us-fda-guidance-7067225a3b6d3e6dd278df669aa9ffb8•ICH Q12: Implementation Considerations for FDA-Regulated Products: Draft Guidance for Industry
ICH Q12: Implementation Considerations for FDA-Regulated Products: Draft Guidance for Industry
us-fda-guidance-7067225a3b6d3e6dd278df669aa9ffb8ICH Q12: Implementation Considerations for FDA-Regulated Products: Draft Guidance for IndustryDraft Guidance
Administrative draft guidance, not enacted legislation. This captured publisher edition preserves the listed document and attachments; later developments are not independently consolidated.
- Docket Number:
- FDA-2021-D-0166
- Issued by:
- Guidance Issuing OfficeCenter for Drug Evaluation and ResearchCenter for Biologics Evaluation and ResearchCenter for Devices and Radiological Health
The International Council for Harmonisation (ICH) guidance for industry Q12 Technical and Regulatory Considerations for Pharmaceutical Product Lifecycle Management and its Annexes (ICH Q12, May 2021) provide a framework to facilitate the management of postapproval chemistry, manufacturing, and controls (CMC) changes in a more predictable and efficient manner. ICH Q12 includes regulatory tools and enablers with associated guiding principles that should enhance industry’s ability to manage postapproval changes and increase transparency between industry and regulatory authorities, supporting innovation and continual improvement.
FEDERAL REGISTER NOTICES:
- The original notice announcing availability of the guidance
- The new notice about the reopening of the comment period
Submit Comments
You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))
If unable to submit comments online, please mail written comments to:
Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852
All written comments should be identified with this document's docket number: FDA-2021-D-0166.
ICH Q12: Implementation Considerations for FDA-Regulated Products Guidance for Industry DRAFT GUIDANCE This guidance document is being distributed for comment purposes only. Comments and suggestions regarding this draft document should be submitted within 60 days of publication in the Federal Register of the notice announcing the availability of the draft guidance. Submit electronic comments to https://www.regulations.gov. Submit written comments to the Dockets Management Staff (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852. All comments should be identified with the docket number listed in the notice of availability that publishes in the Federal Register. For questions regarding this draft document, contact (CDER) Ashley Boam 301-796-6341, (CBER) Office of Communication, Outreach and Development, 800-835-4709 or 240-402-8010, or (CDRH) CDRH product jurisdiction officer at CDRHProductJurisdiction@fda.hhs.gov. U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for Biologics Evaluation and Research (CBER) Center for Devices and Radiological Health (CDRH) Office of Combination Products (OCP) May 2021 Pharmaceutical Quality/CMC
ICH Q12: Implementation Considerations for FDA-Regulated Products Guidance for Industry Additional copies are available from: Office of Communications, Division of Drug Information Center for Drug Evaluation and Research Food and Drug Administration 10001 New Hampshire Ave., Hillandale Bldg., 4 th Floor Silver Spring, MD 20993-0002 Phone: 855-543-3784 or 301-796-3400; Fax: 301-431-6353 Email: druginfo@fda.hhs.gov https://www.fda.gov/drugs/guidance-compliance-regulatory-information/guidances-drugs and/or Office of Communication, Outreach and Development Center for Biologics Evaluation and Research Food and Drug Administration 10903 New Hampshire Ave., Bldg. 71, Room 3128 Silver Spring, MD 20993-0002 Phone: 800-835-4709 or 240-402-8010 Email: ocod@fda.hhs.gov https://www.fda.gov/vaccines-blood-biologics/guidance-compliance-regulatory-information-biologics/biologics-guidances and/or Office of Policy Center for Devices and Radiological Health Food and Drug Administration 10903 New Hampshire Ave., Bldg. 66, Room 5431 Silver Spring, MD 20993-0002 Email: CDRH-Guidance@fda.hhs.gov https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/guidance-documents-medical-devices- and-radiation-emitting-products U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for Biologics Evaluation and Research (CBER) Center for Devices and Radiological Health (CDRH) Office of Combination Products (OCP) May 2021 Pharmaceutical Quality/CMC
Contains Nonbinding Recommendations Draft — Not for Implementation TABLE OF CONTENTS I. INTRODUCTION............................................................................................................. 1 II. CONSIDERATIONS FOR IMPLEMENTATION ....................................................... 2 A. Introduction .................................................................................................................................... 2 B. Categorization of Postapproval CMC Changes .......................................................................... 2 C. Established Conditions .................................................................................................................. 2 1. Submission of ECs ........................................................................................................................... 3 2. Identification of ECs ........................................................................................................................ 4 3. Identification of ECs for a Drug Substance or Drug Product in a Drug Master File ..................... 5 4. Identification of ECs for Device Constituent Parts of Combination Products ................................ 6 5. Changes to Approved ECs ............................................................................................................... 7 6. Postapproval Submissions in Accordance With Approved ECs ...................................................... 7 7. Maintenance of the Application ....................................................................................................... 7 D. Postapproval Change Management Protocol .............................................................................. 8 E. Product Lifecycle Management Document.................................................................................. 8 F. Pharmaceutical Quality System and Change Management ....................................................... 9 G. Relationship Between Regulatory Assessment and Inspection .................................................. 9 H. Structured Approaches for Frequent CMC Postapproval Changes and Stability Data Approaches To Support the Evaluation of CMC Changes ............................................................... 10 APPENDIX A. ESTABLISHED CONDITIONS FOR COMBINATION PRODUCTS WITH DEVICE CONSTITUENT PARTS .............................................................................. 11 APPENDIX B. DECISION TREE FOR IDENTIFYING ESTABLISHED CONDITIONS AND REPORTING CATEGORIES FOR DEVICE CONSTITUENT PARTS ................... 12 APPENDIX C. PRODUCT LIFECYCLE MANAGEMENT DOCUMENT EXAMPLE ... 13
Contains Nonbinding Recommendations Draft — Not for Implementation 1 ICH Q12: Implementation Considerations for 1 FDA-Regulated Products 2 Guidance for Industry1 3 4 5 This draft guidance, when finalized, will represent the current thinking of the Food and Drug 6 Administration (FDA or Agency) on this topic. It does not establish any rights for any person and is not 7 binding on FDA or the public. You can use an alternative approach if it satisfies the requirements of the 8 applicable statutes and regulations. To discuss an alternative approach, contact the FDA staff responsible 9 for this guidance as listed on the title page. 10 11 12 13 14 I. INTRODUCTION 15 16 The International Council for Harmonisation (ICH) guidance for industry Q12 Technical and 17 Regulatory Considerations for Pharmaceutical Product Lifecycle Management and its Annexes 18 (ICH Q12, May 2021) provide a framework to facilitate the management of postapproval 19 chemistry, manufacturing, and controls (CMC) changes in a more predictable and efficient 20 manner.2 ICH Q12 includes regulatory tools and enablers with associated guiding principles that 21 should enhance industry’s ability to manage postapproval changes and increase transparency 22 between industry and regulatory authorities, supporting innovation and continual improvement. 23 24 This guidance should be read in conjunction with ICH Q12, which this guidance complements 25 by clarifying how the ICH Q12 tools and enablers can be implemented within the U.S. regulatory 26 system. These guidances apply to drug substances and drug products3 that are the subject of new 27 drug applications (NDAs), biologics license applications (BLAs), abbreviated new drug 28 applications (ANDAs), and supplements to these applications regulated by the Center for Drug 29 Evaluation and Research (CDER) and the Center for Biologics Evaluation and Research 30 (CBER). They also apply to combination products4 with device constituent parts that are the 31 subject of NDAs, BLAs, ANDAs, and supplements to these applications regulated by CDER and 32 CBER. 33 34 The contents of this document do not have the force and effect of law and are not meant to bind 35 the public in any way, unless specifically incorporated into a contract. This document is intended 36 1 This guidance has been prepared by the Office of Pharmaceutical Quality in the Center for Drug Evaluation and Research in cooperation with the Center for Biologics Evaluation and Research, Center for Devices and Radiological Health, and Office of Combination Products at the Food and Drug Administration. 2 We update guidances periodically. For the most recent version of a guidance, check the FDA guidance web page at https://www.fda.gov/regulatory-information/search-fda-guidance-documents. 3 For the purposes of this guidance, drug substance and drug product include biological drug substances and drug products. 4 See 21 CFR 3.2(e).
Contains Nonbinding Recommendations Draft — Not for Implementation 2 only to provide clarity to the public regarding existing requirements under the law. FDA 37 guidance documents, including this guidance, should be viewed only as recommendations, unless 38 specific regulatory or statutory requirements are cited. The use of the word should in Agency 39 guidances means that something is suggested or recommended, but not required. 40 41 42 II. CONSIDERATIONS FOR IMPLEMENTATION 43 44 The considerations below follow the order of the sections in ICH Q12 and are specific to FDA’s 45 implementation. 46 47 A. Introduction 48 49 As stated above, ICH Q12 and this guidance apply to drug substances and drug products that are 50 the subject of NDAs, BLAs, ANDAs, and supplements to these applications regulated by CDER 51 and CBER. They also apply to combination products with device constituent parts that are the 52 subject of NDAs, BLAs, ANDAs, and supplements to these applications regulated by CDER and 53 CBER. 54 55 B. Categorization of Postapproval CMC Changes 56 57 ICH Q12 describes two categories for regulatory communications: prior approval and 58 notification. In the U.S. regulatory system, prior approval means a prior approval supplement 59 (PAS), notification moderate means a changes being effected-30 (CBE-30) supplement, and 60 notification low means a CBE-0 supplement or annual report.5 As indicated in ICH Q12, the 61 lowest risk changes are managed and documented within the pharmaceutical quality system 62 (PQS) and do not need to be reported, but they may be verified during a surveillance or other 63 inspection. 64 65 C. Established Conditions 66 67 ICH Q12 defines established conditions (ECs) as legally binding information considered 68 necessary to assure product quality. As a consequence, any change to ECs necessitates a 69 submission to the regulatory authority. This is consistent with FDA’s regulations at 21 CFR 70 314.70(a)(1)(i), 314.97(a), and 601.12(a)(1). Although these regulations do not explicitly specify 71 what constitutes an EC, they do set forth a risk-based paradigm for reporting changes. In 72 addition, existing FDA guidance documents describe a broad set of postapproval changes and 73 make recommendations for how they should be reported.6 The risk-based paradigm set forth in 74 the regulations and the recommendations in these associated guidance documents can help 75 5 See 21 CFR 314.70, 314.97, and 601.12. 6 See, e.g., FDA’s Scale-Up and Postapproval Changes (SUPAC) guidances and the guidances for industry Changes to an Approved Application for Specified Biotechnology and Specified Synthetic Biological Products (July 1997), Changes to an Approved NDA or ANDA (April 2004), and CMC Postapproval Manufacturing Changes To Be Documented in Annual Reports (March 2014). Insofar as a guidance adjusts reporting categories pursuant to section 506A of the Federal Food, Drug, and Cosmetic Act (FD&C Act) and 21 CFR 314.70 and 601.12, these adjustments do establish legally enforceable responsibilities and are not only recommendations.
Contains Nonbinding Recommendations Draft — Not for Implementation 3 applicants determine which elements of an application that FDA would typically consider to be 76 ECs. ICH Q12 further helps applicants gain clarity around which elements of the product, 77 manufacturing process, facilities and equipment, and control strategy in their applications are 78 considered to be ECs and therefore require reporting if changed. 79 80 Proposing elements to be considered ECs that may differ from those FDA typically considers to 81 be ECs based on the risk-based paradigm set forth in the regulations and the recommendations 82 contained in guidance is voluntary. Applicants may propose ECs in their original applications or 83 in a PAS.7 If specific ECs are not proposed, ECs would be those that FDA typically considers to 84 be ECs based on the risk-based paradigm set forth in the regulations and the recommendations 85 contained in guidance regarding postapproval changes. 86 87 Applicants may also propose reporting categories for changes to ECs. The reporting categories 88 proposed may be consistent with the risk-based paradigm set forth in the regulations and the 89 recommendations contained in guidance. Alternatively, an increased understanding of the risk to 90 product quality posed by a change to an EC may support a proposal for reduced reporting 91 categories. 92 93 1. Submission of ECs 94 95 To ensure clarity regarding ECs when submitting an original NDA, BLA, or ANDA, applicants 96 should: 97 98 • Include one of the following statements in the cover letter: 99 100 o Specific ECs are proposed. 101 102 o Specific ECs are not proposed; postapproval changes will follow the regulations and 103 the recommendations in guidance.8 104 105 • Include one of the following statements in eCTD section 3.2.R of the application9: 106 107 o Specific ECs are proposed. Specific reporting categories for changes to those ECs are 108 not proposed and therefore will follow the regulations and the recommendations in 109 guidance. 110 7 Plans to include specific ECs in an original application or PAS are an appropriate topic for presubmission meetings with FDA, if available. 8 In this guidance, FDA recommends as an option that the applicant state that it intends to follow both the regulations and the recommendations in guidance. In most cases, an applicant could choose to follow the regulations while taking a different approach than recommended in guidance and explain how its deviations from the guidance would satisfy the applicable legal requirements. However, if a guidance adjusts reporting categories pursuant to section 506A of the FD&C Act and 21 CFR 314.70 and 601.12, these adjustments do establish legally enforceable responsibilities and are not only recommendations. 9 eCTD=electronic common technical document, the standard format for submitting applications, supplements, and other information to FDA. For more information, see https://www.fda.gov/drugs/electronic-regulatory-submission- and-review/electronic-common-technical-document-ectd.
Contains Nonbinding Recommendations Draft — Not for Implementation 4 111 o Specific ECs are proposed. Specific reporting categories for changes to those ECs are 112 proposed. 113 114 o Specific ECs are not proposed; postapproval changes will follow the regulations and 115 the recommendations in guidance. 116 117 When proposing ECs in a PAS, applicants should: 118 119 • State in the cover letter that ECs are proposed. 120 121 • Include one of the following statements in eCTD section 3.2.R: 122 123 o Specific ECs are proposed. Specific reporting categories for changes to those ECs are 124 not proposed and therefore will follow the regulations and the recommendations in 125 guidance. 126 127 o Specific ECs are proposed. Specific reporting categories for changes to those ECs are 128 proposed. 129 130 If applicants choose to propose specific ECs, they can propose them for the entire CMC section 131 of the application (module 3: Quality) or for a subset of information provided in module 3 (e.g., 132 for eCTD section 3.2.S or for eCTD section 3.2.P.3.3). If a limited set of ECs is proposed, such 133 as for an individual unit operation in the manufacturing process, the applicant should list for the 134 applicable eCTD sections all of the ECs for that unit operation. Additionally, the applicant 135 should include a statement that, for the appropriate eCTD sections, changes to those unit 136 operations for which ECs are not proposed will be reported according to the regulations and the 137 recommendations in guidance. 138 139 A complete list of proposed ECs, their reporting categories (if proposed), and the eCTD locations 140 for their scientific justification should be included in the Product Lifecycle Management (PLCM) 141 document in eCTD section 3.2.R. See section II.E in this guidance for more information about 142 the PLCM document. See section II.C.2 for information about the justification for proposed ECs, 143 including where such justifications should reside in the application. 144 145 2. Identification of ECs 146 147 When proposing specific ECs, applicants should include a scientific justification for their 148 selection in the relevant parts of module 3 of the application. In this justification, applicants 149 should address both the identification of particular parameters or attributes as ECs and the 150 proposed reporting categories (if applicable). 151 152 a. Parameters and attributes identified as ECs 153 154 In the justification, applicants should explain how they identified the parameters or attributes that 155 are proposed to be ECs and why others that might typically be considered to be ECs (considering 156
Contains Nonbinding Recommendations Draft — Not for Implementation 5 the regulations and the recommendations in guidance as described above) were not. For example, 157 in a manufacturing unit operation for which three of five process parameters are proposed to be 158 ECs (in a parameter-based approach), the applicant should explain how the three parameters 159 were identified as being ECs and why the other two were not. A description of the applicant’s 160 risk assessment process (including tools) used to identify particular parameters or attributes as 161 ECs, the criticality assessment conducted to determine the level of impact of each parameter on 162 product quality, and the supporting information for each (e.g., fundamental knowledge, empirical 163 investigation, prior knowledge from experience with other products, commercial experience) 164 should be provided. See also sections 3.2.3.1 “Identification of ECs for Manufacturing 165 Processes” and 3.2.3.2 “Identification of ECs for Analytical Procedures” in ICH Q12. Applicants 166 should not propose to manage attributes as ECs if changes to those attributes would require 167 submission of a new original application (e.g., change in dosage form for an NDA). 168 169 For parameters or attributes identified as ECs that have associated acceptable ranges, applicants 170 should include a justification for the proposed ranges. 171 172 b. Reporting categories 173 174 For each proposed EC, the applicant may identify a proposed reporting category: PAS, CBE-30, 175 CBE-0, or annual report. If this reporting category differs from that recommended in existing 176 guidance, the applicant should provide a justification in the relevant section of module 3. This 177 justification can be part of or complementary to the justification provided to support 178 identification of the EC. If the applicant chooses not to propose reporting categories, the 179 applicant should include a statement that reporting categories for changes to those ECs will 180 follow the regulations and the recommendations in guidance. 181 182 3. Identification of ECs for a Drug Substance or Drug Product in a Drug Master 183 File 184 185 If information about the drug substance or drug product is incorporated in an application by 186 reference to a Type II drug master file (DMF),10 ECs associated with that drug substance or drug 187 product can be proposed, but they should only be proposed as part of the application.11 In such 188 cases, the DMF holder will need to share sufficient information with the applicant12 so that 189 proposed ECs and their associated reporting categories can be specified in and approved as part 190 of the application. The justification in support of the identification of ECs can be located in the 191 DMF, as long as the application refers to the specific location of the justification in the DMF. 192 The applicant should provide a justification for proposed reporting categories in the application 193 because the justification will generally be specific to the final drug product and its conditions of 194 approval (e.g., dosing, route of administration) rather than generalizable across different products 195 10 FDA has proposed rulemaking to codify the expectation that BLA applicants submit drug substance, drug substance intermediate, and drug product information directly to the BLA rather than incorporating it by reference to a master file. See 84 FR 30968. 11 The same approach applies to information incorporated by reference from other types of DMFs (e.g., Types III, IV, or V, where applicable). 12 See guidance for industry Contract Manufacturing Arrangements for Drugs: Quality Agreements (November 2016).
Contains Nonbinding Recommendations Draft — Not for Implementation 6 incorporating the same drug substance or drug product. When ECs are approved for an 196 application that incorporates by reference information from a DMF, the applicant maintains 197 responsibility for reporting changes to ECs; however, the applicant should share the relevant 198 approved ECs with the DMF holder to ensure that the DMF holder is aware of how changes to 199 the referenced drug substance or drug product will be managed by the applicant. 200 201 4. Identification of ECs for Device Constituent Parts of Combination Products 202 203 Applicants may propose ECs for the device constituent part of a combination product in an 204 NDA, BLA, or ANDA. The combination product as a whole, including the roles and interactions 205 of the constituent parts, should be considered in proposing ECs and reporting categories for the 206 product, including in relation to each constituent part. The focus of this section is identifying 207 ECs for device constituent parts in particular. One approach is to assess the “characteristics of 208 the product that are essential for its safe and proper use”13 (primary characteristics) relating to 209 the device constituent part and to identify the associated ECs. The primary characteristics for the 210 device constituent part of a combination product generally include: 211 212 • Functions essential for safe use based on risk management principles (see, e.g., ICH 213 guidance for industry Q9 Quality Risk Management (June 2006); ISO 14971:2019, 214 Medical devices—Application of risk management to medical devices). 215 216 • Design features essential to achieve delivery of the labeled dose to a specific body site 217 (e.g., for a device constituent part that provides drug delivery). 218 219 • Characteristics that impact the drug constituent part’s critical quality attributes. 220 221 Applicants should consider the following as potential ECs: 222 223 • Design features that are primary characteristics. 224 225 • Manufacturing process elements for the device constituent part that need to be controlled 226 to ensure a primary characteristic. 227 228 • Other control strategy elements for the device constituent part that ensure a primary 229 characteristic. 230 231 See appendix A for a description of elements that are generally considered to be ECs for device 232 constituent parts. As with identifying ECs for drug constituent parts, whether it may be 233 appropriate to propose ECs for device constituent parts or a lesser reporting category for them 234 will vary based on extent of product and process understanding, knowledge gained from design 235 and development to manage the risks, and evidence to support designation and justification of 236 the primary characteristics. Similarly, the level of risk associated with a change to an EC will 237 13 See ISO 13485:2016, Medical devices—Quality management systems—Requirements for regulatory purposes, Section 7.3.4 Product realization—Design and development—Design and development outputs. ISO=International Organization for Standardization.
Contains Nonbinding Recommendations Draft — Not for Implementation 7 determine the appropriate reporting category for that change. See appendix B for an approach 238 that can be followed for device constituent parts when identifying ECs and their associated 239 reporting categories. 240 241 5. Changes to Approved ECs 242 243 As indicated in ICH Q12, applicants may propose to add, eliminate, or make changes to 244 approved ECs or revisions to their associated reporting categories through: 245 246 • Submission of a supplement. 247 248 o Addition of an EC that provides increased assurance of the quality of the drug 249 substance or product with a reporting category provided for in the regulations or 250 recommended in guidance should be submitted as a CBE-0 (see § 314.70(c)(6)). 251 252 o All other changes should be submitted as a PAS. 253 254 • Submission of a postapproval change management protocol (PACMP; see section II.D in 255 this guidance). 256 257 • Fulfillment of a previously approved postapproval commitment submitted in a 258 supplement or annual report. For example, this may be done if an EC was approved with 259 an allowance for a modification pending the availability—and submission, where 260 specified—of certain information. 261 262 6. Postapproval Submissions in Accordance With Approved ECs 263 264 When submitting postapproval supplements and annual reports to report CMC changes in 265 accordance with approved ECs, applicants should state in the cover letter that the submission 266 contains changes made in accordance with previously approved ECs. These submissions should 267 also include an updated PLCM (see section II.E in this guidance). 268 269 7. Maintenance of the Application 270 271 As indicated in ICH Q12, maintenance of the application is subject to regional requirements. To 272 ensure that FDA has access to up-to-date analytical procedures, applicants should include in the 273 annual report a copy of all analytical procedures that have been appropriately modified during 274 the reporting period without a submission (i.e., managed only through the PQS as changes did 275 not relate to ECs). This information is intended to be for information only and typically is not 276 subject to review unless it is determined that changes to ECs for the procedure were in fact made. 277 Applicants should include this information in eCTD section 1.13.5 or 1.13.7 with a statement 278 such as “For information only: Changes made to analytical methods, including those not 279 requiring submission.” 280 281
Contains Nonbinding Recommendations Draft — Not for Implementation 8 D. Postapproval Change Management Protocol 282 283 In the U.S. regulatory system, PACMPs are referred to as comparability protocols and are 284 voluntary. These protocols differ from the use of ECs in one important respect. Both 285 comparability protocols and the use of ECs address the element to be reported if changed and the 286 reporting mechanism, but a comparability protocol requires the tests and studies to support a 287 future change to an EC to be specified at the time of the protocol submission (21 CFR 288 314.70(e)). 289 290 FDA recommends that comparability protocols be submitted in eCTD section 3.2.R. For a 291 comparability protocol that includes one or more changes that apply to more than one product, 292 see draft guidance for industry Comparability Protocols for Human Drugs and Biologics: 293 Chemistry, Manufacturing, and Controls Information (April 2016)14 for information on the 294 appropriate content and format of the submission and the circumstances in which a comparability 295 protocol might be useful. 296 297 E. Product Lifecycle Management Document 298 299 As described in ICH Q12, the PLCM document should include proposed ECs, reporting 300 categories for making changes to approved ECs, a list of comparability protocols (if submitted), 301 and postapproval CMC commitments, if applicable. FDA recommends that the PLCM document 302 be provided in tabular format in eCTD section 3.2.R, with specific references to the submission 303 sequence, eCTD section number, and page number where each EC’s scientific justification can 304 be found. FDA further recommends that the PLCM indicate the manufacturing sites (preferably 305 by facility establishment identifier (FEI) number) where an EC will be implemented.15 For 306 example, if there are two drug product manufacturing sites named in the application, but the 307 manufacturing process-related ECs will only be associated with the manufacturing at one of 308 those two sites, the FEI number of this site should be specified in the PLCM for those process- 309 related ECs. It is then assumed that ECs for operations at the other facility will follow the 310 regulations and the recommendations in guidance. If the ECs will be implemented at both sites, 311 both FEI numbers should be listed. Applicants should attribute FEIs to ECs in the PLCM at the 312 most inclusive level (e.g., one FEI listed at the beginning of 3.2.P if all subsequent ECs in the 313 section are associated with one drug product facility). See appendix C in this guidance for an 314 example of a PLCM document. 315 316 Applicants should provide an updated PLCM document with each supplement or annual report 317 that reports changes to approved ECs. If no specific ECs are proposed, submission of a PLCM 318 document is not necessary. 319 320 14 More specifically, see the appendix, section A, question 6. When final, this guidance will represent FDA’s current thinking on this topic. 15 Such facilities may include facilities responsible for design control for a combination product. See guidance for industry Current Good Manufacturing Practice Requirements for Combination Products (January 2017).
Contains Nonbinding Recommendations Draft — Not for Implementation 9 F. Pharmaceutical Quality System and Change Management 321 322 As indicated in ICH Q12, in addition to compliance with current good manufacturing practice 323 (CGMP) requirements,16 an effective PQS is necessary to support the use of the tools in that 324 guidance. However, also as noted in that guidance, FDA will not require a specific inspection 325 before an applicant can make use of the principles in the guidance. FDA’s assessment of the 326 effectiveness of the PQS will generally be informed by routine inspections conducted by FDA 327 and capable foreign regulators17 and other available information. Management should employ 328 the principles of ICH guidance for industry Q10 Pharmaceutical Quality System (April 2009) to 329 ensure the continued capability of the PQS to support the management of ECs. 330 331 Because of the importance of the PQS in supporting the use of ICH Q12 tools, when an applicant 332 proposes to introduce a new manufacturing site after approval, the applicant should not assume 333 that the initially approved ECs will apply. Instead, if ECs are proposed, applicants should include 334 the following in supplements that propose a new site: 335 336 • Reassessment of the relevant ECs. 337 338 • Justification for proposed changes to ECs as a result of this reassessment in the relevant 339 parts of module 3 of the application. 340 341 FDA will also consider information included in a supplement that supports a determination that 342 the new site deserves the same level of regulatory flexibility regarding postapproval changes as 343 the site included in the original application. The determination of PQS capability will consider 344 factors such as whether the new site is operated under the same PQS as the original; the state of 345 regulatory compliance determined by FDA and other national drug regulatory authorities; 346 conformance with ICH Q10, especially as concerns change management practices; and 347 conformance with other applicable change management regulations and policies. 348 349 As indicated in ICH Q12, inspection observations that raise concern regarding the effectiveness 350 of the PQS, and change management in particular, may lead to a need to modify previously 351 approved ECs, reporting categories, or comparability protocols until such time as the PQS 352 effectiveness has returned to an acceptable state. In these cases, FDA intends to communicate the 353 impact of such findings on previously approved ECs, reporting categories, and comparability 354 protocols with facilities and applicants, as appropriate. 355 356 G. Relationship Between Regulatory Assessment and Inspection 357 358 The use of ICH Q12 tools, such as ECs, is not expected to change FDA’s processes for how 359 information is assessed as part of the application or from a facility inspection. Similarly, it does 360 16 See section 510(a)(2)(B) of the FD&C Act; 21 CFR parts 4, 210, 211, and 600; and guidance for industry Current Good Manufacturing Practice Requirements for Combination Products. 17 Capability determinations are made in accordance with section 809 of the FD&C Act; see also https://www.fda.gov/international-programs/international-arrangements/mutual-recognition-agreement-mra.
Contains Nonbinding Recommendations Draft — Not for Implementation 10 not change the expectations regarding the type of information to be submitted in an application 361 or the information that is to be available for an inspection. 362 363 H. Structured Approaches for Frequent CMC Postapproval Changes and 364 Stability Data Approaches To Support the Evaluation of CMC Changes 365 366 Sections VIII and IX of ICH Q12 and section II of the ICH Q12 Annexes provide alternative 367 approaches for certain CMC postapproval changes. FDA supports the use of such approaches. 368 FDA also encourages applicants to gain its feedback before proposing or implementing novel 369 approaches. 370
Contains Nonbinding Recommendations Draft — Not for Implementation 11 APPENDIX A. ESTABLISHED CONDITIONS FOR COMBINATION PRODUCTS 371 WITH DEVICE CONSTITUENT PARTS 372 373 The combination product as a whole, including the roles and interactions of the constituent parts, 374 should be considered in proposing established conditions (ECs) and reporting categories for the 375 product, including in relation to each constituent part. This appendix provides general guidance 376 about the elements that are generally considered ECs for the device constituent part of 377 combination products.1 It does not contain a complete list of ECs for a device constituent part; 378 each application should include a justification for the identification of proposed ECs. 379 380 The following are generally considered ECs for the device constituent part of combination 381 products: 382 383 • Identification of the device: If purchased from a third party, manufacturer identifiers for 384 the device (e.g., brand name); references to device clearance or approval (if applicable). 385 386 • Description and design features: Device description; principle (e.g., mechanical, 387 electrical) and mechanism (e.g., spray, mixing) of operation for delivery of the drug 388 product; design features that are primary characteristics; materials of construction in 389 direct or indirect contact with the drug product and patient. 390 391 • Manufacturers: Name, address, and responsibilities for sites that perform assembly, 392 packaging, and testing of the device constituent part. 393 394 • Manufacturing (e.g., assembly): Unit operations and sequence in the manufacturing 395 process; manufacturing process parameters, material attributes, and in-process controls, 396 where variability impacts primary characteristics. 397 398 • Release/expiry specification and associated test methods for attributes that ensure 399 primary characteristics. 400 1 Similar to appendix 1 in International Council for Harmonisation guidance for industry Q12 Technical and Regulatory Considerations for Pharmaceutical Product Lifecycle Management (May 2021).
Contains Nonbinding Recommendations Draft — Not for Implementation 12 APPENDIX B. DECISION TREE FOR IDENTIFYING ESTABLISHED CONDITIONS 401 AND REPORTING CATEGORIES FOR DEVICE CONSTITUENT PARTS 402 403 The decision tree below can be used to guide the identification of established conditions and 404 associated reporting categories for the device constituent part of a combination product. 405 Answer these questions about the characteristic: • Is it essential for safe use based on risk management?* • Is it essential to achieve delivery of the labeled dose? • Does it impact the drug product’s CQA? Primary characteristic: Essential for safe and proper use** Other characteristic: Not essential for safe and proper use Design features that are primary characteristics Manufacturing process elements that need to be controlled to ensure primary characteristics Other control strategy elements that ensure primary characteristics Define the design, process, and control strategy elements that are ECs Elements of design, process, and controls for other characteristics are not ECs PAS CBE-30 or CBE-0 Not Reported What is the level of potential risk associated with the proposed change? “Yes” to any “No” to all Moderate to Low High *As determined using risk management principles in ICH guidance for industry Q9 Quality Risk Management (June 2006) or ISO 14971:2019, Medical devices—Application of risk management to medical devices. **See ISO 13485:2016, Medical devices—Quality management systems—Requirements for regulatory purposes, Section 7.3.4 Product realization—Design and development—Design and development outputs. Note: CQA=critical quality attribute; EC=established condition; PAS=prior approval supplement; CBE=changes being effected; ICH=International Council for Harmonisation; ISO=International Organization for Standardization. 406
Contains Nonbinding Recommendations Draft — Not for Implementation 13 APPENDIX C. PRODUCT LIFECYCLE MANAGEMENT DOCUMENT EXAMPLE 407 408 In this example,1 where the applicant proposes to follow FDA regulations and the 409 recommendations in guidance for a change to a particular established condition, the reporting 410 category has been left blank. 411 412 1 Adapted from annex IF in the International Council for Harmonisation guidance for industry Q12 Technical and Regulatory Considerations for Pharmaceutical Product Lifecycle Management: Annexes (May 2021). eCTD Section Established Conditions (Note that identification and justification of each EC is presented in the relevant eCTD section) Reporting Category When Making a Change to the EC The ECs below are to be implemented at the following sites: FEI xxxxxx FEI yyyyyy Seq 0001, 3.2.P.3.3, p. 4 The manufacturing process consists of the following sequence of unit operations: 1. Powder blending 2. Roller compaction 3. Tablet compression 4. Film coating PAS Seq 0001, 3.2.P.3.3, p. 44 1. Powder Blending The active substance and three excipients are mixed together. The following process parameters are defined as ECs. Seq 0003, 3.2.P.3.3, p. 45 Operating principle: Diffusion mixing PAS Seq 0001, 3.2.P.3.3, pp. 45–47 Equipment type: V-blender Change to equipment of same operating principle: AR Seq 0001, 3.2.P.3.2, p. 8, and 3.2.P.3.3, pp. 48–49 Scale: 200 kg Increase up to 10x: AR Increase beyond 10x: CBE-0 Seq 0004, 3.2.P.3.4, pp. 10–15 Design space for blending process parameters Blend speed: 10-20 rpm CBE-30
Contains Nonbinding Recommendations Draft — Not for Implementation 14 Note: eCTD=electronic common technical document; EC=established condition; FEI=facility establishment 413 identifier; PAS=prior approval supplement; AR=annual report; CBE=changes being effected. 414 Blend time: 15-25 minutes 2. Roller Compaction 3. Tablet Compression 4. Film Coating