Medical Gases — Current Good Manufacturing Practice: Draft Guidance for Industry
us-fda-guidance-602d22b5965edc8fd6a8c9732737c3efMedical Gases — Current Good Manufacturing Practice: Draft Guidance for IndustryDraft Guidance
Administrative draft guidance, not enacted legislation. This captured publisher edition preserves the listed document and attachments; later developments are not independently consolidated.
- Docket Number:
- FDA-2003-D-0431
- Issued by:
- Guidance Issuing OfficeCenter for Drug Evaluation and ResearchCenter for Veterinary MedicineOffice of Inspections and InvestigationsOffice of the Commissioner, Office of Clinical Policy and Programs, Office of Combination Products
The Food and Drug Administration (FDA, Agency, or we) is announcing the availability of a draft guidance for industry entitled “Medical Gases -- Current Good Manufacturing Practice.” This guidance is intended to assist manufacturers of medical gases in complying with regulations for current good manufacturing practice (CGMP) that become effective on December 18, 2025 (note, the conforming amendments to the CGMP requirements for combination products became effective February 2, 2026). These regulations are specific to medical gases for human and animal use and, like all CGMP requirements, contain the minimum requirements to ensure that manufacturing processes operate under a state of control to meet prespecified quality standards for identity, strength, quality, and purity, but are tailored more narrowly to how medical gases are manufactured, packaged, labeled, stored, and distributed. This draft guidance is being issued to reflect new and revised regulations in several areas to reduce the regulatory burden, as appropriate, for the medical gas industry. This draft guidance revises and replaces the draft guidance entitled “Current Good Manufacturing Practice for Medical Gases” issued in June 2017.
Submit Comments
You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))
If unable to submit comments online, please mail written comments to:
Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852
All written comments should be identified with this document's docket number: FDA-2003-D-0431 .
Medical Gases — Current Good Manufacturing Practice Guidance for Industry DRAFT GUIDANCE This guidance document is being distributed for comment purposes only. Comments and suggestions regarding this draft document should be submitted within 60 days of publication in the Federal Register of the notice announcing the availability of the draft guidance. Submit electronic comments to https://www.regulations.gov. Submit written comments to the Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852. All comments should be identified with the docket number listed in the notice of availability that publishes in the Federal Register. For questions regarding this draft document, contact (CDER) at cder-quality- policy@fda.hhs.gov. U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for Veterinary Medicine (CVM) Office of Inspections and Investigations (OII) Office of Combination Products (OCP) in the Office of the Commissioner November 2025 Pharmaceutical Quality/Manufacturing Standards (CGMP) Revision 2
Medical Gases — Current Good Manufacturing Practice Guidance for Industry Additional copies are available from: Division of Drug Information Center for Drug Evaluation and Research Food and Drug Administration Phone: 855-543-3784 or 301-796-3400 Email: druginfo@fda.hhs.gov https://www.fda.gov/drugs/guidance-compliance-regulatory-information/guidances-drugs and/or Policy and Regulations Staff Center for Veterinary Medicine Food and Drug Administration 5001 Campus Dr., College Park, MD 20740-3835 https://www.fda.gov/animal-veterinary/guidance-regulations/guidance-industry U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for Veterinary Medicine (CVM) Office of Inspections and Investigations (OII) Office of Combination Products (OCP) in the Office of the Commissioner November 2025 Pharmaceutical Quality/Manufacturing Standards (CGMP) Revision 2
Contains Nonbinding Recommendations Draft — Not for Implementation i TABLE OF CONTENTS I. INTRODUCTION............................................................................................................. 1 II. BACKGROUND ............................................................................................................... 2 A. Definition of Medical Gas and Designated Medical Gas ............................................................ 2 B. CGMP Statutory and Regulatory Requirements ........................................................................ 2 III. ORGANIZATION AND PERSONNEL ......................................................................... 3 A. Quality Unit Responsibilities ........................................................................................................ 3 B. Quality Agreements With Suppliers ............................................................................................ 4 C. Personnel Qualifications and Responsibilities............................................................................. 5 IV. BUILDINGS AND FACILITIES .................................................................................... 5 V. EQUIPMENT .................................................................................................................... 6 A. Equipment Maintenance and Cleaning........................................................................................ 6 B. Automatic, Mechanical, and Electronic Equipment ................................................................... 7 1. Equipment Qualification .................................................................................................................. 7 2. Equipment Calibration..................................................................................................................... 7 3. Computerized Systems ..................................................................................................................... 8 VI. CONTROL OF INCOMING DMG, COMPONENTS, AND MEDICAL GAS CONTAINERS AND CLOSURES.................................................................................. 9 A. Receipt of Incoming DMGs and Components ............................................................................. 9 1. Certificate of Analysis ...................................................................................................................... 9 2. Supplier Qualification.................................................................................................................... 10 B. Medical Gas Containers and Closures ....................................................................................... 10 1. General Requirements ................................................................................................................... 10 2. Prefill Inspections .......................................................................................................................... 11 VII. PRODUCTION AND PROCESS CONTROLS........................................................... 14 A. Charge-In of Components and Incoming DMGs ...................................................................... 15 B. Sampling and Testing of In-Process Materials.......................................................................... 15 C. Vacuum Evacuation of High-Pressure Cylinders ..................................................................... 15 D. Filling Procedure Checks ............................................................................................................ 16 1. Temperature and Pressure Readings ............................................................................................. 16 2. Valve Assembly Leak Testing ......................................................................................................... 16 3. Heat-of-Compression Check .......................................................................................................... 17 VIII. PACKAGING AND LABELING CONTROL............................................................. 17 A. Materials Examination and Usage ............................................................................................. 17 B. Labeling Issuance ......................................................................................................................... 18
Contains Nonbinding Recommendations Draft — Not for Implementation ii C. Packaging and Labeling Operations .......................................................................................... 18 IX. HOLDING AND DISTRIBUTION ............................................................................... 19 X. LABORATORY CONTROLS ...................................................................................... 20 A. Instrument Calibration ............................................................................................................... 20 B. Testing and Release for Distribution.......................................................................................... 20 C. Test Method and Alternative Test Method Validation............................................................. 21 D. Stability Testing and Expiration Dating .................................................................................... 22 XI. RECORDS ....................................................................................................................... 23 A. General Requirements ................................................................................................................. 23 1. Record Retention and Availability ................................................................................................. 23 2. Maintenance of Written Records ................................................................................................... 23 3. Equipment Calibration, Checks, and Inspections .......................................................................... 24 4. Computer Validation Data............................................................................................................. 24 5. Process Validation Data ................................................................................................................ 24 B. Equipment Cleaning and Use Log .............................................................................................. 24 C. Records for Components, Medical Gas Containers and Closures, and Labeling .................. 25 D. Master Production and Control Records................................................................................... 25 E. Batch Production and Control Records ..................................................................................... 25 F. Production Record Review.......................................................................................................... 27 G. Laboratory Records ..................................................................................................................... 27 H. Distribution Records.................................................................................................................... 29 I. Complaint Files ............................................................................................................................ 29 XII. RETURNED AND SALVAGED MEDICAL GASES ................................................. 30 XIII. ADAPTERS ..................................................................................................................... 31 GLOSSARY................................................................................................................................. 32
Contains Nonbinding Recommendations Draft — Not for Implementation 1 Medical Gases — Current Good Manufacturing Practice 1 Guidance for Industry1 2 3 4 5 This draft guidance, when finalized, will represent the current thinking of the Food and Drug 6 Administration (FDA or Agency) on this topic. It does not establish any rights for any person and is not 7 binding on FDA or the public. You can use an alternative approach if it satisfies the requirements of the 8 applicable statutes and regulations. To discuss an alternative approach, contact the FDA staff responsible 9 for this guidance as listed on the title page. 10 11 12 13 I. INTRODUCTION 14 15 This guidance is intended to assist manufacturers2 of medical gases in complying with current 16 good manufacturing practice (CGMP) requirements. CGMP requirements, when adequately put 17 into practice, help to ensure that medical gases meet appropriate quality standards, and prevent 18 mix-ups, deviations, failures, and errors.3 19 20 On June 18, 2024, FDA established new and revised regulations tailored to medical gases,4 21 including CGMP requirements codified in part 213 (21 CFR part 213). Before these 22 requirements were implemented, medical gas manufacturers were subject to the CGMP 23 regulations for finished pharmaceuticals in parts 210 and 211 (21 CFR parts 210 and 211). The 24 draft guidance for industry Current Good Manufacturing Practice for Medical Gases (June 25 2017) described how to comply with those requirements. This guidance revises and replaces the 26 2017 draft and describes how to comply with the CGMP requirements for medical gases in part 27 213. This revision includes among its recommendations clarification on ensuring the reliability 28 of a supplier’s capabilities; protection against container closure leaks; appropriate cleaning and 29 maintenance of buildings, facilities, and equipment used in medical gas manufacture; prevention 30 of labeling and product mix-ups; circumstances requiring stability testing, expiration testing, or 31 both; and handling of returned and salvaged medical gases. 32 33 1 This guidance has been prepared by the Center for Drug Evaluation and Research, in cooperation with the Center for Veterinary Medicine, the Office of Inspections and Investigations, and the Office of Combination Products in the Office of the Commissioner at the Food and Drug Administration. 2 The Glossary defines many of the terms for purposes of this guidance. Words or phrases found in the Glossary appear in bold italics at first mention. 3 For more information about CGMP regulations, see FDA’s Facts About Current Good Manufacturing Practice (CGMP) web page at https://www.fda.gov/drugs/pharmaceutical-quality-resources/facts-about-current-good- manufacturing-practice-cgmp. 4 See 89 FR 51738.
Contains Nonbinding Recommendations Draft — Not for Implementation 2 In general, FDA’s guidance documents do not establish legally enforceable responsibilities. 34 Instead, guidances describe the Agency’s current thinking on a topic and should be viewed only 35 as recommendations, unless specific regulatory or statutory requirements are cited. The use of 36 the word should in Agency guidances means that something is suggested or recommended, but 37 not required. 38 39 40 II. BACKGROUND 41 42 A. Definition of Medical Gas and Designated Medical Gas 43 44 This guidance applies to medical gases as defined in section 575(2) of the Federal Food, Drug, 45 and Cosmetic Act (FD&C Act) (21 U.S.C. 360ddd(2)). In this section, the term medical gas 46 means a drug that is manufactured or stored in a liquefied, nonliquefied, or cryogenic state and 47 administered as a gas. Medical gases include designated medical gases (DMGs) as defined in 48 section 575(1) of the FD&C Act; medically appropriate combinations of DMGs; medical gases 49 that are approved under an application submitted to FDA under section 505 of the FD&C Act (21 50 U.S.C. 355) for administration to humans or under section 512 of the FD&C Act (21 U.S.C. 51 360b) for administration to animals; and any marketed unapproved drugs that are medical gases. 52 Additionally, some medical gases are marketed as part of a combination product, such as a 53 medical gas marketed with a device constituent part.5 54 55 Section 575(1) of the FD&C Act defines a DMG as any of the following gases that meet the 56 standards in an official compendium:6 oxygen, nitrogen, nitrous oxide, carbon dioxide, helium, 57 carbon monoxide, and medical air. The United States Pharmacopeia and National Formulary 58 (USP-NF) is the applicable compendium for DMGs. Section 575(1)(H) of the FD&C Act 59 authorizes FDA to add to the list of DMGs any other medical gas it deems appropriate. 60 61 The term medical gas does not include gases that are used as excipients in drug products that are 62 not medical gases (e.g., a propellant for an inhalation drug); gases that serve as processing aids in 63 drug manufacturing (e.g., a nitrogen overlay to prevent oxidation of an active pharmaceutical 64 ingredient during manufacturing); or gases that do not meet the definition of a drug under section 65 201(g)(1) of the FD&C Act (21 U.S.C. 321(g)(1)) (e.g., gases for food or industrial applications, 66 gases used for device testing and verification activities, or gases used to clean or purge medical 67 gas containers or medical gas pipelines). Gases that are not medical gases are outside the scope 68 of this guidance. 69 70 B. CGMP Statutory and Regulatory Requirements 71 72 Under section 501(a)(2)(B) of the FD&C Act (21 U.S.C. 351(a)(2)(B)), a drug, including a 73 medical gas, is deemed to be adulterated if “the methods used in, or the facilities or controls used 74 5 A combination product is a product comprised of two or more different types of products (i.e., a combination of a drug, device, and/or biological product with one another), as described in 21 CFR Part 3. A constituent part is a drug, device, or biological product that is part of a combination product (see 21 CFR 4.2). 6 See section 201(j) of the FD&C Act.
Contains Nonbinding Recommendations Draft — Not for Implementation 3 for, its manufacture, processing, packing, or holding do not conform to or are not operated or 75 administered in conformity with current good manufacturing practice.” Section 501 of the 76 FD&C Act states that “the term current good manufacturing practice includes the 77 implementation of oversight and controls over the manufacture of drugs to ensure quality, 78 including managing the risk of and establishing the safety of raw materials, materials used in the 79 manufacturing of drugs, and finished drug products.” To implement oversight and controls, 80 FDA strongly encourages medical gas manufacturers to establish an effective pharmaceutical 81 quality system (PQS).7 An effective PQS enables the manufacturer’s quality unit to use 82 information obtained about the product and the process to assess risks to manufacturing 83 performance and drug quality and to identify opportunities to address those risks through 84 changes to manufacturing practices.8 This approach reflects the principle that quality should be 85 built into the product; testing alone cannot ensure product quality. 86 87 The regulations implementing the statutory requirements for medical gas are located in part 213. 88 They cover the same categories as the general CGMP drug regulations in parts 210 and 211 for 89 finished pharmaceuticals but reflect differences in how medical gases are manufactured, 90 processed, packed, and held. Differences include the fact that generally, medical gases are 91 manufactured, stored, combined, and distributed under pressure in closed systems, which reduce 92 the risk of contamination (e.g., dust, dirt, moisture, and unacceptable levels of impurities); are 93 not expected to expire or chemically degrade under ordinary storage conditions; and are 94 marketed in containers and closures that are typically reused many times. 95 96 CGMP requirements for combination products are addressed in 21 CFR part 4, subpart A. The 97 combination product CGMP requirements that apply to each constituent part apply to the 98 combination product they constitute. Because of the different CGMP requirements, a 99 streamlined approach to demonstrate CGMP compliance while avoiding unnecessary 100 redundancies is available to combination products, as described in part 4A. The medical gas 101 final rule (see footnote 4) amended certain requirements in part 4A to reflect the new CGMP 102 requirements for medical gases under part 213. The amendments to part 4A include the CGMP 103 requirements applicable to a combination product that includes a drug constituent part that is a 104 medical gas (see 21 CFR 4.3(e)) and how certain combination products that include a drug 105 constituent part that is a medical gas can comply with the CGMP requirements, including use of 106 a streamlined approach (see 21 CFR 4.4). 107 108 109 III. ORGANIZATION AND PERSONNEL 110 111 A. Quality Unit Responsibilities 112 113 The quality unit is responsible for quality oversight throughout manufacturing. Specifically, the 114 quality unit has the responsibility and authority to approve or reject all components, medical gas 115 7 See the International Council for Harmonisation (ICH) guidance for industry Q10 Pharmaceutical Quality System (April 2009). We update guidances periodically. For the most recent version of a guidance, check the FDA guidance web page at https://www.fda.gov/regulatory-information/search-fda-guidance-documents. 8 See the ICH guidance for industry Q9(R1) Quality Risk Management (May 2023).
Contains Nonbinding Recommendations Draft — Not for Implementation 4 containers and closures, in-process materials, packaging material, labeling, and medical gases, 116 and the authority to review production records to ensure that no errors have occurred or, if errors 117 have occurred, that they have been fully investigated (§ 213.22(a)). Quality unit responsibilities 118 also include approving or rejecting medical gases manufactured, processed, packed, or held 119 under contract by another company (§ 213.22(a)). To fulfill its responsibilities, the quality unit 120 must have available to them adequate laboratory facilities for the testing and approval or 121 rejection of components, medical gas containers and closures, packaging materials, in-process 122 materials, and medical gases (§ 213.22(b)). The quality unit is also responsible for approving or 123 rejecting all procedures or specifications impacting on the identity, strength, quality, and purity 124 of the medical gas (§ 213.22(c)). The quality unit’s responsibilities and procedures must be in 125 writing, and its procedures must be followed (§ 213.22(d)). 126 127 A quality unit’s size and complexity can vary with the size and number of operations for which 128 the quality unit is responsible. For example, a manufacturer that fills only oxygen and has very 129 few employees might have only one person as the quality unit. However, larger manufacturers 130 with establishments at several locations might employ a corporate quality structure to oversee 131 their staff at multiple locations or opt for a separate quality unit at each establishment. Either 132 arrangement may satisfy the requirements of § 213.22 as long as all responsibilities of the quality 133 unit are fulfilled. 134 135 Production personnel and the quality unit should remain independent. Some medical gas 136 manufacturing establishments, however, have limited personnel, and in such cases, the 137 individuals assigned quality unit responsibilities can also perform other functions (§ 213.22(e)). 138 These individuals, regardless of their production functions or other roles, implement the controls 139 and review the results of manufacture to ensure that product quality standards established by the 140 manufacturer are met. Appropriate written controls must be in place to ensure any other 141 functions are performed separately from quality unit responsibilities and that such other 142 functions do not interfere with the quality unit’s responsibilities or subordinate the quality unit’s 143 responsibilities to any other unit (§ 213.22(e)). For example, in instances where there are limited 144 personnel, it would not be appropriate for personnel to check their own work. 145 146 Each person who performs quality unit functions must be adequately trained and experienced in 147 all quality unit tasks assigned (§ 213.25(a)). For additional information see section III.C., 148 Personnel Qualifications and Responsibilities. In addition, FDA recommends that each person 149 who is part of the quality unit be identified by function and title in written procedures to ensure 150 that quality unit responsibilities are fulfilled. 151 152 B. Quality Agreements With Suppliers 153 154 The quality unit’s procedures should provide for written quality agreements with suppliers of 155 goods and services to ensure compliance with CGMP. The quality agreement should clearly 156 describe the provided goods or services, quality specifications, and communication mechanisms 157 between the contracting parties.9 FDA recommends that CGMP responsibilities and the 158 9 For more information on quality agreements, see the guidance for industry Contract Manufacturing Arrangements for Drugs: Quality Agreements (November 2016). The term quality specifications refers to any quality controls (e.g., drug specifications, process parameters, responsibilities) established to ensure quality during manufacturing.
Contains Nonbinding Recommendations Draft — Not for Implementation 5 communication processes for reporting complaints and changes that could be critical to drug 159 quality be defined in written quality agreements with suppliers. For example, timely 160 communications about process changes are important because such changes could affect the 161 composition of the gas supplied. 162 163 C. Personnel Qualifications and Responsibilities 164 165 Under § 213.25(a), all personnel engaged in the manufacture, processing, packing, or holding of 166 a medical gas, including production personnel, repackers, transfillers, and those driving to 167 customer sites to distribute medical gas, must have sufficient education, training, and experience, 168 or any combination thereof, necessary to perform their assigned functions. Section 213.25(a) 169 specifies the following: 170 171 • Such training must be in both the particular operations that the employee performs and in 172 CGMP requirements related to their assigned functions. 173 174 • The CGMP training must be conducted by qualified individuals on a continuing basis and 175 with sufficient frequency to assure that employees remain familiar with CGMP 176 requirements applicable to their job functions. To comply with this requirement, FDA 177 recommends that CGMP training be provided at least annually. 178 179 • Manufacturers must maintain written documentation, for each employee, of the 180 completion of employee training, including the date of the training, the type of training, 181 and the results of any completion criteria, such as test results. 182 183 Similarly, any consultants hired to advise on medical gas manufacturing must have the necessary 184 education, training, and experience, or any combination thereof, to advise on the subject for 185 which they are retained (§ 213.34). 186 187 Manufacturers must have an adequate number of qualified personnel to perform the manufacture, 188 processing, packing, or holding of each medical gas (§ 213.25(b)). What would constitute 189 adequate personnel would depend in part on the size and complexity of the performed 190 operations. Additionally, any personnel entering limited-access areas of a manufacturing facility 191 (e.g., certain storage areas) must be authorized to avoid improper manufacture or mishandling of 192 medical gas (§ 213.25(c)). 193 194 195 IV. BUILDINGS AND FACILITIES 196 197 The standard industry practice of manufacturing multiple gases at the same facility and refilling 198 labeled medical gas containers underscores the importance of building and facility design as a 199 control method for proper operation. Buildings and facilities must be of adequate design, 200 including adequate space, for the orderly placement of equipment and materials to prevent mix- 201 ups between components, incoming DMGs, medical gas containers and closures, labeling, in- 202 process materials, and medical gases (§ 213.42(a)(1)). For example, the design should include 203
Contains Nonbinding Recommendations Draft — Not for Implementation 6 sufficient lighting for personnel to read labels. The design and construction must also allow for 204 adequate cleaning, maintenance, and proper operations (§ 213.42(a)(2)). 205 206 Manufacturers must perform operations within specifically defined areas of adequate size to 207 prevent contamination or mix-ups during manufacturing (§ 213.42(b)(1)). For example, because 208 facilities may reuse labeled medical gas containers, filled and unfilled containers with identical 209 labeling may only be distinguishable if they are in well-defined areas. Manufacturers must 210 design the flow of components, incoming DMGs, medical gas containers and closures, labeling, 211 in-process, materials, and medical gases through the buildings and facilities in a manner to 212 prevent contamination and mix-ups (§ 213.42(b)(2)). Manufacturers must maintain 213 establishments in a clean condition, as specified in written procedures, so as to assure the safety, 214 identity, strength, quality, and purity of the medical gas (§ 213.42(c)). 215 216 Outdoor spaces and delivery truck beds can be appropriate areas to conduct certain operations 217 (e.g., storage and handling) for medical gases in pressurized containers. For example, industrial 218 and medical gases could be physically separated in the warehouse or in the delivery truck. To 219 separate these areas from other spaces as required under § 213.42(b)(1), manufacturers should 220 use identifiers such as signage, floor demarcation, physical dividers, or tagging. 221 222 223 V. EQUIPMENT 224 225 Equipment used in the manufacture, processing, packing, or holding of a medical gas must be of 226 appropriate design and adequate size and be suitably located to facilitate operations for its 227 intended use and any necessary cleaning and maintenance (§ 213.63). In addition, equipment 228 must be constructed so that surfaces that contact components, in-process materials, or medical 229 gases are not reactive, additive, or absorptive (§ 213.65(a)), and any substances required for 230 manufacturing operations, such as lubricants or coolants, do not come into contact with 231 components, containers, closures, in-process materials, or medical gases (§ 213.65(b)). 232 233 A. Equipment Maintenance and Cleaning 234 235 Under § 213.67 manufacturers are required to establish, maintain, and follow written procedures 236 for adequate cleaning and maintenance of equipment used in the manufacture, processing, 237 packing, or holding of medical gases. These procedures must include, but are not limited to the 238 following: 239 240 • Assignment of responsibility for cleaning and maintaining equipment 241 242 • Maintenance and cleaning schedules, including, where appropriate, sanitizing schedules 243 244 • A description in sufficient detail of the methods, equipment, and materials used in 245 cleaning and maintenance operations, and the methods of disassembling and 246 reassembling equipment as necessary to assure proper cleaning and maintenance 247 248 • Removal or obliteration of previous batch identification 249
Contains Nonbinding Recommendations Draft — Not for Implementation 7 250 • Protection of clean equipment from contamination before use 251 252 • Inspection of equipment for cleanliness immediately before use 253 254 FDA recommends that medical gas manufacturers ensure that personnel can access all equipment 255 that must be cleaned, and tailor their equipment maintenance, cleaning, and inspection 256 procedures to match the type and complexity of the particular operation and to prevent 257 malfunctions or contamination. 258 259 Manufacturers should ensure equipment used in the manufacture of medical gases (e.g., 260 manifolds, pigtails, valve assemblies, hoses, gauges) is cleaned or verified as clean (e.g., 261 ensuring no residual cleaning agents are present) before initial use and after potential exposure to 262 a contaminant. Closed pressurized systems used for filling medical gases (e.g., manifolds) can 263 be considered acceptable without cleaning between batches, unless exposed to a contaminant. 264 To prevent contamination, manufacturers should ensure that open ends are appropriately covered 265 (e.g., with physical caps) and that valves that are critical for preventing contamination are 266 properly maintained and cleaned. Because industrial and medical gases can be filled on the same 267 manifold rack, there should be procedures in place and followed to prevent cross-contamination. 268 269 B. Automatic, Mechanical, and Electronic Equipment 270 271 Automatic, mechanical, and electronic equipment must be routinely calibrated, inspected, and 272 checked according to a written program designed to ensure proper performance (§ 213.68(a)). 273 274 1. Equipment Qualification 275 276 Manufacturers should perform equipment qualification to verify that equipment used in medical 277 gas manufacturing is installed, operates, and performs as intended. FDA recommends that 278 equipment be qualified for the anticipated temperatures and pressures used during filling. 279 Manufacturers should also qualify manifold valves to ensure they are appropriately designed to 280 prevent mix-ups during filling operations and shown to prevent contamination of the medical 281 gas. Similarly, other valves that are critical to the prevention of drug contamination, such as 282 check valves used in filling systems, should be qualified for the particular use. 283 284 2. Equipment Calibration 285 286 Manufacturers can calibrate equipment by either following the equipment manufacturer’s 287 recommended calibration schedule or a schedule based on the medical gas manufacturer’s 288 experience using the equipment (e.g., the manufacturer’s frequency of use). Medical gas 289 manufacturers can reference the equipment manufacturer’s instruction manual in their written 290 procedures if the manual is available for on-site use. 291 292 FDA recommends that manufacturers do the following: 293 294
Contains Nonbinding Recommendations Draft — Not for Implementation 8 • Check the performance of vacuum gauges daily to ensure that the needle on the gauge 295 returns to zero when there is no vacuum or pressure (above atmospheric pressure). 296 297 • Calibrate vacuum and pressure gauges at least annually against an established standard 298 (e.g., a standard from the National Institute of Standards and Technology). Low-pressure 299 gauges and flow meters used in filling portable cryogenic medical gas containers do not 300 require calibration, but manufacturers should ensure that they function properly for their 301 intended use. 302 303 • Calibrate thermometers according to the equipment manufacturer’s instructions. If the 304 instructions do not specify frequency, calibration should be conducted at least annually. 305 306 3. Computerized Systems 307 308 Manufacturers must appropriately validate computerized systems that record, store, or use data in 309 the manufacturing, processing, and holding of medical gases (e.g., computerized systems used in 310 test analyses) (e.g., § 213.68(b)). The validation should address the intended use of the 311 computerized system, and the extent of validation studies should be commensurate with the risk 312 posed by the automated system.10 In addition, manufacturers must maintain a backup file of any 313 data entered into the computer system unless certain data, such as calculations performed in 314 connection with laboratory analysis, are eliminated by computerization or automated processes 315 (§ 213.68(c)). 316 317 Manufacturers must use appropriate change controls whenever modifications are made to 318 computer systems so that changes do not adversely affect data integrity11 or product quality, and 319 records of such modifications must be maintained (§ 213.68(d)).12 Appropriate personnel with 320 expertise should assess the potential impact these changes could have on quality both before the 321 change is approved and after the change is implemented, as well as any need for revalidation. 322 323 Manufacturers should enable audit trail functions for computerized systems to capture changes 324 and maintain complete data about CGMP activities. This information is important for the 325 manufacturer to trace CGMP activities throughout the supply chain when retrospective review or 326 investigation is warranted. 327 10 To comply with the requirement in § 213.68(b) when using commercially available software (e.g., off-the-shelf software), the medical gas manufacturer, as the party with regulatory responsibility, should assess the adequacy of the off-the-shelf software and must ensure it is validated, so that the software specifications conform to user needs and intended uses. See sections 6.1 through 6.3 of the guidance for industry and FDA staff General Principles of Software Validation (January 2002). Although the guidance focuses on the validation of medical device software, it states that “this document is based on generally recognized software validation principles and, therefore, can be applied to any software.” 11 See the guidance for industry Data Integrity and Compliance With Drug CGMP: Questions and Answers (December 2018). 12 For FDA’s current thinking on the use of computerized systems for maintaining electronic records, see the guidance for industry Part 11, Electronic Records; Electronic Signatures—Scope and Application (August 2003).
Contains Nonbinding Recommendations Draft — Not for Implementation 9 328 329 VI. CONTROL OF INCOMING DMG, COMPONENTS, AND MEDICAL GAS 330 CONTAINERS AND CLOSURES 331 332 Manufacturers must control and assess the quality of incoming DMGs, components, containers, 333 and closures as specified in §§ 213.80, 213.82, and 213.84 because materials of poor quality can 334 adversely affect the quality and safety of the medical gas. Materials can contain impurities (e.g., 335 caused by poor air quality conditions in air separation units or inadequate purification during 336 manufacturing), or a change to the manufacturing process could affect quality (e.g., an 337 adjustment in a manufacturing step could alter the composition of a supplied gas). Any rejected 338 incoming DMGs, components, and medical gas containers and closures must be handled in 339 accordance with § 213.89, which requires rejected items to be identified and quarantined to 340 prevent their use in manufacturing or processing operations for which they are unsuitable. In 341 addition, manufacturers must document and assess the data on rejected material and take 342 appropriate corrective action (§ 213.192(a)). 343 344 A. Receipt of Incoming DMGs and Components 345 346 Upon receipt of each shipment of each incoming DMG, the manufacturer must perform an 347 identity test (§ 213.82(b)). The manufacturer must also perform either full compendial testing on 348 the gas and record the results, or verify and record that a signed certificate of analysis (COA) 349 from the supplier accompanied the shipment (§ 213.82(a)(1)). For incoming DMGs, if the 350 supplier is not the original manufacturer, the supplier must include complete information from 351 the original manufacturer’s COA (§ 213.82(a)(2)). Manufacturers must sample, test, and 352 approve or reject each component, as appropriate, prior to use either by performing testing for 353 conformance with written specifications or by an identity test on the component accompanied by 354 an acceptable COA from the supplier (§ 213.84(c)). 355 356 1. Certificate of Analysis 357 358 For incoming DMGs, if a manufacturer relies on the supplier’s COA, the COA must include (i) 359 the supplier’s name; (ii) the name of the incoming DMG; (iii) the lot number (also referred to as 360 control number or batch number) or other unique identification number; (iv) the actual 361 analytical result obtained for strength, as well as the results of other performed tests; (v) 362 identification of the test method(s) used for analysis; (vi) the incoming DMG’s new drug 363 application (NDA) and/or new animal drug application (NADA) number; and (vii) the supplier 364 representative’s signature and the date of signature (§ 213.82(a)(1)). 365 366 For a component, if a manufacturer relies on the supplier’s COA, it should include the following 367 data: (i) the supplier’s name; (ii) the name of the component; (iii) the lot number or other unique 368 identification number; (iv) the type of test or examination to evaluate conformance with written 369 procedures and specifications; (v) the method(s) used; (vi) specification limits; (vii) all test 370 results; and (viii) the supplier representative’s signature and the date of signature. 371 372
Contains Nonbinding Recommendations Draft — Not for Implementation 10 2. Supplier Qualification 373 374 Reliance on information provided in a COA requires a high degree of confidence in the supplier. 375 For both incoming DMGs and components, when the manufacturer is relying on a supplier’s 376 COA to verify that the incoming DMG or components meet specifications beyond identity 377 testing, the manufacturer must establish and maintain a program to ensure the reliability of the 378 supplier’s capabilities through appropriate assessment and testing procedures (§§ 213.82(a)(2), 379 213.84(a), and 213.84(c)). To establish the reliability of the supplier’s analyses, the 380 manufacturer, a contract-testing laboratory, or another third party can perform testing at 381 appropriate intervals. Manufacturers can begin the supplier qualification process, for example, 382 by fully testing several batches and examining the provided gas, containers, and closures. 383 Written procedures should address how to qualify and approve suppliers and to periodically 384 verify the qualifications of approved suppliers. This can be done by conducting audits (e.g., on- 385 site for the initial audit and, as needed, remotely for subsequent audits), analyzing trends in the 386 quality of received goods, testing, and evaluating the timeliness of the supplier’s responses to 387 complaints. 388 389 B. Medical Gas Containers and Closures 390 391 1. General Requirements 392 393 The quality unit must examine medical gas containers and closures, including valves, for 394 conformance with appropriate written procedures and specifications, and approve or reject them, 395 before manufacturing or filling (§§ 213.22(a), 213.84(a) and (c) and 213.89). A manufacturer 396 should reexamine a medical gas container or closure when it is exposed to adverse storage 397 conditions that could cause deterioration or contamination of the medical gas. Manufacturers 398 must also protect against container and closure leaks (§ 213.84(b)). This includes performing 399 leak tests on containers and closures and investigating any container closure defects identified 400 either during production or upon receipt of a leak complaint (§ 213.198(a)). Valve replacements 401 (e.g., due to a valve defect) should be performed as part of a planned program, with mitigation 402 strategies, and investigated as a deviation. See section VIII.D.2., Valve Assembly Leak Testing 403 for a discussion of leak testing. 404 405 Medical gas containers and closures must at all times be handled and stored in a manner to 406 prevent contamination and mix-ups (§ 213.80(b)). Accordingly, high-pressure cylinders exposed 407 to the elements and hoses used to fill cryogenic containers should have caps or other protective 408 means to prevent contamination. Medical gas containers and closures must not be reactive, 409 additive, or absorptive so as to alter the safety, identity, strength, quality, or purity of a medical 410 gas beyond the official or established requirements and must provide adequate protection against 411 foreseeable external factors and use that can cause deterioration or contamination (§ 213.94(a) 412 and (b)). Medical gas containers and closures must also be clean for their intended use 413 (§ 213.94(c)). Thus, medical gas containers and closures should be cleaned before initial use and 414 after exposure to a contaminant. To comply with the requirements in §§ 213.84 and 213.94(c), 415 manufacturers should implement appropriate cleaning and retesting procedures when converting 416 a container’s use from industrial grade gas to medical gas, or if there is reason to believe there 417 was previous industrial use. 418
Contains Nonbinding Recommendations Draft — Not for Implementation 11 419 Portable cryogenic containers that are not manufactured with permanent gas-specific use outlet 420 connections (e.g., those that have been silver-brazed) must have gas-specific use outlet 421 connections that are attached to the valve body so that they cannot be readily removed or 422 replaced (without making the valve inoperable and preventing the container’s use) except by the 423 manufacturer (§ 213.94(e)(1)). If a gas-specific use outlet connection can be readily removed 424 and replaced, a container holding one gas could be inadvertently connected to the supply system 425 for another type of gas, which could cause serious injury or death because of the gas mix-up. 426 427 Manufacturers should not use vapor recovery systems during carbon dioxide delivery unless the 428 manufacturer has adequately addressed potential contamination in a risk management program. 429 Contaminants present in the gaseous head space of the storage tank or container could be drawn 430 into the tank or container, thereby contaminating the carbon dioxide. 431 432 2. Prefill Inspections 433 434 Prefill inspections ensure that containers and closures are acceptable for use before filling begins. 435 This section addresses prefill inspections that evaluate containers and closures used to hold 436 incoming medical gases and store medical gases. Prefill inspections, like filling and postfilling 437 inspections, are a significant step in the manufacturing, processing, packing, or holding of the 438 medical gas produced and, therefore, must be properly documented (§ 213.189(b)). Medical gas 439 containers and closures that are rejected must be quarantined and assessed (§§ 213.84 and 440 213.89). Manufacturers should address the problem by repairing, cleaning, or replacing 441 unsuitable parts and then reinspecting. 442 443 a. External inspection of the container 444 445 Manufacturers should examine each container for dents, burns, dings, oil, grease, and other signs 446 of damage or contamination that can cause a container to be unsafe for use. 447 448 b. External inspection of valves, inlets, outlets, gauges, and connectors 449 450 Manufacturers should inspect each container’s valve assembly, connectors, and fittings to ensure 451 that they are appropriate for the medical gas. Manufacturers should examine valves, inlets, 452 outlets, gauges, and connectors for signs of damage (including fire damage), unusual wear, 453 corrosion, or the presence of debris, oil, or grease. This inspection should cover connections that 454 are brazed, welded, or equipped with a locking device. Manufacturers must ensure that portable 455 cryogenic medical gas containers and small cryogenic gas containers for use by individual 456 patients have a working gauge sufficient to assist the user in determining whether the container 457 contains an adequate supply of medical gas for continued use (§ 213.94(e)(2)). 458 459 c. Label inspection 460 461 Product labels on medical gas containers may be reused if they are legible, properly affixed to the 462 container, and otherwise meet all applicable requirements (§ 213.122(h)). Labels that are obsolete, 463 outdated, or that do not meet applicable requirements must be destroyed (§ 213.122(e)). 464
Contains Nonbinding Recommendations Draft — Not for Implementation 12 465 Manufacturers must ensure that the labeling specified for portable cryogenic medical gas 466 containers in § 201.328(a) (21 CFR 201.328(a)) is affixed to the container in a manner that does 467 not interfere with (e.g., obscure) other labeling and that each label, as well as materials used for 468 coloring medical gas containers, are reasonably resistant to fading, durable when exposed to 469 atmospheric conditions, and not readily soluble in water (§ 213.94(e)(3)). 470 471 Portable cryogenic medical gas containers must be conspicuously marked with a 360° 472 wraparound label identifying their contents (§ 201.328(a)(1)). The wraparound label must be 473 placed on the sidewall of the container as close to the top portion of the container as possible, but 474 below the top weld seam (§ 201.328(a)(1)(iv)). The name of the gas must be printed 475 continuously around the 360° wraparound label so that it can be read around the entire container 476 (e.g., Oxygen USP, Oxygen USP, Oxygen USP) (§ 201.328(a)(1)(iii)), and the lettering for the 477 name of the gas on the label must be at least 2 inches high (§ 201.328(a)(1)(ii)). For containers 478 that hold a single gas, either the lettering or the label’s background must be in the appropriate 479 color (e.g., green for oxygen) with contrasting background or lettering (i.e., lettering in the 480 designated color against a white background, or white lettering on a background of the 481 designated color) (§ 201.328(a)(1)(i)).13 482 483 The 360° wraparound label or a separate label on the portable cryogenic medical gas container 484 must include, in conspicuous lettering, the phrase For Medical Use, Medical Gas, or some 485 similar phrase that indicates the gas is for medical use (§ 201.328(a)(2)). 486 487 Although permanently mounted cryogenic containers are not required to have a 360° label, any 488 content labeling should be easily readable from all readily accessible viewing angles. 489 490 d. Color code inspection 491 492 The shoulder of each high-pressure medical gas cylinder must be colored in the color or colors 493 that correspond to the gas held in the cylinder; furthermore, the shoulder’s color or colors must 494 be visible when viewed from the top of the cylinder (§ 201.328(b)). The FDA-designated colors 495 identifying medical gases in high-pressure medical gas containers and portable cryogenic 496 medical gas containers are (§ 201.328(c)): 497 498 13 There are no specific color requirements for the 360° wraparound label for portable cryogenic medical gas containers containing a mixture of gases. As discussed in this guidance, however, there are requirements under § 201.328(a)(1)(v) addressing the color of the cryogenic container itself.
Contains Nonbinding Recommendations Draft — Not for Implementation 13 Labeling of Medical Gas Containers1 499 Medical Gas Color Medical Air Yellow Carbon Dioxide Gray Helium Brown Nitrogen Black Nitrous Oxide Blue Oxygen Green Mixture or Blend2 Colors corresponding to each component gas 3 1 Adapted from 21 CFR 201.328(c) 500 2 The terms mixture and blend refer to combinations of medical gases. 501 3 For example, green and gray for a blend of oxygen and carbon dioxide. 502 503 Unlike high-pressure cylinders, portable cryogenic medical gas containers are not required to be 504 colored in whole or in part. Portable cryogenic medical gas containers can be colored in a light- 505 reflective color (e.g., white), any color that is not an FDA-designated gas color, or no color. 506 However, if a manufacturer chooses to color a portable cryogenic container in an FDA- 507 designated color, it may only be colored, in whole or in part, with the color corresponding to the 508 gas or gases held in the container (§ 201.328(a)(1)(v)). 509 510 Manufacturers should not rely solely or primarily on color coding to identify medical gases; the 511 label should be used as the primary means of identifying the medical gas. Color coding provides 512 an additional safeguard to facilitate accurate identification and detection of potential errors. 513 514 A container filled with a DMG or medically appropriate combination of DMGs can bear a 515 statement identifying the name of the owner of the container or the address to which the 516 container should be returned after use. If the owner of the medical gas container is not the 517 manufacturer, packer, or distributor, it must be clearly stated on the container (§ 201.328(d)). 518 519 e. Inspection of high-pressure cylinders 520 521 For high-pressure cylinders, manufacturers should include the following as part of prefill 522 inspections: 523 524 • Examination of the U.S. Department of Transportation (DOT) requalification date 525 • Hammer or dead-ring test 526 • Odor inspection 527 • Venting or blow-down of cylinders 528 529 Manufacturers should examine each high-pressure cylinder for the DOT date stamped on the 530 cylinder before use to verify that each cylinder conforms with DOT requirements under 49 CFR 531 180.209 for the requalification and marking of cylinders. If the DOT requalification date has 532 been exceeded, the manufacturer should quarantine the cylinder until either DOT requirements 533 have been satisfied, or the cylinder is removed from inventory. 534 535
Contains Nonbinding Recommendations Draft — Not for Implementation 14 Manufacturers should also conduct a hammer or dead-ring test to provide information about 536 internal corrosion of steel cylinders that could lead to cylinder failure. The test consists of 537 lightly tapping the cylinder sidewall with a hammer-like instrument and listening to the sound 538 produced. A clear bell tone indicates that the cylinder is clean and free of internal corrosion. A 539 dull ring indicates the possibility of internal corrosion and warrants further investigation. 540 Manufacturers should not perform a hammer or dead-ring test on aluminum or composite 541 cylinders because the test would not indicate internal corrosion and could damage the cylinder 542 wall. 543 544 Manufacturers can use an odor test to detect the presence of a foreign gas or odor remaining in 545 the container. This test should not be performed on carbon dioxide, nitrous oxide, or toxic or 546 corrosive gases for safety reasons. This test should not be confused with odor testing conducted 547 as part of finished product testing, which may be included in a USP-NF monograph. If a 548 cylinder is empty at atmospheric pressure, manufacturers can introduce Nitrogen NF into the 549 cylinder at a predetermined pressure (i.e., a pressure that is high enough to introduce gas into the 550 container) and perform an odor test on the resulting gas. Residual pressure valves prevent the 551 cylinder from emptying completely and prevent backflow. A prefill odor test on a cylinder with 552 a qualified residual pressure valve is not necessary if the cylinder has residual pressure. 553 Manufacturers should document verification of residual pressure on the batch record. 554 555 High-pressure cylinders that manufacturers receive for refilling should be vented or blown down 556 to remove gas remaining in the cylinders. Manufacturers can omit this step if the cylinder is 557 equipped with a qualified residual pressure valve and has residual pressure. Manufacturers 558 should address the continued suitability of medical gas containers and closures after extended 559 storage by having procedures in place to ensure that they (1) are not exposed to conditions that 560 render them unfit for use, and (2) have successfully completed prefill tests. 561 562 563 VII. PRODUCTION AND PROCESS CONTROLS 564 565 Under § 213.100(a) manufacturers are required to have written procedures for production and 566 process controls designed to assure that medical gases have the identity, strength, quality, and 567 purity they purport or are represented to have. This regulation requires manufacturers to design a 568 process, including operations and controls, so that products meet these attributes, and is therefore 569 the foundation for process validation.14 For example, validation of automated filling systems 570 provides assurance that filling will be done properly. This regulation also requires that the 571 procedures, and any changes, be drafted, reviewed, and approved by the appropriate 572 organizational units and reviewed and approved by the quality unit. 573 574 During production, if contamination is found, or it seems likely that contamination occurred, the 575 manufacturer must quarantine the batch or lot of medical gas in which the contamination was 576 found until an investigation has been completed (§§ 213.84, 213.89, 213.110(c)). Given that 577 products with a USP monograph must meet the monograph standards if tested or be considered 578 14 For more information on FDA recommendations regarding process validation generally, see the guidance for industry Process Validation: General Principles and Practices (January 2011).
Contains Nonbinding Recommendations Draft — Not for Implementation 15 adulterated under section 501(b) of the FD&C Act, any tests for detecting contaminants in 579 medical gas should be equivalent to, or better than, USP compendial methods and suitable for 580 their intended use.15 For additional information, see section X.C., Test Method and Alternative 581 Test Method Validation. 582 583 When monitoring medical gas production, adverse trends of errors or deviations can indicate a 584 drift in the state of control, which warrants an appropriate investigation and an effective 585 corrective and preventative action (CAPA).16 Manufacturers should pay particular attention to 586 critical defects (e.g., leaking units, contamination) and assess whether marketed units were 587 potentially impacted. 588 589 A. Charge-In of Components and Incoming DMGs 590 591 Under § 213.101(a), a batch must be formulated with the intent to provide 100 percent of the 592 labeled or established amount of each medical gas unless a monograph or formulary specifies a 593 range. Where a monograph or formulary specifies a range for the contents of a medical gas, the 594 medical gas must be formulated with the intent to provide an amount within that specified range. 595 596 Under § 213.101(b) components and incoming DMGs added to in-process supply or final 597 product containers must be weighed or measured as appropriate. In-process and final product 598 containers must identify (1) the name of the component or DMG, or if there are multiple, the 599 name and percentage of each component or DMG; and (2) the unique lot number assigned. 600 601 B. Sampling and Testing of In-Process Materials 602 603 During production, in-process materials must be tested for identity, strength, quality, and purity 604 as appropriate, and the quality unit must approve or reject the in-process materials 605 (§ 213.110(a)). In-process control procedures and tests or examinations on appropriate samples 606 of in-process materials of each batch must be established to monitor output and to validate the 607 performance of manufacturing processes that can cause variability in the quality of the medical 608 gas (§ 213.110(b)). For example, in-process controls could include validation of processes to 609 remove contaminants. Manufacturers must identify and quarantine any rejected in-process 610 materials to prevent their use in manufacturing or processing operations for which they are 611 unsuitable (§ 213.110(c)). 612 613 C. Vacuum Evacuation of High-Pressure Cylinders 614 615 For cylinders that are not equipped with a residual pressure valve with backflow prevention, 616 FDA recommends vacuum evacuation to remove residual gases if cylinders are reused. FDA 617 recommends manufacturers use 25 or more inches of mercury vacuum to evacuate the residual 618 gases. If using less than 25 inches of vacuum (e.g., when compensating for altitude), 619 15 USP-NF General Notices, Section 5.60, Impurities and Contaminants in Official Articles is an example of an appropriate reference for a contaminant test method. 16 To identify an adverse trend, we recommend that the quality unit establish appropriate alert and action limits to ensure investigation expansion to reassess process design and control.
Contains Nonbinding Recommendations Draft — Not for Implementation 16 manufacturers should maintain data demonstrating sufficient evacuation). Changes to the 620 amount of vacuum used must be documented (§§ 213.184, 213.186, and 213.189) and available 621 for inspection (§ 213.180(a)), and the changes should be scientifically justified. 622 623 Manufacturers must maintain records of their testing or examination of any problems that occur 624 with container evacuation, such as the inability to adequately empty the cylinder of residual 625 gases (§ 213.184). 626 627 D. Filling Procedure Checks 628 629 When filling medical gas containers, filling to a predetermined and acceptable temperature or 630 pressure limit, along with the required finished product testing at § 213.165, indicates that the 631 medical gas or medical gas mixture in the container is the amount17 and type indicated by the 632 label and required by the final product specifications. Manufacturers should include the 633 following checks: 634 635 1. Temperature and Pressure Readings 636 637 A medical gas in a container will increase in pressure as the temperature of the gas rises. 638 Overfilled containers can reach dangerously high pressures if exposed to elevated temperatures, 639 even if the pressure at room temperature is at a safe level. To ensure that high-pressure 640 containers filled with DMG(s) in a gaseous state are filled correctly (i.e., contents as indicated on 641 the label), the manufacturer can attach a thermometer to one container per manifold-filling 642 sequence, or to each container if filling one at a time, and adjust the final filling pressure 643 according to a temperature/pressure chart or temperature/pressure calculations (Boyle’s Law) to 644 achieve proper content (expressed on the label in liters or cubic feet based on the filled pressure 645 at 70°F as required under 21 CFR 201.51(b)(1)). 646 647 The manufacturer must record the temperature and/or pressure readings on the batch production 648 record as part of the required documentation that each significant step in the manufacture, 649 processing, packing, or holding of the medical gas produced was accomplished (§ 213.189(b)). 650 See section XI.E., Batch Production and Control Records. 651 652 2. Valve Assembly Leak Testing 653 654 Manufacturers must take appropriate actions to identify and protect against container and closure 655 leaks, which include leak testing at the time of fill and after fill but prior to release (§ 213.84(b)). 656 Leak testing, performed at additional time points, may be warranted (e.g., if leaks are detected 657 after leaving the manufacturer) to provide sufficient assurance of the durability and suitability of 658 the container closure system throughout its period of use. 659 660 In the event the manufacturer receives a complaint involving a possible leak in a medical gas 661 container or closure, that complaint must be reviewed, evaluated, and investigated in accordance 662 with § 213.192 (§ 213.198(a)). An appropriate investigation includes a determination of the 663 17 For requirements related to the measure of contents for DMGs, see 21 CFR 201.51(b), which explains how net quantity must be expressed on the label, depending on container type.
Contains Nonbinding Recommendations Draft — Not for Implementation 17 source of a leak. Based on the results of the investigation, additional leak testing, such as upon 664 pickup, could be warranted as part of an appropriate corrective action to prevent recurrence. 665 666 During leak testing, manufacturers should test each valve assembly for leaks by spraying or 667 brushing an appropriate leak detection solution on and around the entire assembly.18 The 668 solution should be safe for use with the particular medical gas, leave no residue that is 669 flammable, and be noncorrosive to the valve assembly. Among other things, it should not 670 contain hydrocarbons, ammonia, ethylene glycol, or halide ions. Solutions containing soap are 671 not recommended because they can corrode the valve stem and leave a residue. Manufacturers 672 should perform this test while the cylinder is under pressure with the cylinder valve open. If 673 bubbles appear in the solution, there is a leak. If leaks are detected, the cylinder must be 674 removed from service and quarantined until repaired (see §§ 213.84(a) and 213.89). 675 676 3. Heat-of-Compression Check 677 678 During or immediately after filling a high-pressure cylinder, manufacturers should perform a 679 heat-of-compression check by lightly touching the exterior of the cylinder or an alternative 680 method that verifies temperature change. A warm cylinder indicates that the cylinder is filling 681 properly; a cool or cold cylinder indicates that the cylinder is not filling properly. Manufacturers 682 should investigate, in accordance with § 213.192, any cool or cold cylinders. 683 684 685 VIII. PACKAGING AND LABELING CONTROL 686 687 A. Materials Examination and Usage 688 689 To ensure the quality and legibility of medical gas labels, in particular because labels may be 690 reused (§ 213.122(h)), manufacturers must do the following: 691 692 • Representatively sample, and examine or test, labeling and packaging materials upon 693 receipt and before use in packaging or labeling of a medical gas (§ 213.122(a)). 694 695 • Approve and release for use any labeling or packaging materials that meet appropriate 696 written specifications and reject any labeling or packaging materials that do not meet 697 specifications to prevent their use in operations for which they are unsuitable 698 (§ 213.122(b)). We note that previous lot numbers on any labeling must be removed or 699 obliterated as part of equipment maintenance and cleaning requirements (§ 213.67(a)(4)). 700 701 • Maintain records of each shipment of each different labeling and packaging material, 702 indicating receipt, examination, and whether accepted or rejected (§ 213.122(c)). 703 704 • Store labels and other labeling materials for each different medical gas, strength, or 705 quantity of contents with suitable identification to avoid mix-ups (§ 213.122(d)). Labels 706 18 See ASTM International, 2021, G188-05 (Reapproved 2021), Standard Specification for Leak Detector Solutions Intended for Use on Brasses and Other Copper Alloys.
Contains Nonbinding Recommendations Draft — Not for Implementation 18 and labeling materials can be stored in the same cabinet provided they are adequately 707 separated and identified. Labels for nonmedical purposes (e.g., industrial gas) should be 708 stored in a separate area. 709 710 • Secure labeling by limiting access to authorized personnel (§ 213.122(d)). 711 712 • Destroy obsolete and outdated labeling (§ 213.122(e)). 713 714 • Include one of the following special control procedures: 715 716 - Dedicate labeling and packaging lines to each different strength of each different 717 medical gas (§ 213.122(f)(1)) 718 719 - Use appropriate electronic or electromechanical equipment to conduct a 720 100 percent examination for correct labeling during or after completion of 721 finishing operations (§ 213.122(f)(2)) 722 723 - Use visual inspection to conduct a 100-percent examination for correct labeling 724 during or after completion of labeling operations for hand-applied labeling, which 725 must be independently verified by a second person (§ 213.122(f)(3)) 726 727 • Monitor any printing devices on, or associated with, manufacturing lines used to imprint 728 labeling on the unit label or case for conformance with the print specified in the batch 729 production record (§ 213.122(g)). 730 731 B. Labeling Issuance 732 733 Manufacturers must strictly control labeling issued for use in medical gas operations to prevent 734 labeling and product mix-ups (§ 213.125(a) (21 CFR 125(a)). Appropriate controls include 735 labeling reconciliation (§ 213.125(b)) and the discarding of all excess lot number stickers or 736 decals that are unused (§ 213.125(c)). FDA recommends that labeling be issued for use in 737 medical gas labeling operations by authorized personnel only. 738 739 Manufacturers must reconcile the number of labels issued, used, and returned and evaluate any 740 discrepancies outside narrow preset limits based on historical operating data (§ 213.125(b)) and 741 investigate those discrepancies (§ 213.192). 742 743 Labeling reconciliation is waived for cut or roll labeling if manufacturers perform a 100 percent 744 examination for correct labeling in accordance with § 213.122(f)(2) and is also waived for 360° 745 wraparound labels on portable cryogenic medical gas containers (§ 213.125(b)). 746 747 C. Packaging and Labeling Operations 748 749 Manufacturers must handle packaging and labeling operations in a manner that assures that 750 correct labels, labeling, and packaging materials are used for medical gases (§ 213.130). 751 Measures include the following: 752
Contains Nonbinding Recommendations Draft — Not for Implementation 19 753 • Physically or spatially separating labeling and packaging operations for medical gas from 754 operations on other products (e.g., gases for industrial use) (§ 213.130(a)). 755 756 • Identifying any filled, unlabeled containers that have been set aside for future labeling 757 operations to prevent mislabeling (§ 213.130(b)). Each container does not need to be 758 individually identified. This requirement can be met by having signage for an area 759 designated to store these containers. 760 761 • Identification of a batch medical gas with a lot or control number (§ 213.130(c)). A 762 separate sticker or decal can be used to meet the requirement as long as the sticker or 763 decal remains adhered to the container and is legible. The sticker or decal should be 764 readily visible and should not obscure required drug information. Manufacturers should 765 consider as a batch each (1) manifold filling sequence, (2) uninterrupted filling 766 sequence, and (3) filled rail tank car, trailer, and cryogenic container. For continuous 767 manufacturing operations, a batch is the amount of medical gas produced in a unit of time 768 (e.g., in 24 hours) or quantity in a manner that assures its having uniform character and 769 quality within specified limits (§ 213.3(b)(9)), with the unit of time or quantity defined by 770 the manufacturer. 771 772 • Inspection of packaging and labeling facilities immediately before use to assure that all 773 medical gases have been removed from previous operations and any packaging or 774 labeling materials unsuitable for subsequent operations have been identified and removed 775 (§ 213.130(d) and (e)). 776 777 If a manufacturer includes an expiration date on the medical gas product labeling, it must appear 778 on the labeling in accordance with 21 CFR 201.17. The date must be supported by a stability 779 testing program as described in § 213.166. See section X.D., Stability Testing and Expiration 780 Dating for additional information and requirements. 781 782 783 IX. HOLDING AND DISTRIBUTION 784 785 Manufacturers must establish and follow written procedures describing medical gas distribution 786 (§ 213.150(a)). These procedures must include a system by which the distribution of each lot 787 can be readily determined to facilitate its recall if necessary (§ 213.150(a)). The procedures 788 should explain (1) who would evaluate distribution information, (2) how a recall would be 789 initiated, (3) who would be informed about the recall, (4) what would be done with the recalled 790 product, and (5) how these steps would be accomplished. 791 792 Manufacturers must also establish and follow written procedures describing the warehousing of 793 medical gas, including quarantine before release by the quality unit (§ 213.150(b)). 794 795 796
Contains Nonbinding Recommendations Draft — Not for Implementation 20 X. LABORATORY CONTROLS 797 798 Laboratory control requirements in § 213.160 include establishment of scientifically sound and 799 appropriate specifications, standards, sampling plans, and test procedures designed to assure that 800 components, medical gas containers and closures, in-process materials, labeling, and medical 801 gases conform to appropriate standards of identity, strength, quality, and purity (§ 213.160(b)). 802 These laboratory control mechanisms must be drafted by the appropriate organizational unit, 803 reviewed and approved by the quality unit, followed, documented at the time of performance, 804 and any deviations from the written specifications, standards, sampling plans, test procedures, or 805 other laboratory control mechanisms recorded and justified (§ 213.160(a)). For example, to 806 comply with this requirement, manufacturers must record and justify changes made to the 807 analytical method, such as a different column length or a different carrier gas (§ 213.160(a)). 808 809 A. Instrument Calibration 810 811 Laboratory controls must include a written program for the calibration, or verification of 812 calibration, for instruments, apparatus, gauges, and recording devices at suitable intervals in 813 accordance with an established written program containing specific directions, schedules, limits 814 for accuracy and precision, and provisions for remedial action in the event accuracy and/or 815 precision limits are not met (§ 213.160(b)(4)). Under this requirement, any downstream 816 entities19 that do not perform their own calibration must verify calibration. It is of particular 817 importance that instruments measuring a quality characteristic are calibrated to ensure proper 818 performance. 819 820 If using a calibration gas, manufacturers should verify that it is traceable to a nationally 821 recognized standard and that it ensures the appropriate level of precision and accuracy. The 822 COA for the calibration gas should be specific to the cylinder of calibration gas received and 823 should contain the following: 824 825 • Name and address of the supplier 826 • Name of the calibration gas 827 • Lot number or unique identification number 828 • Description of the analytical method used to assay the calibration gas 829 • Analytical results expressed quantitatively (e.g., 99.9 percent nitrogen) 830 • Statement that the calibration gas is traceable to a nationally recognized standard 831 • Responsible person’s signature and the date signed 832 833 B. Testing and Release for Distribution 834 835 For each batch of medical gas, there must be an appropriate laboratory determination of 836 satisfactory conformance to final specifications for the medical gas, including the identity and 837 strength, before release (§ 213.165(a)). An appropriate laboratory determination would include 838 19 Downstream entities are entities that manufacture, process, pack, or hold medical gases, including firms that combine, commingle, refill, or distribute DMGs and medically appropriate combinations of DMGs but are downstream from the manufacturer that initially manufactures the medical gas.
Contains Nonbinding Recommendations Draft — Not for Implementation 21 meeting the standards listed in an applicable compendium (e.g., a USP-NF monograph for 839 certified DMGs,20 relevant general USP NF chapters21) or the specifications established in an 840 application approved under section 505 or section 512 of the FD&C Act. 841 842 Written procedures must describe the method of sampling and testing, including the number of 843 units per batch that will be sampled and tested and the acceptance criteria (§ 213.165(b)). The 844 method of sampling must be suitable and verified under actual conditions of use (§ 213.165(c)). 845 Test results that fail to meet specifications must be rejected (§ 213.165(d)). Samples can be 846 retested, but retesting is not recommended until the manufacturer has conducted a thorough 847 investigation according to established written procedures.22 848 849 For cylinders filled on a multiple-outlet manifold, at least one high-pressure cylinder from each 850 manifold filling sequence should be tested for identity and strength. 851 852 For cylinders filled individually, one high-pressure cylinder per uninterrupted filling sequence 853 should be tested for identity and strength. 854 855 For mixtures containing two gases, each high-pressure cylinder in a batch should be tested for 856 both the identity and strength of one of the gases, and one cylinder from each batch should be 857 tested for the identity of the second gas. For mixtures containing three gases, each high-pressure 858 cylinder in a batch should be tested for both the identity and strength of two of the gases, and one 859 cylinder from each batch should be tested for the identity of the third gas. If mixtures contain 860 oxygen, each cylinder in the batch should be tested for the identity and strength of the oxygen. 861 862 C. Test Method and Alternative Test Method Validation 863 864 The manufacturer must demonstrate that test methods are accurate, sensitive, specific, and 865 reproducible under actual conditions of use (§ 213.165(c)) and must be approved by the quality 866 unit (§ 213.22(c)). Validation and documentation must be accomplished as described in 867 § 213.194(a)(2). 868 869 Because USP-NF monograph drug products must meet USP-NF monograph standards (section 870 501(b) of the FD&C Act), FDA recommends that manufacturers use the test methods in the 871 appropriate USP-NF monograph for the medical gas. For USP-NF test methods, a full test 872 method validation study is unnecessary, as long as the referenced method is not modified. Data 873 that verify that the USP-NF test method is accurate and reliable for testing the particular product 874 (i.e., suitable for its intended purpose) should be generated on the appropriate equipment and 875 20 See section 576(a)(2)(A) and (B) of the FD&C Act (21 U.S.C. 360ddd-1(a)(2)(A) and (B)) (a certification will not be granted if the Secretary finds that the medical gas is not a DMG, or if the request does not contain required information or otherwise lacks sufficient information to permit the Secretary to determine that the medical gas is a DMG); section 575(1) of the FD&C Act (defining each DMG as a gas that “meets the standards set forth in an official compendium”). 21 See footnote 15. 22 See the guidance for industry Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production (May 2022).
Contains Nonbinding Recommendations Draft — Not for Implementation 22 readily available. If a medical gas manufacturer relies on the equipment manufacturer’s study, 876 the medical gas manufacturer should retain a copy of the study, including the protocol and data. 877 878 Manufacturers that use approved test methods that are not USP-NF monograph methods must 879 maintain a copy of the full and complete test method validation (§ 213.165(c)). When a test is 880 approved as part of an NDA submitted under section 505(b)(1) and approved under section 881 505(c) of the FD&C Act, an abbreviated new drug application (ANDA) submitted and approved 882 under section 505(j) of the FD&C Act, a NADA submitted under section 512(b)(1) and approved 883 under section 512(c)(1) of the FD&C Act, or an abbreviated new animal drug application 884 (ANADA) submitted under section 512(b)(2) and approved under section 512(c)(2) of the FD&C 885 Act, it becomes the approved analytical test method for the medical gas. 886 887 For drug products with a USP-NF monograph, a manufacturer can establish alternative test 888 methods, as long as the USP-NF monograph standards are met or exceeded. Alternative test 889 methods must be validated (§§ 213.160(b) and 213.165(c)). The validation can be performed in 890 accordance with USP-NF General Chapter <1225> Validation of Compendial Procedures, and 891 the validation study should include data comparing it to the official test method.23 892 893 The suitability of all testing methods must be verified under the actual conditions of use 894 (§ 213.165(c)). For example, paramagnetic oxygen analyzers can give inaccurate readings when 895 used at high altitudes unless special adjustments are made. The results of these tests must be 896 fully documented (§ 213.165(c)). 897 898 Certain changes made to instrumentation can be substantive enough to be considered changes to 899 the test method itself; these changes would require additional documentation of the accuracy and 900 reliability of the method or a new validation study (see § 213.194(b)). 901 902 D. Stability Testing and Expiration Dating 903 904 For medical gases marketed under applications submitted under section 505 or section 512 of the 905 FD&C Act, any stability testing performed and any expiration dating established must be in 906 accordance with § 213.166(b), subject to the conditions established in their approved 907 applications, if any. For example, if an approved application states that both stability testing and 908 expiration dating are conditions for the safe and effective use of the medical gas, the 909 manufacturer must have a stability program to assess the stability characteristics of the medical 910 gas and the integrity of its container closure (§ 213.166(b)(1)). The stability program must 911 include the appropriate sample size, test intervals, container closure systems, and storage 912 conditions for samples retained for testing, and the results of the stability testing would be the 913 basis for determining appropriate storage conditions and any expiration date included on the 914 label (§ 213.166(b)(1)). 915 916 If an approved application does not require stability testing or expiration dating for the safe and 917 effective use of a medical gas, but the manufacturer nonetheless chooses to include an expiration 918 date on the labeling, the expiration date must be based on the results of a stability program as 919 described in § 213.166(b)(1). Similarly, stability testing and expiration dating are not required 920 23 See the guidance for industry Analytical Procedures and Methods Validation for Drugs and Biologics (July 2015).
Contains Nonbinding Recommendations Draft — Not for Implementation 23 for DMGs, but if a manufacturer of a DMG chooses to include an expiration date on the label, 921 the date should be determined by stability testing. 922 923 Manufacturers must evaluate stability periodically to ensure that the medical gas continues to 924 meet the standards for identity, strength, quality, and purity stated on the labeling to support the 925 expiration date (§ 213.166(b)(3)). 926 927 928 XI. RECORDS 929 930 This section discusses general records requirements and requirements related to records 931 maintenance for equipment cleaning and use; components, containers and closures, and labeling; 932 master production and control; batch production and control; laboratory records; distribution 933 records; record review; and complaint files. 934 935 A. General Requirements 936 937 1. Record Retention and Availability 938 939 Manufacturers must maintain all records required under part 213 for at least 3 years after the 940 distribution of the batch of medical gas, except as otherwise provided in this part (§ 213.180(c)). 941 942 All records must be readily available for authorized inspection during the retention period at the 943 establishment where the activities described in such records occurred (§ 213.180(a)). Records 944 must be originals or accurate copies of original records that are legible and stored to prevent 945 deterioration or loss (§ 213.180(b)). Electronic records that can be immediately retrieved satisfy 946 this requirement.24 947 948 2. Maintenance of Written Records 949 950 Manufacturers must maintain written records so that data therein can be used for evaluating, at 951 least annually, the quality standards of each medical gas to determine the need for changes in 952 specifications or manufacturing or control procedures (§ 213.180(d)). There must be written 953 procedures for the review of the following: 954 955 • A representative number of batches, whether approved or rejected, and where applicable, 956 records associated with the batch (§ 213.180(d)(1)) 957 958 • Complaints, recalls, returned or salvaged medical gases, and investigations conducted 959 under § 213.192 for each medical gas (§ 213.180(d)(2)) 960 961 Also, under § 213.180(e), the manufacturer must have written procedures to notify responsible 962 officials of the firm, in writing, of any recalls, reports of inspectional observations by FDA, 963 regulatory actions related to good manufacturing practices brought by FDA, or investigations 964 24 Electronic records are subject to the requirements of 21 CFR part 11. See the guidance for industry Part 11, Electronic Records; Electronic Signatures—Scope and Application.
Contains Nonbinding Recommendations Draft — Not for Implementation 24 resulting from adverse event complaints. Manufacturers should use these records to identify 965 potential product quality issues and opportunities for continuous process improvement. This can 966 be done on a company-wide or site-by-site basis. 967 968 3. Equipment Calibration, Checks, and Inspections 969 970 Manufacturers must keep records of calibration, checks, and inspections of automatic, 971 mechanical, and electronic equipment used in the manufacture, processing, packing, and holding 972 of medical gas (§ 213.68(a)). See section V.B.2., Equipment Calibration for more information 973 about equipment calibration. 974 975 4. Computer Validation Data 976 977 Manufacturers must have documentation that their automated, mechanical, and electronic 978 equipment—including computers used in the manufacturing or holding of a gas—demonstrates 979 proper performance as required under § 213.68(a) through (d), including data after modifications 980 are made to computer systems. 981 982 5. Process Validation Data 983 984 Records required under part 213 include process validation records. Such data must show that 985 the manufacturer has production and process controls designed to assure that medical gases have 986 the identity, strength, quality, and purity that they purport or are represented to possess as 987 required under § 213.100. 988 989 B. Equipment Cleaning and Use Log 990 991 Under § 213.182, manufacturers must retain cleaning records for major equipment cleaning, 992 nonroutine maintenance, and equipment use, including the date, time, product, and lot number of 993 each batch processed. 994 995 Manufacturers must document the work on individual equipment logs, which are separate 996 cleaning or maintenance records that are not associated with specific batch records. The people 997 who perform and double-check cleaning and maintenance must date and either sign or initial the 998 log indicating that the work was performed, with entries made in chronological order 999 (§ 213.182). If using equipment dedicated to the manufacture of a single medical gas, and lots or 1000 batches follow in numerical order and are manufactured in numerical sequence, individual 1001 equipment logs are not required. However, records for cleaning, maintenance, and use would be 1002 required as a part of the batch record (§ 213.182). If automated equipment is used in cleaning 1003 and maintenance in accordance with § 213.68, only the person verifying the cleaning and 1004 maintenance needs to date and sign or initial the use log. 1005 1006
Contains Nonbinding Recommendations Draft — Not for Implementation 25 C. Records for Components, Medical Gas Containers and Closures, and 1007 Labeling 1008 1009 Records associated with component, container and closure systems, and labeling requirements 1010 covered in §§ 213.84, 213.122, and 213.130 must include the following (§ 213.184): 1011 1012 • Results of any test or examination performed on components, containers, and container 1013 closure systems, including those performed as required by §§ 213.84 or 213.122, and the 1014 conclusions derived from the results 1015 1016 • Documentation of the examination and review of labels and labeling for conformity with 1017 established specifications in accordance with §§ 213.122 and 213.130 1018 1019 • Disposition of rejected medical gas components, containers and closures, and labeling 1020 1021 D. Master Production and Control Records 1022 1023 Master production and control records for each medical gas must be prepared, dated, and signed 1024 to assure uniformity from batch to batch. The manufacturer must have written procedures for the 1025 preparation of these records that must include the following (§ 213.186): 1026 1027 • The name and strength of the medical gas 1028 1029 • A complete list of components and any incoming DMGs used in manufacturing 1030 designated by names or codes sufficiently specific to indicate any special quality 1031 characteristic 1032 1033 • A description of the medical gas containers and closures, packaging materials, and labels 1034 1035 • Complete manufacturing and control instructions, sampling and testing procedures, 1036 specifications, special notations, and precautions to be followed 1037 1038 E. Batch Production and Control Records 1039 1040 Batch production and control records must be prepared for each batch of medical gas produced 1041 (§ 213.189(a)). 1042 1043 The records must contain documentation that each significant step in batch manufacture, 1044 processing, packing, or holding was accomplished, including: 1045 1046 • Dates of each significant step, including in-process and laboratory tests as applicable 1047 (§ 213.189(b)(1)). Manufacturers should not use a single entry to indicate that all 1048 significant manufacturing, processing, packing, and holding steps have been performed. 1049 1050 • Description of the container for the medical gas, including the number and size of the 1051 containers filled as applicable (§ 213.189(b)(2)). 1052
Contains Nonbinding Recommendations Draft — Not for Implementation 26 1053 • Specific identification of each component and its source or in-process material used as 1054 applicable (§ 213.189(b)(3)). 1055 1056 • Measures of components used in the course of processing, as applicable 1057 (§ 213.189(b)(4)). 1058 1059 • Testing results, including in-process test results and finished product test results 1060 (§ 213.189(b)(5)). 1061 1062 • Dated signature or initials of each person performing and directly supervising or checking 1063 each significant step in the operation (§ 213.189(b)(6)). 1064 1065 • Inspection of the packaging and labeling area before and after use (§ 213.189(b)(7)). 1066 1067 • Complete labeling control records, including specimens or copies of all labeling used and 1068 label application and reconciliation records as appropriate (§ 213.189(b)(8)). 1069 Considering that all written, printed, and graphic matter upon or accompanying an article 1070 is labeling,25 any accompanying labeling (e.g., booklets, brochures) is to be part of the 1071 complete labeling control records. Manufacturers should consider the following: 1072 1073 - A photocopy or printed digital image can be an alternative to a label specimen. 1074 Manufacturers should include a specimen of the specific lot number labeling (e.g., 1075 lot number sticker) in the batch record. 1076 1077 - At the time a container is transferred from one entity to another, each entity is 1078 responsible for the labeling on or accompanying the container. This labeling 1079 should be part of the complete labeling control records. For example, 1080 § 201.328(d) states that a container filled with a DMG or medically appropriate 1081 combination of DMGs may bear a statement identifying the name of the owner of 1082 the container or the address to which the container should be returned after use 1083 and that this statement may appear on a separate sticker or decal. In addition, this 1084 regulation requires that if the owner of the medical gas container is not the 1085 manufacturer, packer, or distributor of the medical gas, this fact must be clearly 1086 stated on the container. The above information is considered labeling and 1087 therefore is expected to be part of complete labeling control records. 1088 1089 - Labeling control records maintained by original manufacturers that fill bulk 1090 trailers may or may not include a finished product label. These manufacturers 1091 should maintain both product and lot-specific labeling. 1092 1093 • Any investigation made according to § 213.192 (§ 213.189(b)(9)). 1094 1095 25 Section 201(m) of the FD&C Act defines the term labeling to mean “all labels and other written, printed, or graphic matter (1) upon any article or any of its containers or wrappers, or (2) accompanying such article.”
Contains Nonbinding Recommendations Draft — Not for Implementation 27 Batch production records should reflect actual production practices and conditions at the time of 1096 manufacture. Transfiller pumper or filler logs can serve as batch production records if they 1097 contain the relevant information specified in § 213.189. 1098 1099 In addition, FDA recommends that batch records maintained by transfillers, including curbside 1100 vendors, document the following information: 1101 1102 • Findings from the transfiller’s inspections conducted prefill, during fill, and postfill 1103 • Number and size of the containers filled 1104 • Final temperature and pressure results 1105 • Other inspection results 1106 1107 All required batch record information should be easily located and traceable to each specific 1108 batch manufactured. Manufacturers need not maintain batch production records as one 1109 document. Each lot number should be traceable in records for batch manufacturing, labeling, 1110 testing, and release. Manufacturers that use computer-controlled equipment during manufacture 1111 should establish and follow procedures to maintain, review, and approve the manufacturing data. 1112 1113 A checkmark or other symbol should not be used in place of an actual value, such as for 1114 temperature and pressure readings, or results from purity and identity testing. 1115 1116 FDA recommends that records of nonconforming medical gas describe the rejection relative to 1117 the rest of the batch to ensure that the scope of the investigation is appropriate. For example, 1118 medical gases rejected for container or manifold leaks during filling must be documented 1119 (§§ 213.84, 213.165, and 213.189) and investigated (§ 213.192), and the scope of the 1120 investigation should extend to other batches of medical gas that could have been associated with 1121 the same failure. 1122 1123 F. Production Record Review 1124 1125 The quality unit must review and approve all manufacturing production and control records, 1126 including those for packaging and labeling, before a batch is released or distributed 1127 (§ 213.192(a)). Unexplained discrepancies or the failure of a batch to meet its specifications, 1128 including any test results falling outside of specifications, must be investigated (by the quality 1129 unit) and corrective actions taken (e.g., removing a faulty cylinder from circulation) 1130 (§ 213.192(a)). 1131 1132 Manufacturers must maintain written records of investigations that include conclusions and 1133 follow-up information (§ 213.192(a)). In the event that filling occurs when the quality unit 1134 official(s) is temporarily not on site, the quality unit may review and approve production and 1135 control records within one business day after fill (§ 213.192(b)). 1136 1137 G. Laboratory Records 1138 1139 Laboratory records related to the manufacturer of a medical gas must include complete data 1140 derived from all tests necessary to ensure compliance with established specifications and 1141
Contains Nonbinding Recommendations Draft — Not for Implementation 28 standards, including examinations and assays (§ 213.194(a)). These records include the 1142 following: 1143 1144 • Description of the sample, batch or lot number to be tested; date the sample was taken; 1145 and date the sample was tested (§ 213.194(a)(1)). 1146 1147 • The test method, the test result, how the results compare with established standards of 1148 identity, strength, quality, and purity for the component, container, in-process materials 1149 (as applicable), and medical gas tested; a record of any calculations performed and any 1150 calculated results; and the unit of measurement of the result. It would not be necessary to 1151 provide the actual calculation where the result is evident through use of simple addition 1152 and subtraction (§ 213.194(a)(2)). 1153 1154 • Where applicable, any graphs, charts, and spectra from laboratory instrumentation, 1155 properly identified to show the specific component, in-process material, or medical gas 1156 for each lot tested (§ 213.194(a)(3)). If the analytical equipment only provides a direct 1157 reading, recording the result for each test would fulfill the requirement. 1158 1159 • Initials or signature of the person who performs each test and a second person showing 1160 that the original records have been reviewed for accuracy, completeness, and compliance 1161 with established standards (§ 213.194(a)(4)). 1162 1163 In addition, manufacturers must maintain complete records of the following: 1164 1165 • Any modification of an established test method, including the reason for the modification 1166 and the data to verify that the modification produced results that are at least as accurate 1167 and reliable for the material being tested as the established method (§ 213.194(b)). 1168 1169 • Any testing and standardization of laboratory reference standards, reagents, and standard 1170 solutions (e.g., a reference gas or calibration gas used as a standard) (§ 213.194(c)). 1171 1172 • Periodic calibration or verification of calibration of laboratory instruments, apparatus, 1173 gauges, and recording devices required by § 213.160(b)(4) (§ 213.194(d)). For example, 1174 medical gas equipment calibrated by a supplier before it arrives at the laboratory can be 1175 verified rather than performed again. 1176 1177 • Stability testing performed in accordance with § 213.166 (§ 213.194(e)). 1178 1179 When a chromatographic method specified in the USP-NF (e.g., the assay method for Nitrogen 1180 NF) is used for testing, the chromatographic system must meet all system suitability 1181 requirements listed in the monograph (see section 501(b) of the FD&C Act). If the USP-NF 1182 monograph lacks specific suitability requirements, manufacturers should use USP-NF General 1183 Chapter <621> Chromatography as a reference. 1184 1185
Contains Nonbinding Recommendations Draft — Not for Implementation 29 H. Distribution Records 1186 1187 Distribution records must contain the product name, the lot or batch number, consignee’s name 1188 and address, shipping date, and quantity shipped (§ 213.196). For medically appropriate 1189 combinations of DMGs, the distribution record must include the percentage of each gas in the 1190 combination (§ 213.196). Record maintenance according to written procedures is critical for 1191 batch traceability, particularly in the event a safety-related problem leads to a product recall. 1192 1193 I. Complaint Files 1194 1195 Under § 213.198(a), the quality unit must review all written and oral complaints involving the 1196 possible failure of a medical gas to meet any of its specifications. The quality unit must also 1197 determine the need for an investigation in accordance with § 213.192, as well as determine 1198 whether the complaint contains an event reportable under 21 CFR part 230. For example, any 1199 complaint involving a possible leak of a container or closure must be reviewed, evaluated, and 1200 investigated in accordance with § 213.192. Further, if multiple complaints indicate an adverse 1201 trend, it is important that the investigation extend to evaluating state of process control and 1202 ensure appropriate CAPAs to resolve any operational problems. 1203 1204 Manufacturers must maintain written records of each complaint concerning a medical gas. These 1205 records must include (§ 213.198(b)) the following: 1206 1207 • Name of the gas 1208 • Lot or batch number 1209 • Name of the complainant 1210 • Date the complaint was received 1211 • Nature of the complaint 1212 • Response provided to the complainant 1213 • Findings of any investigation and follow-up 1214 1215 In addition to the name of the complainant, FDA recommends that manufacturers list any known 1216 contact information and the date of the FDA response to the complainant. The description of the 1217 nature of the complaint should contain sufficient information to facilitate investigative follow-up 1218 and identify adverse trends or patterns (e.g., a higher-than-expected rate of valve replacement 1219 indicating a recurring container closure defect). If the manufacturer conducts an investigation, a 1220 record of the findings should include the following: 1221 1222 • Nature of the complaint and date the complaint was received 1223 1224 • Action initially taken, including dates and the identity of the person taking the action 1225 1226 • Follow-up action taken, which can include both corrective action and preventative 1227 measures to address the complaint 1228 1229 • Whether other batches of medical gas were potentially affected 1230 1231
Contains Nonbinding Recommendations Draft — Not for Implementation 30 • Outcome regarding the problem(s) raised by the complainant 1232 1233 If an investigation was not conducted, there must be a record of the reason an investigation was 1234 found to be unnecessary and the name of the responsible person making such a determination 1235 (§ 213.198(b)). 1236 1237 1238 XII. RETURNED AND SALVAGED MEDICAL GASES 1239 1240 The requirements pertaining to the handling of returned medical gas are in 21 CFR 213.204. If 1241 the conditions under which a returned gas had been held, stored, or shipped before or during their 1242 return casts doubt (e.g., is unknown) on the safety, identity, strength, quality, or purity of the gas, 1243 or if the condition of the gas, its container, carton, or labeling, as a result of storage or shipping, 1244 casts doubt on the safety, identity, strength, quality, or purity of the gas, then the manufacturer 1245 must destroy the returned gas. However, the returned gas does not need to be destroyed if the 1246 manufacturer can prove through examination, testing, or other investigations that the gas meets 1247 appropriate standards. If the reason a medical gas was returned is because of a quality issue or 1248 implicates associated batches, the manufacturer must conduct an investigation as required under 1249 § 213.192. FDA recommends that manufacturers vent or completely evacuate the container 1250 holding returned gas. 1251 1252 FDA does not consider gas remaining in high-pressure cylinders or cryogenic containers that are 1253 returned for refilling in the normal course of business to be returned medical gas. In these cases, 1254 a small amount of gas is expected to remain in a returned container and can be reused if the 1255 medical gas meets specifications. 1256 1257 Manufacturers must have written procedures for the holding, testing, and use of returned medical 1258 gases and must maintain records of returned medical gases that include the following 1259 (§ 213.204): 1260 1261 • Name of the returned medical gas 1262 • Lot number (or control number or batch number) 1263 • Reason for the return 1264 • Quantity returned 1265 • Date of disposition 1266 • Ultimate disposition of the returned gas 1267 1268 The requirements pertaining to the salvaging of medical gas are in 21 CFR 213.208. 1269 Manufacturers may salvage medical gases that have been improperly stored and return the 1270 medical gas to the marketplace unless the container was subjected to adverse conditions that 1271 impact the identity, strength, quality, and purity of the medical gas, or the integrity of the 1272 container closure. Examples of such adverse conditions include natural disasters, facility 1273 structural damage, and changes in external factors such as temperature. In these instances, 1274 manufacturers must demonstrate through laboratory tests that the gases continue to meet all 1275 applicable standards of identity, strength, quality, and purity, and that the adverse conditions did 1276
Contains Nonbinding Recommendations Draft — Not for Implementation 31 not compromise the integrity of the container closure system. Manufactures must have written 1277 procedures for the holding, testing, and use of salvaged medical gases. 1278 1279 1280 XIII. ADAPTERS 1281 1282 For safety reasons, FDA recommends avoiding the use of adapters of any kind to circumvent the 1283 specific medical gas valves and connections associated with a medical gas. 1284 1285 In limited circumstances, such as when filling medical gas mixtures, manufacturers can use 1286 adapters (e.g., to reduce or expand the connection size for a specific medical gas while still 1287 maintaining the connection system). Manufacturers should have written procedures detailing 1288 system checks and controls to prevent mix-ups or contamination when using adapters, and to 1289 promptly identify and quarantine compromised gases if a mix-up or contamination should occur. 1290 1291
Contains Nonbinding Recommendations Draft — Not for Implementation 32 GLOSSARY 1292 1293 The following terms are defined for the purposes of this guidance: 1294 1295 Air separation units: Air separation units separate atmospheric air into constituent gases of 1296 oxygen, nitrogen, and argon through a purification process of precleaning, compression, cooling, 1297 and fractional distillation of liquefied air. 1298 1299 Batch: A specific quantity of a medical gas or other material that is intended to have uniform 1300 character and quality, within specified limits, and is produced according to a single 1301 manufacturing order during the same cycle of manufacture.1 1302 1303 Component: Any ingredient intended for use in the manufacture of a medical gas, including 1304 those that may not appear in such gas. It does not include an incoming designated medical gas.2 1305 1306 Designated medical gas (DMG): A drug that is manufactured or stored in a liquefied, 1307 nonliquefied, or cryogenic state; is administered as a gas; and is defined in section 575(1) of the 1308 Federal Food, Drug, and Cosmetic Act (FD&C Act) (21 U.S.C. 360ddd(1)).3 1309 1310 In-process material: Any material fabricated, compounded, blended, or derived by chemical 1311 reaction that is produced for, and used in, the preparation of the medical gas.4 1312 1313 Incoming DMG: A DMG received from one source that, after receipt, is commingled with the 1314 same gas from another source, used in a medically appropriate combination of DMGs or in the 1315 production of another medical gas, or further distributed.5 1316 1317 Lot: A batch, or a specific identified portion of a batch, having uniform character and quality 1318 within specified limits; or, in the case of a medical gas produced by continuous process, it is a 1319 specific identified amount produced in a unit of time or quantity in a manner that assures its 1320 having uniform character and quality within specified limits.6 1321 1322 Lot number, control number, or batch number: Any distinctive combination of letters, 1323 numbers, or symbols, or any combination of them, from which the complete history of the 1324 manufacture, processing, packing, holding, and distribution of a batch or lot of medical gas or 1325 other material can be determined.7 1326 1327 Manufacturer: Any person or firm that manufactures, processes, packs, or holds a medical gas, 1328 which includes packaging and labeling operations, testing, and quality control.8 1329 1 See 21 CFR 213.3(b)(2). 2 See 21 CFR 213.3(b)(4). 3 See 21 CFR 213.3(b)(5). 4 See 21 CFR 213.3(b)(7). 5 See 21 CFR 213.3(b)(8). 6 See 21 CFR 213.3(b)(9). 7 See 21 CFR 213.3(b)(10). 8 This definition is based on the definition of manufacture, processing, packing, or holding of medical gases in 21 CFR 213.3(b)(11).
Contains Nonbinding Recommendations Draft — Not for Implementation 33 1330 Medical gas: A drug that is manufactured or stored in a liquefied, nonliquefied, or cryogenic 1331 state and administered as a gas.9 1332 1333 Original manufacturer: The person that initially produces a DMG by chemical reaction, 1334 physical separation, compression of atmospheric air, purification (e.g., reprocessing an industrial 1335 gas into a medical gas), or other means.10 1336 1337 Portable cryogenic medical gas containers: Containers that are capable of being transported 1338 and are intended to be attached to a medical gas supply system within a hospital, health care 1339 entity, nursing home, other facility, or home health care setting, or is used to fill small cryogenic 1340 gas containers for use by individual patients. The term excludes cryogenic containers that are 1341 not designed to be connected to a medical gas supply system, e.g., tank trucks, trailers, rail cars, 1342 or small cryogenic gas containers for use by individual patients (including portable liquid oxygen 1343 units, as defined in 21 CFR 868.5655).11 1344 1345 Quality unit: Any person or persons designated with the authority and responsibility for overall 1346 quality management and other responsibilities as defined in 21 CFR 213.22.12 1347 1348 Strength: The concentration of the medical gas (for example, weight/weight, weight/volume, or 1349 unit dose/volume basis); and/or the potency, that is, the therapeutic activity of the medical gas as 1350 indicated by appropriate laboratory tests or by adequately developed and controlled clinical data 1351 (expressed, for example, in terms of units by reference to a standard).13 1352 1353 Transfillers: Transfillers manufacture medical gas by transferring a finished medical gas, either 1354 in a liquid or gaseous state, from one container to another, including a container that contains the 1355 same medical gas. Manufacturers that combine different medical gases are considered both 1356 transfillers and original manufacturers. 1357 1358 Uninterrupted filling sequence: A single, continuous filling sequence with no breaks or 1359 shutdowns occurring during the filling operation. This sequence uses the same personnel, 1360 equipment, and batch of component(s). 1361 1362 9 See section 575(2) of the FD&C Act; 21 CFR 213.3(b)(12). 10 See 21 CFR 213.3(b)(13). 11 See 21 CFR 213.94(e)(1); 21 CFR 201.328(a). 12 See 21 CFR 213.3(b)(14). 13 See 21 CFR 213.3(b)(15).