Revised Recommendations Regarding Invalidation of Test Results of Licensed and 510(k) Cleared Bloodborne Pathogen Assays Used to Test Donors: Guidance for Industry

us-fda-guidance-3acb9e05ff4c8f183b6391e4ef088b7dRevised Recommendations Regarding Invalidation of Test Results of Licensed and 510(k) Cleared Bloodborne Pathogen Assays Used to Test Donors: Guidance for IndustryGuidance

Administrative guidance, not enacted legislation. This captured publisher edition preserves the listed document and attachments; later developments are not independently consolidated.

Docket Number:
FDA-1999-D-0743
Issued by:
Guidance Issuing Office
Center for Biologics Evaluation and Research

To reduce infectious disease transmission by blood and blood products, donor samples from blood donations are tested for markers of pathogenic bloodborne infections, including antibodies, antigens, and nucleic acids that may indicate the presence of etiologic agents such as human immunodeficiency virus (HIV), hepatitis B virus (HBV), hepatitis C virus (HCV), human T-cell lymphotropic virus (HTLV), cytomegalovirus (CMV), syphilis, and others. The validity of screening and supplemental (confirmatory) test assay results is determined by the performance of test kit manufacturer supplied reagents labeled as Acontrols@, used in accordance with the test kit instructions.


Submit Comments

You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))

If unable to submit comments online, please mail written comments to:

Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852

All written comments should be identified with this document's docket number: FDA-1999-D-0743.

Questions?

Contact Point
Office of Communication, Outreach and Development (OCOD)
Center for Biologics Evaluation and Research
Food and Drug Administration
10903 New Hampshire Ave WO71-3128
Silver Spring, MD 20993-0002
ocod@fda.hhs.gov
(800) 835-4709
(240) 402-8010
Guidance for Industry
Revised Recommendations Regarding
Invalidation of Test Results of Licensed
and 510(k) Cleared Bloodborne
Pathogen Assays Used to Test Donors
Additional copies are available from:
Office of Communication, Training and Manufacturers Assistance (HFM-40)
1401 Rockville Pike, Rockville, MD 20852-1448
(Tel) 1-800-835-4709 or 301-827-1800
http://www.fda.gov/cber/guidelines.htm
U.S. Department of Health and Human Services
Food and Drug Administration
Center for Biologics Evaluation and Research (CBER)
July 2001
TABLE OF CONTENTS
I. 	INTRODUCTION ................................................................................................................. 1
II. 	BACKGROUND.................................................................................................................... 1
III. SPECIFIC RECOMMENDATIONS ................................................................................... 3
A. 	CLIA REQUIREMENTS FOR C ONTROL REAGENTS................................................................... 3
B. 	INVALIDATION OF TEST RESULTS BASED ON TEST K IT P ACKAGE INSERT................................. 3
C. 	INVALIDATION OF TEST RESULTS BASED ON CLIA REGULATIONS .......................................... 4
IV. 	REFERENCES ...................................................................................................................... 5
i
GUIDANCE FOR INDUSTRY
Revised Recommendations Regarding Invalidation of Test Results of
Licensed and 510(k) Cleared Bloodborne Pathogen Assays Used to
Test Donors
This guidance document represents FDA’s current thinking on invalidating test results based on
the Clinical Laboratory Improvement Act of 1988 (CLIA) required control reagents. It does not
create or confer any rights for or on any person and does not operate to bind FDA or the public.
An alternative approach may be used if such approach satisfies the requirements of the
applicable statutes and regulations.
I. 	INTRODUCTION
To reduce infectious disease transmission by blood and blood products, donor samples from blood
donations are tested for markers of pathogenic bloodborne infections, including antibodies, antigens,
and nucleic acids that may indicate the presence of etiologic agents such as human immunodeficiency
virus (HIV), hepatitis B virus (HBV), hepatitis C virus (HCV), human T-cell lymphotropic virus
(HTLV), cytomegalovirus (CMV), syphilis, and others. The validity of screening and supplemental
(confirmatory) test assay results is determined by the performance of test kit manufacturer supplied
reagents labeled as Acontrols@, used in accordance with the test kit instructions. Historically, FDA has
not required additional assay-specific quality control procedures beyond those documented in the test
kit package inserts. FDA now recommends implementation of quality assurance and quality control
procedures that involve use of external control reagents beyond those provided by the test kit
manufacturer in order to contribute to overall testing accuracy and therefore to blood safety. This
guidance finalizes the draft guidance entitled AGuidance for Industry-Revised Recommendations for the
Invalidation of Test Results When Using Licensed and 510(k) Cleared Bloodborne Pathogen Assays to
Test Donors@ that was announced in the Federal Register of September 1, 1999 (64 FR 47847). This
guidance also supercedes the January 3, 1994, guidance document on this topic (Ref. 1). These
invalidation procedures should be part of blood establishments= comprehensive quality assurance
programs (Ref. 2) and documented in blood establishments= Standard Operating Procedures (SOPs).
II. 	BACKGROUND
On May 28, 1992, September 25, 1992, September 26, 1996, and December 13, 1996, the Blood
Products Advisory Committee (BPAC) discussed in open public meetings circumstances under which
test results may be invalidated, including actions based on the use of additional quality control reagents
in donor testing. These discussions included general aspects of quality assurance related to bloodborne
pathogen testing, invalidation procedures based on test kit product insert requirements, and additional
1
procedures to invalidate test results based on criteria developed by the user. FDA issued guidance on
this subject on January 3, 1994 (Ref. 1). In an effort to further clarify FDA=s guidance and to help
ensure compliance with the Clinical Laboratory Improvement Act of 1988 (CLIA), at the December
13, 1996, BPAC meeting, the FDA, the Centers for Disease Control and Prevention (CDC), and the
Health Care Financing Administration (HCFA) discussed elements of CLIA. CDC described the
differences between a calibrating reagent and a reagent serving as an assay control. Based on CLIA
regulations, CDC stated that if a test kit supplied Acontrol@ reagent is used to calculate the assay cutoff,
regardless of its container label title, the reagent serves the function of a calibrator and may not serve
simultaneously as a control reagent. In such instances, an additional reagent should be included in the
assay run to meet CLIA control reagent requirements.
For example, if a test kit includes two reagents labeled as negative control and positive control
respectively, and the negative control reagent is used to calculate the assay cutoff, the negative control
serves as a calibrator and may not also serve as an assay control reagent. Since CLIA regulations
require use of a minimum of two controls (one negative and one positive) each day of testing, a negative
control from an external source should be included in the run to meet CLIA requirements. Such
external controls may be developed in-house or procured, and the blood establishment must develop
acceptance/rejection criteria for their use. For further information, blood establishments should refer to
CLIA regulations under Title 42, Code of Federal Regulations (CFR), Part 493, or contact the State
Survey Agency in the state where the laboratory is located or the accrediting organization with which the
laboratory is affiliated.
It is important to note that the test kit package insert instructions must continue to be followed in order
for runs to be considered valid for donor testing. 21CFR 606.65(e). Accordingly, externally supplied
reagents should not be substituted for test kit supplied reagents to calculate the assay cutoff, but may be
used additionally to satisfy CLIA requirements.
As a result of discussions held during BPAC meetings, and additional discussions among CDC, HCFA,
and FDA, FDA has developed revised recommendations for blood establishments, integrating current
CLIA provisions for invalidating test results based on CLIA designated control reagents.
Because blood safety relies significantly on the accuracy of testing, user-developed acceptance/rejection
rules should be designed only with the objective of enhancing blood safety. Accordingly, rules
developed by blood establishments to meet either CLIA requirements or other in-house developed
requirements should only be used to invalidate nonreactive results. Reactive results obtained from
a screening test or confirmatory/supplemental test run that satisfies the test kit== s criteria for
acceptance should not be invalidated. Such reactive results remain as the initial test of record, and,
in the case of a screening test, these specimens should be retested in duplicate as specified in the test kit
manufacturer=s instructions. The duplicate retest of a reactive specimen (duplicate rescreen) is designed
to reduce the probability of obtaining a false positive result for a truly negative, but non-specifically
reactive specimen. A donor sample with an invalidated nonreactive result may be retested singly.
2
III. 	SPECIFIC RECOMMENDATIONS
These recommendations for procedures to invalidate test results, based on: 1) test kit package insert
rejection criteria; and 2) CLIA provisions, are applicable to all blood establishments and should be
included in their SOPs.
A. 	CLIA Requirements for Control Reagents
CLIA regulations require control reagents to be used according to 42 CFR 493. If a
bloodborne pathogen test kit uses any of its manufacturer supplied reagents to serve as a
calibrator function, i.e., either or both of the test kit controls (negative or positive) are used to
calculate the assay cutoff, then CLIA regulations require that (an) additional Acontrol@ reagent(s)
be included in each run (see table below). Such reagents may be procured or developed in
house. In any case, prior to placing the additional controls in routine use, each lot of such
reagents should have: 1) a known dating period, i.e., established stability (supplied by a control
reagent manufacturer or established by the user on in-house developed control reagent); and 2)
known performance parameters, i.e., specifications for acceptance. Prior to implementation,
additional control reagents should be qualified, i.e., evaluated for suitability by the user applying
generally accepted laboratory quality control procedures to establish acceptance/rejection
criteria, to minimize possible incompatibilities that may exist with particular test kits before
putting control reagents into use.
TEST KIT
REAGENT(S)
USED IN CALCULATION OF
THE CUTOFF?
ADDITIONAL CONTROL
REAGENT REQUIRED
Negative Control 	Yes 	Yes (Negative Control)
Negative Control 	No 	No
Positive Control 	Yes 	Yes (Positive Control)
Positive Control 	No 	No
Positive and Negative
Control
Yes 	Yes (Positive and Negative Controls)
B. 	Invalidation of Test Results Based on Test Kit Package Insert
All results, both reactive and non-reactive, must be considered invalid when it has been
determined that the assay, or part of the assay, has not been performed according to the test kit
package insert instructions, (e.g., improper procedure, compromised reagent, faulty equipment)
or that the test kit package insert defined run acceptance criteria have not been met, (e.g.,
controls out of specified range) [21 CFR 606.65(e)]. Whenever such situations occur, test
results must be invalidated and the subsequent valid assay becomes the initial test of record.
The determination of nonadherence to the test kit package insert requirements may be a result
3
of an investigation into an unexplained discrepancy, a routine review of the assay data, or
concerns raised by a blood establishment's quality control procedures. For example, an
investigation of procedures may be performed in response to unexpected external control
reagent values or failure to meet the blood establishment's additional statistical acceptance
criteria for rate of reactivity. Documentation of each incident of invalidation should include the
basis for invalidation, the details of an investigation, including records of supervisory oversight,
the outcome of the investigation, and, if indicated, any corrective action taken. All of these
actions should be taken prior to repeat testing of donor samples.
C. 	Invalidation of Test Results Based on CLIA Regulations
When the test kit package insert instructions have been met, but CLIA control requirements
have not been met: 1) any reactive results must not be considered invalid and 2) non
reactive results should be invalidated. All specimens giving initially reactive results should
be re-tested in duplicate, or as the package insert directs [21 CFR 606.65(e)]. If either of the
repeated duplicates is reactive, the unit should be classified as repeatedly reactive and no further
repeat testing should be performed. Specimens giving repeatedly reactive test results then
should be further assayed and reported as recommended in FDA Recommendations to Blood
Establishments (Ref. 3, 4, 5, 6, 7, 8, 9) regarding viral marker testing of donations. Invalidated
non-reactive specimens should be re-tested singly and those results, if valid, should become the
test of record.
Invalidation criteria pursuant to CLIA requirements should be based on site-user developed
rules, e.g., population reactive prevalence data, or standard deviation based limits, which may
be derived from generally accepted clinical laboratory quality control algorithms.
Documentation for all incidents of invalidation should include the basis for invalidation, the
details of an investigation, including records of supervisory oversight, the outcome of the
investigation, and, if indicated, any corrective action taken. All of these actions should be taken
prior to repeat testing of donor samples.
Reactive results should only be invalidated when an assay run either fails to meet test kit
package insert acceptance criteria or the assay was not performed in accordance with the test
kit package insert. Accordingly, invalidation of reactive test results should not be based solely
on either CLIA required quality control reagent performance or site supplemental quality control
rules. Examples of circumstances when reactive results should not be invalidated include CLIA
required control reagent values exceeding site established quality control acceptance limits, or
reactive rates exceeding site defined upper limits.
4
IV. 	REFERENCES
1. 	FDA Recommendations for the Invalidation of Test Results When Using Licensed Viral Marker
Assays to Screen Donors, 3 January 1994, Center for Biologics Evaluation and Research,
FDA.
2. 	Guideline for Quality Assurance in Blood Establishments, 11 July 1995, Center for Biologics
Evaluation and Research, FDA.
3. 	Revised Recommendations for Testing Whole Blood, Blood Components, Source Plasma and
Source Leukocytes for Antibody to Hepatitis C Virus Encoded Antigen (Anti-HCV), 23 April
1992, Center for Biologics Evaluation and Research, FDA.
4. 	Revised Recommendations for Testing Whole Blood, Blood Components, Source Plasma and
Source Leukocytes for Antibody to Hepatitis C Virus Encoded Antigen (Anti-HCV), 5 August
1993, Center for Biologics Evaluation and Research, FDA.
5. 	Additional Recommendations for Testing Whole Blood, Blood Components, Source Plasma
and Source Leukocytes for Antibody to Hepatitis C Virus Encoded Antigen (Anti-HCV), 16
May 1996, Center for Biologics Evaluation and Research, FDA.
6. 	Recommendations for the Management of Donors and Units that are Initially Reactive for
Hepatitis B Surface Antigen (HBsAg), 2 December 1987, Center for Biologics Evaluation and
Research, FDA.
7. 	Revised Recommendations for the Prevention of Human Immunodeficiency Virus (HIV)
Transmission by Blood and Blood Products, 23 April 1992, Center for Biologics Evaluation
and Research, FDA.
8. 	FDA Recommendations Concerning Testing for Antibody to Hepatitis B Core Antigen (Anti-
HBc), 10 September 1992, Center for Biologics Evaluation and Research, FDA.
9. 	HTLV-I Antibody Testing, 29 September 1989, Center for Biologics Evaluation and Research,
FDA.
5