non-clinical bench
testing information, please see FDA’s guidance Recommended Content and Format of Non-
Clinical Bench Performance Testing Information in Premarket Submissions.7
Mechanical Testing
To assess mechanical strength of the worst-case facet screw(s) in the system, static cantilever
bending testing should be performed on your final, finished device in conformance with the FDA
currently-recognized version of ASTM F2193 Standard Specifications and Test Methods for
Components Used in the Surgical Fixation of the Spinal Skeletal System-Annex 4: Test Method
for Measuring the static and fatigue bending strength of metallic spinal screws. To assess screw
fixation, axial pullout strength should be evaluated using the engineering analysis method
described below.8 Mechanical testing and engineering analyses should be performed on devices
that represent the worst-case (e.g., most likely to loosen or fail). You should also provide a
rationale identifying how you identified the worst-case design for each test/evaluation. All
mechanical testing should be performed on the final, finished versions of the devices unless
certain processes (e.g., sterilization) can be rationalized to have no impact on the mechanical
strength of the device. Acceptance criteria are listed below for each test.
For the mechanical test below, you should provide a report as specified in the relevant reporting
section of ASTM F2193, in addition to a Declaration of Conformity (DoC) to the consensus
standard. Any protocol deviations should be thoroughly described and justified; however, note
that certain protocol deviations may invalidate comparison to the performance criteria listed
below, resulting in the need for submission of a Traditional, Special, or Abbreviated 510(k), as
appropriate.
1. 	Test name: Static Cantilever Bending
Methodology: FDA-recognized version of ASTM F2193 Standard Specifications and
Test Methods for Components Used in the Surgical Fixation of the Spinal Skeletal
System-Annex 4: Test Method for Measuring the static and fatigue bending strength of
metallic spinal screws
Performance Criteria:
6 Available at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/safety-and-performance-
based-pathway
7 Available at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/recommended-content-
and-format-non-clinical-bench-performance-testing-information-premarket
8 It should be noted that although ASTM F2193 is FDA-recognized in full, FDA believes that for the purposes of the
Safety and Performance Based Pathway, the testing, methods and criteria identified in this section on mechanical
bench testing represent the least burdensome approach to demonstrating substantial equivalence for this pathway,
although alternative or additional methods or acceptance criteria are identified in the recognized consensus standard
for some tests.
4
Contains Nonbinding Recommendations
Table 2 – Static cantilever bending acceptance criteria for facet screw systems
Test Parameter 	< 4.5 mm diameter
(Cervical)
≥ 4.5 mm diameter
(Lower
Thoracic/Lumbar)
Static Cantilever
Bending Yield
Moment (Nm)
2.6 	3.7
Performance Criteria Source: Criteria are based on aggregated mechanical testing data
submitted to FDA in 510(k) submissions for facet screw systems previously found to be
substantially equivalent. It should be noted that the values in the table above were
rounded to be the most inclusive and accurate based on the final data.
Additional Considerations: As specified in ASTM F2193, a minimum of five samples
should be tested. In order to be considered a successful result, either: (1) all samples
should meet or exceed the acceptance criteria listed above, or (2) the average of all
samples should meet or exceed the criteria above and the standard deviation should be ≤
10% of the calculated average.
Submission Information: Results summary and DoC
2. 	Test name: Axial Pullout Strength
Methodology: An engineering analysis is recommended to assess axial pullout strength
using the equation described by Chapman et al., 1996.9 Note that for this analysis to be
appropriate, the instrumentation identified in the associated surgical technique manual
should allow for close to idealized thread engagement. If this assumption is not accurate
for your scenario, then the identified engineering analysis may not be appropriate for the
assessment of the proposed device as identified in this guidance.
For all facet screw sizes, extract the relevant dimensions below (i.e., screw major
diameter, screw minor diameter, screw pitch, and axial thread length). These dimensions
will be used to calculate theoretical pullout strengths for the worst-case screws in the
device system using the following equation:
𝐹𝐹𝐹𝐹 = 𝑆𝑆 ∗ 𝐴𝐴 = {𝑆𝑆 ∗ 𝐿𝐿 ∗ 𝜋𝜋 ∗ 𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷 ∗ 𝑇𝑇𝑆𝑆𝐹𝐹}
Fs = predicted shear failure force (N)
S = material ultimate shear stress (MPa)
A = thread shear area (mm2)
9 Chapman, J. R. (1996). Factors Affecting the Pullout Strength of Cancellous Bone Screws. Journal of
Biomechanical Engineering,118(3), 391. doi:10.1115/1.2796022
5
Contains Nonbinding Recommendations
L = axial thread length (mm) including only threads that have the nominal major diameter
where complete purchase is expected (e.g., excluding the screw tip) of thread engagement
in material
Dmajor = major diameter (mm)
TSF = Thread Shape Factor (dimensionless) = (0.5 + 0.57735 d/p)
d = thread depth (mm) = (Dmajor – Dminor)/2
Dminor = minor (root) diameter (mm)
p = thread pitch (mm)
Use a material ultimate shear stress (S) value of 3.395 MPa, which is representative of
Grade 20 polyurethane foam material (per FDA currently-recognized version of ASTM
F1839 Standard Specification for Rigid Polyurethane Foam for Use as a Standard
Material for Testing Orthopaedic Devices and Instruments). The resulting theoretical
pullout strength value obtained for the device should be equivalent or greater to the
following values depending on the nominal major diameter of the worst-case screws. A
justification should be provided to support why the evaluated facet screws selected are
worst-case. Axial pullout performance is heavily influenced by amount of interface.
Factors such as decreasing outer diameter and decreasing axial thread length may help
identify the worst-case.
Dimensions used for calculations should be clearly listed for each theoretical outcome.
Dimensional values used in this calculation should be consistent with the values listed on
the screw engineering drawings.
Performance Criteria: The resulting theoretical pullout strength values obtained for
your worst-case devices should meet or exceed to the values listed in Table 3 depending
on the major diameter of the screw being evaluated.
Table 3 – Axial pullout strength acceptance criteria for facet screw systems
Nominal Major Diameter
(mm)
Theoretical Pullout Strength
in Grade 20 Foam (N)
< 4.5 mm 	190
≥ 4.5 mm 	390
Performance Criteria Source: Criteria are based on aggregated mechanical testing data
and device description information submitted to FDA in 510(k) submissions for facet
screws previously found to be substantial equivalent.
Submission Information: Results summary and engineering analysis
Sterilization (devices labeled as sterile) and Reprocessing (end-user sterilized) Validation
3. 	Test name: Sterilization (devices labeled as sterile) and Reprocessing (end-user
6
Contains Nonbinding Recommendations
sterilized)
Methodology: FDA currently-recognized versions of the following consensus standards
(as applicable):
• 	International Organization for Standardization (ISO) 17665-1 Sterilization of
health care products – Moist heat – Part 1: Requirements for the development,
validation, and routine control of a sterilization process for medical devices
• 	ISO 11135-1 Sterilization of health care products – Ethylene oxide- Part 1:
Requirements for development, validation, and routine control of a sterilization
process for medical devices
• 	ISO 11137-1 Sterilization of health care products—Radiation—Part 1:
Requirements for development, validation, and routine control of a sterilization
process for medical devices
• 	ISO 11607-1 Packaging for terminally sterilized medical devices – Part 1:
Requirements for materials, sterile barrier systems and packaging systems
• 	ISO 11607-2 Packaging for terminally sterilized medical devices – Part 2:
Validation requirements for forming, sealing and assembly processes
Performance Criteria: Validation testing should demonstrate the cleanliness and
sterility of, or the ability to clean and sterilize to a sterility assurance level of 10-6, the
device and device-specific instruments.
Performance Criteria Source: FDA guidance:
• 	Submission and Review of Sterility Information in Premarket Notification
(510(k)) Submissions for Devices Labeled as Sterile10
• 	Reprocessing Medical Devices in Health Care Settings: Validation Methods and
Labeling11
Submission Information: You should provide a description of the packaging (sterile
barrier system) and how it will maintain the device’s sterility, and a description of the
package test methods, but not package test data. With respect to the Established
Sterilization Method, whether using an Established Category A or Established Category
B sterilization method, you should provide the information in Section V.A. of the FDA
guidance Submission and Review of Sterility Information in Premarket Notification
(510(k)) Submissions for Devices Labeled as Sterile; generally, the validation data itself
is not needed to demonstrate substantial equivalence.
Biocompatibility Evaluation:
To identify the biocompatibility endpoints to include as part of your biocompatibility evaluation
you should use Attachment A of the FDA guidance Use of International Standard ISO 10993-1,
“Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk
management process,”12 referred to in the rest of this document as the “FDA Biocompatibility
10 Available at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/submission-and-review-
sterility-information-premarket-notification-510k-submissions-devices-labeled
11 Available at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/reprocessing-medical-
devices-health-care-settings-validation-methods-and-labeling
12 Available at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/use-international-
standard-iso-10993-1-biological-evaluation-medical-devices-part-1-evaluation-and
7
Contains Nonbinding Recommendations
Guidance” for brevity. FDA considers the devices covered by this guidance to be categorized as
Implanted Devices in contact with tissue/bone with a “permanent” contact duration of > 30 days
and you should assess the endpoints below per Attachment A of the FDA Biocompatibility
Guidance.
• 	Cytotoxicity
• 	Sensitization
• 	Irritation or Intracutaneous Reactivity
• 	Acute Systemic Toxicity
• 	Material-Mediated Pyrogenicity
• 	Sub-acute/Sub-chronic Toxicity
• 	Genotoxicity
• 	Implantation
• 	Chronic Toxicity
• 	Carcinogenicity
Rationale in Lieu of Testing: If the subject device is manufactured from the identical raw
materials using identical manufacturing processes as a predicate device with the same type and
duration of tissue contact, and any changes in geometry are not expected to impact the biological
response, this is typically sufficient to establish substantially equivalent biocompatibility if
documentation such as that outlined in Attachment F of the FDA Biocompatibility Guidance is
also provided.
Testing: In rare cases, if you determined that testing is needed to address some or all of the
identified biocompatibility endpoints, FDA recommends that complete test reports be provided
for all tests performed unless a declaration of conformity without supplemental information can
be appropriately provided, as discussed in Attachment E of the FDA Biocompatibility Guidance.
Any test-specific positive, negative, and/or reagent controls should perform as expected, and
protocol deviations should be thoroughly described and justified; however, note that certain
protocol deviations may invalidate comparison to the performance criteria listed below.,
resulting in the need for submission of a Traditional, Special, or Abbreviated 510(k).
4. 	Test name: Biocompatibility endpoints (identified from FDA Biocompatibility
Guidance)
Methodology: FDA currently-recognized versions of biocompatibility consensus
standards
Performance Criteria: All direct tissue contacting components of the device and device-
specific instruments should be determined to have an acceptable biological response.
Performance Criteria Source: The FDA Biocompatibility Guidance
Additional Considerations: For any biocompatibility test samples with an adverse
biological response, the biocompatibility evaluation should explain why the level of
toxicity seen is acceptable. Some comparison testing against a legally marketed predicate
may be necessary (and is considered acceptable under the Safety and Performance Based
Pathway) to support such a rationale as explained in the FDA Biocompatibility Guidance.
For standard biocompatibility test methods that include comparison device control
samples, the legally marketed comparison device control samples should perform as
expected.
8
Contains Nonbinding Recommendations
Submission Information: Refer to FDA Biocompatibility Guidance
9