Technical Considerations for Demonstrating Reliability of Emergency-Use Injectors Submitted under a BLA, NDA or ANDA: Draft Guidance for Industry and Food and Drug Administration Staff
us-fda-guidance-6b13be0854fbc7d8ec7ba3d3c9745727Technical Considerations for Demonstrating Reliability of Emergency-Use Injectors Submitted under a BLA, NDA or ANDA: Draft Guidance for Industry and Food and Drug Administration StaffDraft Guidance
Administrative draft guidance, not enacted legislation. This captured publisher edition preserves the listed document and attachments; later developments are not independently consolidated.
- Docket Number:
- FDA-2019-D-5573
- Issued by:
- Guidance Issuing OfficeOffice of the Commissioner, Office of Clinical Policy and Programs, Office of Combination ProductsCenter for Drug Evaluation and ResearchCenter for Biologics Evaluation and ResearchCenter for Devices and Radiological Health
This guidance covers emergency-use injectors submitted under a biologics license application (BLA), new drug application (NDA), or abbreviated new drug application (ANDA). The term “emergency-use injector” means injectors marketed with an emergency-use drug as a prefilled single entity combination product under 21 CFR 3.2(e)(1) or as a co-packaged combination product under 21 CFR 3.2(e)(2). Emergency-use injector includes pen injectors, autoinjectors, or on-body-wearable delivery systems for drugs for emergency treatment of conditions such as anaphylaxis, opioid overdose, poisoning, or severe hypoglycemia.
Submit Comments
You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))
If unable to submit comments online, please mail written comments to:
Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852
All written comments should be identified with this document's docket number: FDA-2019-D-5573.
Questions?
Technical Considerations for Demonstrating Reliability of Emergency-Use Injectors Submitted under a BLA, NDA or ANDA: Guidance for Industry and Food and Drug Administration Staff DRAFT GUIDANCE This guidance document is being distributed for comment purposes only. Although you can comment on any guidance at any time (see 21 CFR 10.115(g)(5)), to ensure that the agency considers your comment on this draft guidance before it begins work on the final version of the guidance, submit written or electronic comments on the draft guidance within 60 days of publication in the Federal Register of the notice announcing the availability of the draft guidance. Submit written comments to the Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852. Submit electronic comments to http://www.regulations.gov. All comments should be identified with the docket number listed in the notice of availability that publishes in the Federal Register. Additional copies are available from: Office of Combination Products Food and Drug Administration WO32, Hub/Mail Room #5129 10903 New Hampshire Avenue Silver Spring, MD 20993 (Tel) 301-796-8930 (Fax) 301-847-8619 https://www.fda.gov/combination-products For questions regarding this document, contact Patricia Love, Office of Combination Products, at 301-796-8930, patricia.love@fda.hhs.gov or combination@fda.gov. U.S. Department of Health and Human Services Food and Drug Administration Center for Devices and Radiological Health Center for Drug Evaluation and Research Center for Biological Evaluation and Research Office of Combination Products in the Office of the Commissioner April 2020
Contains Nonbinding Recommendations Draft – Not for Implementation i Table of Contents I. INTRODUCTION 2 II. SCOPE 3 III. REGULATORY FRAMEWORK 4 IV. BACKGROUND ON RELIABILITY 5 V. RELIABILTY REPORT DEVELOPMENT AND CONTENT RECOMMENDATIONS FOR PREMARKET SUBMISSIONS 9 VI. RELIABILITY REPORT FORMAT CONSIDERATIONS 17 VII. WHERE TO FIND ADDITIONAL INFORMATION 17 VIII. APPENDIX: FAULT TREE EXAMPLE 19
Contains Nonbinding Recommendations Draft – Not for Implementation 2 Technical Considerations for Demonstrating Reliability of 1 Emergency-Use Injectors Submitted under a BLA, NDA, or ANDA 2 Guidance for Industry and Food and Drug Administration Staff 1 3 4 This draft guidance, when finalized, will represent the current thinking of the Food and Drug 5 Administration (FDA or Agency) on this topic. It does not establish any rights for any person 6 and is not binding on FDA or the public. You can use an alternative approach if it satisfies the 7 requirements of the applicable statutes and regulations. To discuss an alternative approach, 8 contact the FDA staff responsible for this guidance as listed on the title page. 9 10 11 I. INTRODUCTION 12 13 This guidance covers emergency-use injectors submitted under a biologics license application 14 (BLA), new drug application (NDA), or abbreviated new drug application (ANDA). The term 15 “emergency-use injector” means injectors marketed with an emergency-use drug2 as a prefilled 16 single entity combination product under 21 CFR 3.2(e)(1) or as a co-packaged combination 17 product under 21 CFR 3.2(e)(2). Emergency-use injector includes pen injectors, autoinjectors, or 18 on-body-wearable delivery systems for drugs for emergency treatment of conditions such as 19 anaphylaxis, opioid overdose, poisoning, or severe hypoglycemia. 20 21 For injectable drug or biological products that are intended to treat emergent, life-threatening 22 conditions, it is essential to ensure that the emergency-use injector will reliably deliver the drug 23 as intended. Failure of the injector may prevent adequate delivery of a life-saving drug to a 24 patient. In this context, reliability is defined as the probability that the injector will perform as 25 intended, without failure, for a given time interval under specified conditions.3 26 27 The FDA guidance document, “Technical Considerations for Pen, Jet, and Related Injectors 28 Intended for Use with Drugs and Biological Products” provides general recommendations for 29 the technical and scientific information to be included in marketing applications for a range of 30 injectors for various uses.4 In that document the reliability of the injector in delivering the drug 31 product is listed as one of the functional elements FDA assesses in premarket review. This 32 1 This guidance has been prepared by the Center for Devices and Radiological Health in cooperation with the Center for Drug Evaluation and Research, the Center for Biologic Evaluation and Research, and the Office of Combination Products in the Office of Medical Products and Tobacco/ Office of Clinical Policy and Programs/ Office of the Commissioner at the Food and Drug Administration. 2 For purposes of this guidance, unless otherwise stated, the term drug applies to human drug and biological products. 3 Reliability definition source: IEC 61078:2016 - Reliability Block Diagrams. Some Agency guidance documents use the term robustness to convey the reliability concept. 4 The FDA guidance Technical Considerations for Pen, Jet, and Related Injectors Intended for Use with Drugs and Biological Products is accessible at https://www.fda.gov/regulatory-information/search-fda-guidance- documents/technical-considerations-pen-jet-and-related-injectors-intended-use-drugs-and-biological-products.
Contains Nonbinding Recommendations Draft – Not for Implementation 3 document describes additional information and data that FDA recommends be included in 33 marketing applications to demonstrate that an emergency-use injector is reliable. 34 35 In general, FDA’s guidance documents do not establish legally enforceable responsibilities. 36 Instead, guidances describe the Agency’s current thinking on a topic and should be viewed only 37 as recommendations, unless specific regulatory or statutory requirements are cited. The use of 38 the word should in Agency guidances means that something is suggested or recommended, but 39 not required. 40 41 II. SCOPE 42 43 This guidance’s focus is emergency-use injectors marketed with the emergency-use drug as a 44 prefilled single entity combination product or as a co-packaged combination product.5 The 45 recommendations in this guidance are applicable to combination products intended to treat 46 emergent, life-threatening conditions, when it is essential to ensure that the injector will reliably 47 deliver the drug as intended.6 Such products are marketed as combination products assigned to 48 the Center for Drug Evaluation and Research (CDER) or the Center for Biological Evaluation 49 and Research (CBER) with market authorization under an approved NDA, ANDA, or BLA.7 50 51 Although this guidance contains specific recommendations of ways to demonstrate the reliability 52 of emergency-use injectors,8 the recommendations would also be useful in considering how to 53 demonstrate the reliability of other emergency-use drug delivery devices; e.g., intranasal sprays, 54 inhalation devices, topical cutaneous sprays, syringes, or transdermal systems. Questions 55 regarding whether these recommendations would apply to a specific emergency-use drug 56 delivery system and proposed methods to demonstrate the reliability should be discussed with 57 the Agency early in the product development process.9 58 59 5 See 21 CFR part 3. Combination products are comprised of differently regulated articles; i.e., a drug-device, device-biological product, drug-biological product, or a combination of a drug, device, and biological product. See 21 CFR 3.2(e)(1) and (2) for definitions of single entity and co-packaged combination products. 6 The reliability data discussed within this guidance document is limited to assessing functional performance of the device and does not address human factors/user interface considerations. For information on human factors see FDA guidance: Human Factors Studies and Related Clinical Study Considerations in Combination Product Design and Development, https://www.fda.gov/regulatory-information/search-fda-guidance-documents/human-factors- studies-and-related-clinical-study-considerations-combination-product-design-and; or FDA guidance Applying Human Factors and Usability Engineering to Medical Devices, https://www.fda.gov/regulatory-information/search- fda-guidance-documents/applying-human-factors-and-usability-engineering-medical-devices. 7 See 21 CFR 3.4. Combination Products are assigned to a lead center based on the primary mode of action (PMOA). In this instance, the drug or biological product is considered to be the PMOA and the combination products are assigned to CDER or CBER. For additional information contact the Office of Combination Products at combination@fda.gov. 8 Throughout this document the term emergency-use injector applies to the device constituent part of the combination product. 9 Applicants developing an emergency-use injector or similar product for an ANDA should request Agency feedback on the potential applicability of the recommendations in this guidance document.
Contains Nonbinding Recommendations Draft – Not for Implementation 4 III. REGULATORY FRAMEWORK 60 61 Combination products are subject to 21 CFR Part 4, which sets forth current good manufacturing 62 practice (CGMP) requirements for combination products. The constituent parts of a combination 63 product retain their regulatory status (as a drug or device, for example) after they are combined. 64 The CGMP requirements that apply to each of the constituent parts apply to the combination 65 product they constitute. 66 67 For single-entity combination products and co-packaged combination products, such as those 68 covered in this guidance, part 4 identifies two ways to demonstrate compliance with CGMP 69 requirements. Under the first option, manufacturers demonstrate compliance with all CGMP 70 regulations applicable to each of the constituent parts included in the combination product. 71 Under the second option, manufacturers implement a streamlined approach for combination 72 products that include both a drug and device by demonstrating compliance with either the drug 73 CGMPs (21 CFR parts 210 and 211) or the device Quality System (QS) regulation (21 CFR part 74 820) and also demonstrating compliance with specified provisions from the other two sets of 75 CGMP requirements. 76 77 Under the streamlined approach described in 21 CFR 4.4(b), manufacturers of drug-led, drug- 78 device combination products,10 such as those that are the subject of this guidance, may meet the 79 requirements of both the drug CGMPs and device QS regulation by designing and implementing 80 a CGMP operating system that demonstrates compliance with the drug CGMPs and the 81 following provisions from the device QS regulation in accordance with 21 CFR 4.4(b)(1) (drug 82 CGMP-based streamlined approach): 83 84 • 21 CFR 820.20 Management responsibility 85 • 21 CFR 820.30 Design controls 86 • 21 CFR 820.50 Purchasing controls 87 • 21 CFR 820.100 Corrective and preventive action 88 • 21 CFR 820.170 Installation 89 • 21 CFR 820.200 Servicing 90 91 As explained in the FDA guidance for “Current Good Manufacturing Practice Requirements for 92 Combination Products, “the core requirements embedded in these regulations provide for 93 systems that assure proper design, monitoring, and control of manufacturing processes and 94 facilities. This includes establishing a strong quality management system, using appropriate 95 quality raw materials, establishing robust manufacturing and control procedures based on sound 96 design principles, and detecting and investigating product quality deviations. In addition, these 97 10 A biological product regulated under section 351 of the Public Health Service Act is also, by definition, a drug or a device. Accordingly, for combination products that include a biological product, in addition to complying with the drug CGMP and device QS regulation requirements as applicable in accordance with 21 CFR part 4, manufacturers of such products must comply with the CGMP requirements in 21 CFR parts 600 through 680 that would apply to the biological product if it were not part of a combination product. 21 CFR 4.4(b)(3).
Contains Nonbinding Recommendations Draft – Not for Implementation 5 regulations call for ongoing assessment of systems and the implementation of corrective actions 98 where appropriate.”11 99 100 IV. BACKGROUND ON RELIABILITY 101 102 Emergency-use injectors such as those for treatment of anaphylaxis typically are used by the 103 patient, caregiver, or first responder outside of a health care environment. For the patient 104 experiencing the emergency or their assisting lay caregivers, there may be only one opportunity 105 to use the product and for that one opportunity the emergency-use injector needs to successfully 106 inject the drug at that time. Further, because these emergency-use injectors are for a single use, 107 the functional performance cannot be verified before the injector is used. Therefore, to ensure 108 safe and effective use of the emergency-use injector, FDA recommends using the reliability 109 engineering methods described in this guidance to ensure that the injector will function as 110 intended within its expiration date. Designing the combination product to achieve its identified 111 functional performance (reliability) is consistent with the combination product good 112 manufacturing practice design control requirements provisions (see 21 CFR 4.4(b)(1)(ii) and 113 4.4(b)(1)(iv)). 114 115 Although the requirements of both the drug CGMPs and device QS regulation must be met, as 116 described in Section III above, several aspects of the design requirements identified in 21 CFR 117 4.4(b)(1)(ii) are particularly important in the development and reliability of an emergency-use 118 injector. For example, 119 120 • 21 CFR 820.30(c) states that “[e]ach manufacturer shall establish and maintain 121 procedures to ensure that the design requirements relating to a device are appropriate and 122 address the intended use of the device, including the needs of the user and patient.” 123 124 • 21 CFR 820.30(d) states that “[e]ach manufacturer shall establish and maintain 125 procedures for defining and documenting design output in terms that allow an adequate 126 evaluation of conformance to design input requirements. Design output procedures shall 127 contain or make reference to acceptance criteria and shall ensure that those design 128 outputs that are essential for the proper functioning of the device are identified.” 129 130 • 21 CFR 820.30(f) states that “[e]ach manufacturer shall establish and maintain 131 procedures for verifying the device design. Design verification shall confirm that the 132 design output meets the design input requirements. The results of the design verification, 133 including identification of the design, method(s), the date, and the individual(s) 134 performing the verification, shall be documented in the DHF.” Moreover, 21 CFR 135 820.30(g) states that “[e]ach manufacturer shall establish and maintain procedures for 136 validating the device design. Design validation shall be performed under defined 137 operating conditions on initial production units, lots, or batches, or their equivalents. 138 Design validation shall ensure that devices conform to defined user needs and intended 139 11 See Section II.B accessible at https://www.fda.gov/regulatory-information/search-fda-guidance- documents/current-good-manufacturing-practice-requirements-combination-products.
Contains Nonbinding Recommendations Draft – Not for Implementation 6 uses and shall include testing of production units under actual or simulated use 140 conditions.” 141 142 • 21 CFR 820.30(h) states that “[e]ach manufacturer shall establish and maintain 143 procedures to ensure that the device design is correctly translated into production 144 specifications.” 145 146 • 21 CFR 820.30(i) states that “[e]ach manufacturer shall establish and maintain 147 procedures for the identification, documentation, validation or where appropriate 148 verification, review, and approval of design changes before their implementation.” 149 150 • 21 CFR 820.100 states that “each manufacturer shall establish and maintain procedures 151 for implementing corrective and preventive action.” 152 153 FDA considers that “the needs of the user and patient” (21 CFR 820.30(c)) in an emergency-use 154 context would be that, for the patient experiencing the emergency or his/her assisting lay 155 caregiver, there is only one opportunity to use the product and, thus, FDA has found emergency- 156 use injectors acceptable if they would successfully inject the drug on the first try. In this 157 instance, the design input requirements would provide functional measures for performance 158 characteristics, specifications for how reliably the emergency-use injector functions, and the use 159 condition of the patient or caregiver. This would include identification of a reliability 160 specification that is consistent with the level of risk to the patient if the emergency-use injector 161 does not function (e.g., the morbidity or mortality associated with untreated anaphylactic shock). 162 163 In addition, FDA interprets the requirements under 21 CFR 820.30(f) and (g) to mean that a 164 manufacturer of an emergency-use injector must verify and validate the design of the injector to 165 ensure that it works in “one opportunity” situations. These requirements can be met, for 166 example, if available documentation demonstrates that (1) the emergency-use injector has met its 167 design input requirements within the specified reliability targets at expiry and (2) design 168 validation has been conducted on finished products to ensure reliability targets are met and that 169 test conditions were representative of how the product would be exposed up to expiry. 170 171 The preceding design control requirements are intended to ensure that the emergency-use 172 injector performance is as reliable as possible. Consistent with this purpose, FDA has found 173 emergency-use injectors to be acceptable if they would successfully inject on the first try in “one 174 opportunity” (or emergency) situations. This reliability concept has an inherent feasibility 175 consideration. FDA recommends that emergency-use injectors include design control 176 specifications for successful injection reliability of 99.999% with a 95% level of confidence As 177 FDA has found such specifications to be acceptable under applicable standards. This prospective 178 99.999% target is equivalent to post-market detection of failure to successfully inject in 179 1/100,000 injection attempts. This reliability level was found to balance appropriately the 180 objective of ensuring the emergency-use injector performance is as safe and reliable as possible 181 with considerations on feasibility. 182 183 As part of determining an acceptable level of reliability, FDA has considered available 184 information for risk assessment. Specifically, the FDA-recognized standard, ISO 14791 - 185
Contains Nonbinding Recommendations Draft – Not for Implementation 7 Application of risk management to medical devices, provides insight regarding probabilities of 186 occurrence.12 In the standard, examples are provided for semi-quantitative analysis that 187 identifies probable, remote, and improbable events rates. In the standard, events occurring in the 188 range of 1/10,000 detection rate are considered to be probable. In FDA review experience, this 189 probable failure rate is likely associated with unacceptable rates of adverse events for 190 emergency-use injectors that may result in product recalls. In the standard, events occurring less 191 than a 1/1,000,000 detection rate are considered to be improbable. Although the lowest possible 192 failure rates are desirable, FDA believes that based on the standard rates and current technology, 193 that the improbable rate of less than 1/1,000,000 detection rate could result in drug shortage or 194 delayed product availability. In contrast to both of the preceding rates, events occurring within 195 1/100,000 to 1/1,000,000 detection rate are considered as a remote probability of occurrence. 196 Therefore, based on the ISO standard, FDA believes the detection of failure to successfully inject 197 in 1/100,000 injection attempts is an appropriate risk management target for ensuring successful 198 injection and treatment when there is only one opportunity to inject. Further, FDA review of 199 recent marketing applications demonstrates that the reliability target of 99.999% with a 95% 200 level of confidence (i.e., 1/100,000 failure to successfully inject rate) is achievable for these 201 emergency-use injectors. 202 203 The following information provides the details of an example of what FDA currently believes 204 would be an acceptable approach for the mathematical model, statistics, fault tree analysis, and 205 use of combination product current good manufacturing design control requirements provisions 206 (21 CFR 4.4(b)(1)(ii) and 4.4(b)(1)(iv)) to establish reliability of the emergency-use injector. 207 FDA recognizes that as an alternative to the approach discussed in this guidance, applicants may 208 propose other reliability specifications methodologies. For example, based on considerations 209 such as product design, drug being delivered, for emergent unmet medical needs, counter- 210 terrorism considerations, or conditions of use, alternative reliability specifications may be 211 appropriate. During emergency-use injector development FDA encourages applicants to seek 212 FDA meetings to discuss their proposals and rationale (e.g., context of use, risk/benefit, shortage, 213 and supportive data). 214 215 Establishing reliability 216 217 Establishing reliability is an iterative process in which the design controls should focus on 218 emergency-use injector attributes determined to be essential for achieving the emergency-use 219 injector’s intended use.13 The level of reliability necessary to manufacture a safe and effective 220 combination product directly correlates to the level of risk associated with an unreliable emergency- 221 use injector. Because reliability as a mathematical model is defined as R(t) = 1 – F(t), where F(t) 222 represents the cumulative distribution function of failure, the goal should be to define the point at 223 which that distribution, F(t), is adequately controlled. The reliability specification(s), R(t), 224 represents the probability that the emergency-use injector will perform as intended, without failure, 225 12 ISO 14971:2007/R(2016) Medical Devices - Application of risk management to medical devices; Section 3.4.2 Semi-quantitative analysis. 13 For more information on design controls for a combination product see 21 CFR Part 4 Subpart A, and related FDA guidance Current Good Manufacturing Practice Requirements for Combination Products accessible at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/current-good-manufacturing-practice- requirements-combination-products.
Contains Nonbinding Recommendations Draft – Not for Implementation 8 for a given time interval under specified conditions. This level of risk should be identified in the 226 risk analysis conducted as part of device design controls activities.14 227 228 This assessment is specific to the combination product’s intended use because the risks are likely 229 to be impacted by the condition being treated, environments of use, emergency-use injector 230 technology, drug, user characteristics, etc. The emergency-use injector reliability analysis, in 231 addition to the traditional development activities, should incorporate the following device design 232 control activities as applicable to the emergency-use injector: 233 234 • Identification of the reliability requirements and specifications; 235 • Risk analysis; 236 • Design verification and validation of the reliability requirements and specifications; and 237 • Design transfer of the reliability specification to the correct production specifications. 238 239 As described in Section IV Background on Reliability, after transfer of the design into 240 production, the manufacturing controls must be adequate to produce reliable emergency-use 241 injectors as specified.15 These controls should, among other required activities, include the 242 following to ensure the design specifications and tolerances that must be achieved during 243 production and that all sources of potential quality problems are analyzed: 244 245 • Adequate manufacturing in-process controls and release activities to ensure the final 246 finished emergency-use injector conforms to its specifications; 247 • Adequately defined acceptance activities for the supplied components to ensure that the 248 manufactured combination product has the required design attributes to ensure the 249 reliability specifications are achieved; and 250 • Establish and maintain procedures for implementing corrective and preventive action 251 (CAPA) activities (e.g., post-market complaints) to ensure that all sources of quality data 252 are analyzed and, where necessary, preventive or corrective actions are taken.16 253 254 In addition to providing assurance of the emergency-use injector reliability, the advantages of a 255 well-constructed reliability analysis include the following life-cycle management benefits in the 256 linkage between the design and manufacturing controls. 257 258 • The least reliable design elements or manufacturing controls of an emergency-use 259 injector can be identified prior to commercialization of the combination product. This 260 can further inform potential improvements in manufacturing controls, tracking or 261 trending limits of part rejects, and future investigations of failed combination products; 262 • Future changes to either design or manufacturing processes can be evaluated against the 263 reliability analysis to assess the potential impact to the emergency-use injector reliability; 264 • Post-market emergency-use injector failure investigations can directly link to the 265 emergency-use injector reliability analysis to readily identify gaps or deficiencies in the 266 emergency-use injector’s design and/or manufacturing controls; and 267 14 See 21 CFR 820.30(g). 15 See 21 CFR 820.30(h). 16 See 21 CFR 820.100.
Contains Nonbinding Recommendations Draft – Not for Implementation 9 • Once the results of those specific design elements or manufacturing controls are 268 understood within the context of the overall reliability analysis, if the reliability analysis 269 is implemented early in the development cycle, reliability improvements can be made by 270 making changes to the emergency-use injector design or manufacturing processes or 271 controls. 272 273 Section V below provides an example of an acceptable way to demonstrate the reliability of the 274 emergency-use injector and define failure to successfully inject. 275 276 V. RELIABILTY REPORT DEVELOPMENT AND CONTENT 277 RECOMMENDATIONS FOR PREMARKET SUBMISSIONS 278 279 As discussed in Section IV, achieving the necessary reliability specification is based on 280 knowledge of the design, manufacture, and use of the combination product. Generally, the 281 highest risk is failure to successfully inject (e.g., activation and drug delivery functions). This 282 occurs when one or more functional failure modes result in the emergency-use injector failure to 283 deploy the needle to the target site or failure to complete drug delivery as intended. As described 284 in Section IV Background on Reliability, FDA has found emergency-use injectors to be 285 acceptable if they would successfully inject in “one opportunity” (or emergency) situations. As 286 stated above, FDA has found emergency-use injectors to be acceptable under the applicable 287 standards when they include design control specifications for successful injection reliability of 288 99.999% with a 95% level of confidence. 289 290 To establish the emergency-use injector’s safe and effective injection performance, the 291 marketing application should include information to verify and validate that the emergency-use 292 injector achieves its reliability specifications and related information. The following sections 293 identify examples of acceptable activities for developing the verification and validation data. If 294 applicants submit such information, FDA recommends that the applicants provide these data in 295 the form of an emergency-use injector reliability report to facilitate efficient review.17 The 296 following subsections provide examples of the type of information to provide in the reliability 297 report. Section VI of this guidance provides a reliability report format example. 298 299 1. Design Inputs and Design Outputs Necessary18 for Ensuring Reliability 300 301 As described in Section IV, specifications for how reliably the emergency-use injector 302 functions and the use condition of the patient or caregiver need to be identified.19 The design 303 inputs necessary for ensuring reliability should be identified and developed into specified 304 design outputs. Selecting design inputs that may not be relevant to the reliable function of 305 the emergency-use injector could result in an inability to meet the manufacturer’s established 306 reliability specifications. To assist in identifying design input requirements, manufacturers 307 should consider the following information: 308 17 Throughout the remainder of this document the term “reliability report” applies to the emergency-use injector reliability report. 18 See 21 CFR 820.30(c) and 820.30(d). 19 See 21 CFR 820.30(c).
Contains Nonbinding Recommendations Draft – Not for Implementation 10 • Intended use and associated risks; 309 • Emergency-use injector risk analysis, including the drug constituent part 310 characteristics; 311 • Use-related issues to the extent that they could impact the reliability of the 312 combination product, including: 313 o Use tasks, which may include unpacking, preparation, administration and disposal 314 of the combination product; and 315 o Use conditions and environments of use. 316 • Use-condition factors20 to consider include all users of the emergency-use injector, 317 where they are using it, and the possible circumstances under which the emergency- 318 use injector may be used, including: 319 o Use environment (e.g., school, work, public transportation vehicle, harsh climates, 320 first responder chaotic conditions) and associated risks to reliability; and 321 o User characteristics (e.g., self- injection with cognitive or physical impairment 322 associated with the disorder being treated) and associated risks to reliability. 323 324 Table -1 provides an example of emergency-use injector design considerations for 325 emergency-use injector reliability only.21 326 327 328 329 (See next page for Table-1.) 330 331 20 The use condition factors discussed within this guidance document are limited to assessing functional performance of the emergency-use injector and do not address human factors/user interface considerations. For information on human factors see FDA guidance: Human Factors Studies and Related Clinical Study Considerations in Combination Product Design and Development (Draft 2016), https://www.fda.gov/regulatory- information/search-fda-guidance-documents/human-factors-studies-and-related-clinical-study-considerations- combination-product-design-and; or Applying Human Factors and Usability Engineering to Medical Devices, https://www.fda.gov/regulatory-information/search-fda-guidance-documents/applying-human-factors-and-usability- engineering-medical-devices. 21 Other types of input or output considerations for the safe and effective use of the emergency-use injector should continue to be part of the combination product development. For more information see FDA guidance Technical Considerations for Pen, Jet, and Related Injectors Intended for Use with Drugs and Biological Products accessible at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/technical-considerations-pen-jet- and-related-injectors-intended-use-drugs-and-biological-products.
Contains Nonbinding Recommendations Draft – Not for Implementation 11 Table-1: Emergency-Use Injector Design Reliability Development Considerations22 Consideration Category Examples Protective Packaging • Packaging ability to prevent emergency-use injector damage during shipping, daily carry, etc. • Removal from packaging or carrying case (e.g., force to remove) Removal / Deactivation of Safety Mechanisms • Force to remove caps or needle shields • Force to deactivate safety mechanism Activation Force • Force to initiate the injection mechanism Needle Insertion • Needle bevel specifications • Needle material of construction • Needle Insertion Forces (e.g., penetrating clothing, skin, etc.) • Needle resistance to bending and fracture Needle Patency • Particulates Injection Depth • Target tissue for drug delivery • Body habitus, skin and tissue characteristics • Anatomical location(s) for injection • Types of garments to be injected through • Exposed needle length Drug Delivery Initiates as Intended • Needle reaches intended injection depth • Drug delivery begins when needle is at intended injection depth • Drug fluid properties (e.g., viscosity) Drug Delivery Stops as Intended • Needle does not retract before intended dose is delivered • Audible, visual, or tactile feedback does not prematurely signal a completed injection Dose Accuracy • Intended dose delivered to intended injection site or depth 332 The preceding tabular considerations are to assist in identifying the emergency-use injector 333 performance characteristics that inform the design inputs and outputs. Based on such 334 considerations the applicant should develop the design inputs and outputs that are essential 335 for ensuring reliability of the applicant’s proposed product. The reliability report should 336 define all the design inputs and outputs determined to be necessary for achieving reliability. 337 It is important to clearly define the difference between acceptable and unacceptable 338 emergency-use injector performance to determine appropriate design inputs that will inform 339 22 For more information on development design controls for a combination product see 21 CFR Part 4 Subpart A, and see related FDA guidance Current Good Manufacturing Practice Requirements for Combination Products accessible at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/current-good- manufacturing-practice-requirements-combination-products.
Contains Nonbinding Recommendations Draft – Not for Implementation 12 the design outputs. If the design outputs are not correctly defined, then the overall reliability 340 analysis may also be inadequate. 341 342 2. Definition of Emergency-Use Injector Reliability Specifications 343 344 345 346 347 348 349 350 351 352 353 354 355 356 357 358 359 360 361 362 363 364 365 366 367 368 369 370 371 372 373 374 As noted in the preceding discussion on design inputs and outputs, there are multiple design specifications that should be considered with respect to the reliability of the emergency-use injector. Using the mathematical expression identified in Section IV, R(t) = 1 – F(t), the reliability specification is defined as the probability distribution, R(t). An example of acceptable reliability specifications is those that are developed in accordance with the risk assessment as described in Section V.3 – Fault Tree Analysis of this document and, through this analysis, linked to the appropriate manufacturing controls to ensure the reliability of the final finished combination product. The specifications may be one or two-sided depending upon the risk associated with the failure (e.g., risk of overdose and under-dose exists for a two-sided dose-accuracy reliability specification). For the reliability specification analysis, failure to inject should be the primary endpoint. This should be the top-level failure mode of the fault tree analysis as described in Section V.3.23 The reliability analysis and testing should include emergency-use injector performance requirements based on the assessment of the design (see Section V.1). In general, FDA recommends that these include dose accuracy, extended needle length, activation force, and injection time be included as part of emergency-use injector reliability.24 However, manufacturers should assess the specific emergency-use injector design to determine if additional performance attributes are considered to be essential for completing a successful injection. For example, a specific design may have a cap that must be removed to initiate injection and the manufacturer may determine that cap removal is essential to completing a successful injection. 3. Fault Tree Analysis The information described in the preceding sections could be used to develop a model of the reliability using fault tree analysis. The fault tree analysis would focus on failure to achieve the reliability specifications. An example of an acceptable analysis is one that also includes additional fault trees to address other emergency-use injector performance requirements determined to be essential for reliability (e.g., dose accuracy, extended needle length). Manufacturers should consider the following for their fault tree analyses: 375 376 • Design and manufacturing elements should be considered for the fault tree analysis 377 for the purposes of establishing the reliability of the emergency-use injector to 378 perform as intended, without failure, for a given time interval under specified 379 conditions; 380 23 For more information of fault tree analyses see IEC 61025:2006 – Fault Tree Analysis (FTA); NASA Fault Tree Handbook with Aerospace Applications, 2006. 24 The acceptance criteria for the performance attributes in Table-2 should be established based on relevance to clinical performance (i.e., established as design inputs), and not based on manufacturing capability or to facilitate meeting the reliability target.
Contains Nonbinding Recommendations Draft – Not for Implementation 13 • The probability data for each basic event25 should be included in the fault tree; 381 • The analysis should consider potential common cause failures and whether 382 assumptions of independence of events are supportable; 383 • Any risk analyses (e.g., such as design and process failure modes effects analysis)26 384 used to support the fault tree analysis should be included in the reliability report; 385 • Once the fault tree analyses are completely developed, data should be provided to 386 support that the reliability specification for the top-level failure mode is verified and 387 validated (e.g., the probability data for each basic event should be supported with 388 evidence); and, 389 • The basic events in the fault tree analysis should be linked to appropriate design 390 and/or manufacturing controls. 391 392 393 394 395 396 397 398 399 400 401 402 403 404 405 406 407 408 409 410 Based on standard fault tree analysis quantification methods, the reliability of each basic event within the fault tree analysis should be assessed through a cumulative analysis to determine whether the reliability specification for the top-level failure mode is adequately supported. The statistical methods utilized to demonstrate the reliability of each basic event within the fault tree analysis should inform the test sample size necessary for reliability testing of the final finished combination product. To assess the potential for the basic event failure mode of the emergency-use injector, it is important to use a statistical tolerance interval27 method in which the limits of each individual component are analyzed (e.g., dimensions, geometry, material strength, etc.),28 both by itself and in conjunction with its associated components (i.e., stack-up analysis). To effectively use the tolerance interval method, the critical measurable elements of each component contributing to the basic event should be clearly stated and the statistical tolerance limit29 identified. Data to support the tolerance interval methodology should be provided and may include process validation data for individual components. The resultant k factor30 for each basic event should be used to calculate the necessary sample size of the reliability study based on the desired reliability specification and confidence interval. An example of acceptable use of the statistical tolerance interval methodology is one described above.31 An acceptable methodology should evaluate both design and 411 25 The basic event is a failure mode event or state that cannot be further developed; i.e., the lowest level failure mode that cannot be further subdivided. 26 ISO 14971 Second edition 2007-Medical devices - Application of risk management to medical devices; and IEC 60812:2006 - Analysis techniques for system reliability - Procedure for failure mode and effects analysis (FMEA). (FDA recognized standards.) 27 Statistical tolerance interval is an interval determined from a random sample in such a way that one may have a specified level of confidence that the interval covers at least a specified proportion of the sampled population per ISO 3534-1:2006. (FDA recognized standards.) 28 ISO 16269-6 Second edition 2014 – Statistical interpretation of data - Determination of statistical tolerance intervals. (FDA recognized standards.) 29 Statistical tolerance limit is the statistic representing an end-point of a statistic tolerance interval per ISO 3534- 1:2006. (FDA recognized standards.) 30 The k factor is the variable used to determine the limits of a statistical tolerance interval per ISO 16269-6:2014. 31 There are other methods that may also be appropriate to support the fault tree analysis and overall reliability specifications. If a manufacturer intends to use a method not described in this guidance document, the FDA
Contains Nonbinding Recommendations Draft – Not for Implementation 14 manufacturing risks associated with the emergency-use injector such that the fault tree 412 analysis is assessing reliability of the final, finished combination product. (See Section VIII - 413 Appendix for a fault tree analysis template example.) 414 415 4. Reliability Testing 416 417 FDA recommends that a reliability analysis include verifying and validating the adequacy of 418 the data used to support the reliability specifications, and that the data support the reliability 419 specification over the combination product expiry period. The following subsections 420 describe information recommendations for preconditioning and testing reliability samples. 421 422 a. Use Conditions and Preconditioning Recommendations 423 424 To identify the reliability test conditions to ensure achievement of the reliability 425 specification at the end of use-life,32 it is important to define the combination product’s 426 use conditions and preconditioning steps. This ensures that the testing program has 427 adequately challenged the ability of the emergency-use injector to withstand stressors that 428 are likely to occur or to which the product will be exposed during the use-life. 429 Combination product-specific use-life factors to consider for preconditioning steps that 430 influence shelf-life may include the following preconditions: 431 432 • Shipping; 433 • Aging; 434 • Storage orientation and conditions; 435 • Vibration; 436 • Shock (e.g., resistance to impacts, such as being dropped); and 437 • Environmental factors: 438 o Temperature (extremes and cyclic); 439 o Altitude and pressure effects (e.g., airplane, submarine, or other above/below 440 sea-level effects); and 441 o Air particulates (dust/sand). 442 443 444 FDA recommends that the reliability report include the following information to help establish the acceptability of the use of accelerated testing to generate supporting data: 445 446 • Validation data to ensure that the accelerated methods accurately model the time- 447 dependent failure mechanisms of the emergency-use injector; 448 • Results of testing at various time points to identify any trends in emergency-use 449 injector performance; and 450 encourages the manufacturer to request a meeting to discuss the validity of the proposed approach for supporting the reliability specification. 32 Use-life begins when the product manufacture is complete and ends on the date when the product cannot function as intended (the expiration date). For the combination product, the expiration date also may be described as shelf life or end of use-life of the drug-device combination product.
Contains Nonbinding Recommendations Draft – Not for Implementation 15 • A statistical justification to ensure that the sample size at the final time point is 451 adequate to support the pre-specified reliability and confidence interval goals. 452 FDA recommends that the tolerance interval method discussed in Section V.3 be 453 used to provide the sample size justification. 454 455 The report should also describe the use conditions that are important for the emergency- 456 use injector’s reliability and should define the bounded specifications for each condition 457 (e.g., a temperature range). 458 459 b. Testing 460 461 Reliability verification testing should be conducted on the final finished combination 462 product after considering the appropriate preconditioning and use conditions laid out in 463 Section IV.4.a. A reliability report should include the test protocol with validated test 464 methods. 465 466 The sample size for reliability verification testing should be justified with an appropriate 467 statistical method as discussed in Section V.3 and based on the emergency-use injector 468 reliability specifications described in Section V.2. If multiple test groups are included in 469 the manufacturer’s protocol, such as aged and non-aged test groups, then the sample size 470 should ensure adequate statistical results from each group. 471 472 All test failures should undergo root cause analyses which directly link to the fault tree 473 analysis. The fault tree analysis may need to be updated based on any previously 474 unknown failure modes discovered during reliability verification testing. If there is a 475 failure, the reliability testing may need to be redone depending on the conclusions of the 476 root cause analyses. 477 478 5. Total Product Life Cycle Reliability 479 480 Throughout the life cycle of the combination product, manufacturers may become aware 481 of potential emergency-use injector malfunctions, nonconformance or other related 482 quality problems. In these cases, manufacturers must investigate the potential cause and 483 identify any actions that may be needed to correct and prevent recurrence.33 484 485 An example of an acceptable reliability report is one that documents the manufacturer’s 486 plan for maintaining emergency-use injector reliability throughout the product life cycle 487 as part of compliance with 21 CFR part 4 current good manufacturing practice 488 requirements for combination products.34 An example of an acceptable plan is a plan that 489 includes the following: 490 491 33 21 CFR 820.100. 34 See 21 CFR Part 4 Subpart A, and related FDA guidance Current Good Manufacturing Practice Requirements for Combination Products accessible at https://www.fda.gov/regulatory-information/search-fda-guidance- documents/current-good-manufacturing-practice-requirements-combination-products.
Contains Nonbinding Recommendations Draft – Not for Implementation 16 • Procedures that include requirements for analyzing processes, work operations, 492 concessions, quality audit reports, quality records, service records, complaints, 493 returned combination product, and other sources of quality data to identify 494 existing and potential causes of nonconformance, or other quality problems;35 495 496 • As part of defect and/or failure investigations, the procedures use the reliability 497 data and fault tree analysis as part of the root cause analysis; 498 499 • Procedures for when, during complaint investigations and related CAPA 500 activities, it is appropriate to image the emergency-use injector internally or 501 physically open the emergency-use injector to inspect, measure, and test 502 assemblies or individual components and compare results with the specifications 503 and data identified in the reliability analysis; 504 505 • Appropriate steps for linking the reliability data to the appropriate acceptance 506 activities, including the specific emergency-use injector attributes that are 507 evaluated, evaluation methods, and acceptability criteria that should be considered 508 in the context of the emergency-use injector’s reliability; 509 510 • Detailed descriptions of the in-process control and release test sampling plans to 511 ensure that the reliability specification is maintained for each released lot; 512 513 • Established action limits for significant increases in rejections of the emergency- 514 use injector and its components due to incoming inspection, in-process control, or 515 release test failures; 516 517 • The activities triggered by exceeding an action limit should include 1) the need 518 for implementing CAPA, 2) a root cause investigation, and 3) an associated risk 519 analysis of the failure. The reliability data and fault tree analysis are consulted as 520 part of the root cause investigation; 521 522 • Procedures for updating the reliability data when new information is obtained 523 (e.g., previously unidentified failure modes); and 524 525 • A section that addresses disposition of non-released emergency-use injector or 526 combination product pending analysis and mitigation of newly identified failures. 527 528 6. Activities when Implementing Emergency-Use Injector Design or Manufacturing 529 Modifications 530 531 An example of an acceptable reliability report is one that includes a plan for the activities 532 to be completed when implementing an emergency-use injector design or manufacturing 533 modification. After completing the initial reliability report, future design changes and 534 35 21 CFR 820.100.
Contains Nonbinding Recommendations Draft – Not for Implementation 17 manufacturing process changes may occur that necessitate re-evaluation of reliability to 535 ensure emergency-use injector reliability.36 To determine when the reliability data may 536 need to be updated, the impact of the change should be evaluated based on the existing 537 reliability model. 538 539 Manufacturers should consider if the change impacts the design output specifications of 540 the emergency-use injector or the basic events of the fault tree analysis. If it is 541 determined that the change does impact these aspects, creates new design outputs, or 542 creates new risks, then the emergency-use injector reliability report should be updated 543 with data regarding the changes and included within your premarket submission. 544 545 VI. RELIABILITY REPORT FORMAT CONSIDERATIONS 546 547 To facilitate the data assessment, the reliability report should be provided in the following 548 format: 549 550 Section Content 1. Combination Product Definition • Drug Type • Indications for Use • Emergency-Use Injector Technical and Functional Description • Design Inputs and Design Outputs Necessary for Reliability 2. Emergency-Use Injector Reliability Specifications 3. Fault Tree Analysis 4. Reliability Test Plan and Data 5. Total Product Life Cycle Reliability Plan 7. Conclusions 8. Appendices Containing Supporting Data Reports or Risk Analyses 551 When submitting this information to an NDA or BLA, we recommend including the information 552 with other device constituent part information located in eCTD module 3.2.P.7.37 553 554 VII. WHERE TO FIND ADDITIONAL INFORMATION 555 556 As noted in Section II, the drug-delivery emergency-use injectors addressed by this guidance are 557 designed as combination products with a CDER or CBER lead. These combination products are 558 submitted under an IND, NDA, ANDA, or BLA pathway (including supplements). To address 559 any uncertainties for the emergency-use injector reliability development process, FDA strongly 560 encourages early development meetings (e.g., Pre-IND or Pre-ANDA to discuss the initial 561 combination product) as well as subsequent IND meetings throughout the development process. 562 36 See 21 CFR 820.30(i) and 820.70(b). 37 See section-5 in the FDA eCTD Technical Conformance Guide: Technical Specifications Document: Guidance for Industry Providing Regulatory Submissions in Electronic Format -Certain Human Pharmaceutical Product Applications and Related Submissions Using the eCTD Specifications, December 2019 accessible at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/ectd-technical-conformance-guide.
Contains Nonbinding Recommendations Draft – Not for Implementation 18 These meetings can include requests for clarification on the reliability information discussed in 563 this document. For applicants with emergency-use delivery systems that are not emergency-use 564 injectors, these meetings could be used for early consideration of the principles identified in this 565 document. All meeting requests should be submitted to the lead center in accordance with its 566 procedures and should identify the requested participants (e.g., the lead center, CDRH, Office of 567 Combination Products). FDA intends to use its Inter-Center Consult Review Process38 for the 568 assessment of the scientific and technical questions described in this document. The following 569 guidance documents provide procedural information on requesting meetings with FDA and 570 general information on combination products. 571 572 • Guidance for Industry - Formal Meetings Between the FDA and Sponsors or Applicants 573 of PDUFA Products; https://www.fda.gov/regulatory-information/search-fda-guidance- 574 documents/formal-meetings-between-fda-and-sponsors-or-applicants-pdufa-products- 575 guidance-industry 576 577 • Formal Meetings Between FDA and ANDA Applicants of Complex Products Under 578 GDUFA – Guidance for Industry (DRAFT); https://www.fda.gov/regulatory- 579 information/search-fda-guidance-documents/formal-meetings-between-fda-and-anda- 580 applicants-complex-products-under-gdufa-guidance-industry 581 582 • Guidance for Industry and FDA Staff - Technical Considerations for Pen, Jet, and 583 Related Injectors Intended for Use with Drugs and Biological Products; 584 https://www.fda.gov/regulatory-information/search-fda-guidance-documents/technical- 585 considerations-pen-jet-and-related-injectors-intended-use-drugs-and-biological-products 586 587 • eCTD Technical Conformance Guide: Technical Specifications Document: “Guidance 588 for Industry Providing Regulatory Submissions in Electronic Format —Certain Human 589 Pharmaceutical Product Applications and Related Submissions Using the eCTD 590 Specifications” December 2019 accessible at https://www.fda.gov/regulatory- 591 information/search-fda-guidance-documents/ectd-technical-conformance-guide 592 593 • For general information on combination products see 594 https://www.fda.gov/CombinationProducts/default.htm 595 38 For more information on the Inter-Center Consult Review process see https://www.fda.gov/media/81927/download
Contains Nonbinding Recommendations Draft – Not for Implementation 19 VIII. APPENDIX: FAULT TREE EXAMPLE 596 597 The following is a template example of a fault tree analysis for an emergency-use injector. The 598 example uses the following key terms.39 599 600 • Basic Event: A failure mode event or state that cannot be further developed. 601 602 • Failure to successfully inject: Failure of an emergency-use injector to successfully inject 603 occurs when one or more functional failure modes result in the emergency-use injector 604 failing to deploy the needle or failing to complete drug delivery when or as intended. This 605 could include circumstances where the emergency-use injector prematurely activates or 606 does not activate when intended. 607 608 • Failure Mode: The manner in which a failure occurs. 609 610 Fault Tree Analysis: 611 612 For the emergency-use injector, each final element of the fault tree (e.g. A.1.i.a) should directly 613 link to probability data supporting the overall reliability of the product. The steps for fault tree 614 analysis include the following. See the preceding sections of this document for guidance for 615 information on the completion of these steps, conduct of the analysis, and data submission. 616 617 • Definition of the scope of the analysis; 618 • Familiarization with the design, functions and operation of the system; 619 • Definition of the top event; 620 • Construction of the fault tree; 621 • Analysis of the fault tree logic; and, 622 • Reporting on results of the analysis; 623 624 The template defines the top event as the Failure to Successfully Inject. The fault tree should be 625 broken down into all reasonable, identified faults and failure modes that could lead to a Failure- 626 to-Inject event. The lowest level in the example fault tree illustrates how individual components 627 are incorporated into the fault tree. 628 629 This fault tree example is not prescriptive. The actual number of levels and total amount of 630 failure modes identified in a manufacturer’s fault tree will depend on the specific design and 631 manufacturing of that emergency-use injector. 632 633 634 (See next page for example of flow diagram.) 635 636 39 For a more complete glossary and more information see ISO 61025:2006 – Fault tree analysis and the NASA Fault Tree Handbook with Aerospace Applications, 2006.
Contains Nonbinding Recommendations Draft – Not for Implementation 20 637