CPG Sec. 300.400 Contamination of Devices Labeled as Sterile

us-fda-guidance-30034a4ccf7d54e5916691045dbc5367CPG Sec. 300.400 Contamination of Devices Labeled as SterileGuidance

Administrative guidance, not enacted legislation. This captured publisher edition preserves the listed document and attachments; later developments are not independently consolidated.

Docket Number:
FDA-2020-D-0957
Issued by:
Guidance Issuing Office
Office of Regulatory Affairs
Center for Devices and Radiological Health

This compliance policy guide describes FDA’s policy on contamination of devices labeled as sterile.


Submit Comments

You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))

If unable to submit comments online, please mail written comments to:

Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852

All written comments should be identified with this document's docket number: FDA-2020-D-0957.

CPG Sec. 300.400 Contamination of Devices Labeled as Sterile
BACKGROUND:
Certain devices labeled as sterile are frequently used on debilitated patients whose resistance to
microorganisms is lower than that of a normal healthy person.
Recent experience has demonstrated that devices labeled as sterile and intended for use under sterile
conditions may be nonsterile.
POLICY:
*A device labeled as sterile which fails to pass USP tests for sterility is adulterated under Section 50l(c) in
that its purity and quality falls below that which it is represented to possess and is misbranded under
502(a) in that the label statement "sterile" is false or misleading as applied to a product which is not
sterile, but is contaminated with viable organisms.*
Device nonsterility should be treated as top priority, because of the seriousness of the health hazard
posed by the microbial contamination of sterile products. The responsible firm should be notified of our
finding to afford it the opportunity to voluntarily recall its device.
Appropriate legal action and/or recall should be recommended when the examination of a sterile device
reveals defective packaging (separated seams, torn or punctured outer wrap) and there is no other
protection of package contents. Confirmation of nonsterility by sterility testing is unnecessary in these
cases. Inspection reports, exhibits and the results of analysis for defective packaging should be included
in the recommendation of the division within the Office of Medical Device and Radiological Health
Operations (OMDRHO) to CDRH when appropriate. The results of analysis should include the number of
units examined, the number of units found defective, the types of defects identified and the range in
size and numbers of holes or broken seams found per unit.
In situations where there is more than one wrap or there are questions regarding whether the nature of
the packaging defect meets the conditions outlined in this guideline, the appropriate division within
OMDRHO should consult with CDRH when appropriate.
*Material between asterisks is new or revised*
Issued: 8/23/74
Reissued: 6/11/76
Revised: 10/1/80, 9/24/87, 3/95