IN RE: Bard Implanted Port Catheter Products Liability Litigation

2:23-md-03081Azd24 de fev. de 2026

Abrir fonte

Texto completo

1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
WO
IN THE UNITED STATES DISTRICT COURT
FOR THE DISTRICT OF ARIZONA
IN RE: Bard Implanted Port Catheter
Products Liability Litigation MDL No. 3081
ORDER
Steven MacLean, Ph.D., P.E.
This multidistrict litigation (“MDL”) involves thousands of personal injury cases
related to implanted port catheters (“IPCs” or “ports”) designed, manufactured, and
marketed by Defendants Becton Dickenson and Company, C. R. Bard, Inc., Bard Access
Systems, Inc., and Bard Peripheral Vascular, Inc. (collectively, “Defendants” or “Bard”).
Plaintiffs have filed a motion to exclude the expert opinions of Steven MacLean, Ph.D.,
P.E. Doc. 5203. The motion is fully briefed and no party requests oral argument. See
Docs. 5203, 5757, 6024. The Court will deny the motion.
I. Background.
An IPC is a medical device surgically implanted under the skin to provide repeated
access to a patient’s vascular system without the need for multiple needlesticks into
peripheral veins. IPCs are used to deliver chemotherapy and immunotherapy medication,
intravenous fluids, blood transfusions, and parenteral nutrition. IPCs can remain in place
for weeks, months, or even years until treatment is completed.
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 1 of 22
2
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
IPCs consist of a port and a catheter. The port is a small reservoir disc with a self-
sealing silicone center where a needle can be inserted to deliver medication. The port
typically is implanted beneath the chest skin and below the collarbone. Port bodies vary
in shape and size and are made of plastic, silicone, or titanium. The port is attached to a
catheter, which is a thin flexible tube for delivery of the medication to a vein. The catheter
is tunneled under the skin to an insertion point in a central vein. Catheters vary in width
and are made of silicone or polyurethane. IPCs typically are implanted by an interventional
radiologist or a vascular surgeon. A nurse or other medical professional then accesses the
IPC to inject the medication, which flows from the port to the catheter and into the vein.
This MDL involves multiple versions of Bard IPCs, each of which received
premarket clearance from the Food and Drug Administration (“FDA”). Bard IPCs contain
barium sulfate, a radiopaque substance that can be seen on diagnostic imaging such as
X-ray, CT, and MRI scans. Several Bard IPCs are manufactured using polyoxymethylene
(“POM”), an acetal thermoplastic polymer, in the construction of the port. Many Bard
IPCs have a port with raised palpation bumps on the silicone center, which help identify
these IPCs as power-injectable devices.
Each Plaintiff in this MDL had a Bard IPC implanted in his or her body, and claims
it was defective and caused serious injury. Plaintiffs allege that the barium sulfate in the
IPC degrades the mechanical integrity of the catheter, leading to surface irregularities that
cause catheter fracture, infection, and blood clots. Plaintiffs further allege that Bard’s
manufacturing process for the POM-containing IPCs lacks adequate measures to stabilize
the POM to prevent oxidative degradation, which causes cracks, fissures, and other
physical defects in the polymer and increases the risk of fracture, infection, and thrombosis.
Plaintiffs also allege that the palpation bumps on the power-injectable IPCs cause undue
compression stress on the tissue surrounding the port, leading to ulceration and tissue
necrosis. See Docs. 23, 1889.
Plaintiffs claim that safer alternative designs and manufacturing processes were
available to Bard and that Bard IPCs are more dangerous than other available treatment
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 2 of 22
3
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
options. Plaintiffs assert a host of state law claims, including design and manufacturing
defects, failure to warn, breach of warranty, misrepresentation, concealment, and consumer
fraud and unfair trade practices. See Docs. 23, 1889.
Bard disputes Plaintiffs’ allegations and contends its IPCs are safe and effective and
that the medical community is aware of risks associated with IPCs. Bard further contends
that various factors impact the performance and wear of IPCs, including where the catheter
is inserted and how well the device is maintained by medical professionals and the patient.
See Docs. 23, 2023-1.
The parties plan to use various expert witnesses at trial, including materials science
experts. Defendants have identified Steven MacLean, Ph.D., P.E. as a materials science
expert. MacLean is a Principal Engineer and the Practice Director of the Polymers and
Chemistry Practice at Exponent, a technical consulting firm. Doc. 5203-1 at 17. He holds
a Doctor of Philosophy degree in Materials Science, a Master of Science degree in
Materials Science and Engineering, and a Master of Engineering degree in Mechanical
Engineering. Id. Through his advanced education, MacLean’s chosen field of study was
polymer science and engineering, and he has since spent a 30-year career in this field. Id.
at 17, 21. He is a registered Professional Engineer in several states and a Senior Member
of the Society of Plastics Engineers, among other professional memberships. Id. Plaintiffs
move to exclude his opinions under Federal Rule of Evidence 702.
II. Legal Standard.
Rule 702 provides that “[a] witness who is qualified as an expert by knowledge,
skill, experience, training, or education may testify in the form of an opinion” if the
proponent of the expert’s testimony “demonstrates to the court that it is more likely than
not that” (a) “the expert’s scientific, technical, or other specialized knowledge will help the
trier of fact to understand the evidence or to determine a fact in issue,” (b) “the testimony
is based on sufficient facts or data,” (c) “the testimony is the product of reliable principles
and methods,” and (d) “the expert’s opinion reflects a reliable application of the principles
and methods to the facts of the case.” Fed. R. Evid. 702(a)-(d). The proponent of the
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 3 of 22
4
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
expert testimony must show by a preponderance of the evidence that the testimony satisfies
each of these requirements. Fed. R. Evid. 104(a); Fed. R. Evid. 702 advisory committee’s
note to 2023 amendment (“[T]he rule has been amended to clarify and emphasize that
expert testimony may not be admitted unless the proponent demonstrates to the court that
it is more likely than not that the proffered testimony meets the admissibility requirements
set forth in the rule.”); Engilis v. Monsanto Co., 151 F.4th 1040, 1050 (9th Cir. 2025)
(“Consistent with the 2023 amendment, our precedent establishes that Rule 702 requires a
proponent of expert testimony to demonstrate each of the requirements of Rule 702 by a
preponderance of the evidence.”).
The Court acts as a gatekeeper to ensure that expert testimony satisfies Rule 702.
Daubert v. Merrell Dow Pharms., Inc., 509 U.S. 579, 597 (1993). The Court’s task is not
to decide whether the expert is right or wrong, but to ensure the proposed opinions satisfy
Rule 702. Alaska Rent-A-Car, Inc. v. Avis Budget Grp., 738 F.3d 960, 969-70 (9th Cir.
2013); Engilis, 151 F.4th at 1050. If the proponent does not meet its Rule 702 burden, the
expert testimony is not admissible. See Davis v. McKesson Corp., No. CV-18-1157-PHX-
DGC, 2019 WL 3532179, at *3 n.2 (D. Ariz. Aug. 2, 2019) (“[T]he Court may admit expert
opinions only if it can determine, under Rule 104(a), that Plaintiffs have shown each of the
Rule 702 requirements to be satisfied by a preponderance of the evidence.”).
III. Discussion.
Plaintiffs do not dispute that MacLean is qualified under Rule 702 to render his
opinions. They challenge the methodology and factual foundation underlying several of
his opinions.
A. Non-Testing Opinions.
Plaintiffs argue MacLean’s “non-testing” opinions should be excluded because
Defendants cannot show they were the product of a reliable methodology. Doc. 5203 at 4.
They rely primarily on MacLean’s use of assistants during report preparation to contend
his methodology is unreliable, but they also challenge MacLean’s failure to describe his
methodology.
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 4 of 22
5
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
1. Use of Assistants.
It is unclear what Plaintiffs mean by “non-testing” opinions. They explain that
“‘[n]on-testing opinions’ refers to all opinions not based on Exponent’s SEM,
Fouriertransform infrared spectroscopy (‘FTIR’), differential scanning calorimetry
(‘DSC’), or thermogravimetric analysis (‘TGA’) testing,” and they cite 215 pages of
MacLean’s 233-page report. Id. at 4 n.2. Given this broad-stroke approach to identifying
the challenged opinions, the Court will rule on the admissibility of MacLean’s general
opinions outside of those requiring scientific testing.
The Court will not exclude MacLean’s non-testing opinions based on his use of
assistants to prepare his report. “The fact that [an expert’s] opinions are based on data
collected by others is immaterial; Federal Rule of Evidence 703 expressly allows such
opinion testimony.” Southland Sod Farms v. Stover Seed Co., 108 F.3d 1134, 1142 (9th
Cir. 1997); see also Kim v. Benihana, Inc., No. 22-55529, 2023 WL 3674673, at *1 (9th
Cir. May 26, 2023) (“In arriving at an opinion, an expert may rely on . . . reliable data
collected by others[.]” (citation modified)). “Additionally, an expert may collaborate and
converse with colleagues before reaching his or her opinions.” Wolkowitz v. Lerner, No.
SA CV 07-777-CAS, 2008 WL 1885770, at *4 (C.D. Cal. Apr. 21, 2008) (citing Scott v.
Ross, 140 F.3d 1275, 1286 (9th Cir.1998); United States v. Golden, 532 F.2d 1244, 1248
(9th Cir. 1976)). “An expert may [also] use assistants in performing his work, so long as
those assistants do not ‘exercise professional judgment that is beyond the expert’s ken.’”
Id. (quoting Dura Automotive Sys. v. CTS Corp., 285 F.3d 609 (7th Cir. 2002)). But the
expert must “substantially participate” in the report’s preparation. Arrowood Indem. Co.
v. City of W. Sacramento, No. 221CV00397WBSJDP, 2021 WL 5989746, at *2 (E.D. Cal.
Dec. 17, 2021) (citation modified).
Thus, MacLean was free to reach his opinions and prepare his report in reliance on
data collected by others, in collaboration with his colleagues, and with support from
assistants. It is not dispositive that “21 other employees” participated in the report
preparation. Doc. 5203 at 5 (emphasis omitted). Plaintiffs note that MacLean billed only
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 5 of 22
6
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
14% of the 1,800 hours charged by his firm for the report (id.), but this amounts to 252
hours – more than 31 eight-hour days. This is equivalent to six weeks of full-time work
on his report and opinions for this case.
The Court also notes that MacLean’s collaboration with his colleagues is aligned
with his employer’s internal peer-review process. Doc. 5757 at 6. Exponent’s quality
management system mandates that work product be reviewed by peers, including review
of relied-upon documents. Doc. 5757-1 at 3 (“Q. What’s the [quality control] process? A.
So when we generate work product, we have a formal process where peers have to look at
documents after the initial author to confirm that there is accuracy and quality of what was
written.”). It was part of Exponent’s peer review process for MacLean’s colleagues and
assistants to also review and synthesize data relied on by him. Id. at 23 (“If I am going to
have a cited Bard document in my report, I need to have at least two sets of eyeballs that
have also read it, distilled it, synthesized it. So there would have been overlap.”).
There is also evidence MacLean substantially participated in preparing his report.
MacLean testified that he instructed his team to “review documents [and] synthesize them,”
followed by “lots of team discussions on what information people were finding, [and what]
data they were finding.” Doc. 5203-5 at 7; see also Doc. 5757-1 at 24. MacLean also
participated in the review and synthesizing process. Doc. 5757-1 at 12, 23-24, 26. He had
“complete editorial control” over the report. Id. at 27. He reviewed the report thoroughly
before signing it. For example, he reviewed “every single citation that’s in the report[.]”
Doc. 5203-5 at 8. The Court is satisfied that MacLean substantially participated in the
preparation of his report.
The Court finds no evidence that MacLean’s assistants impermissibly “exercise[d]
professional judgment beyond” his ken. Wolkowitz, 2008 WL 1885770, at *4. Plaintiffs
argue MacLean “cannot explain what methodological choices generated his report”
because he acknowledged he did not know everyone on his team or what particular work
they conducted. Doc. 5203 at 5-6; see, e.g., Doc. 5203-5 at 35-37. But in his firm’s
invoices, team members described tasks well within MacLean’s expertise such as
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 6 of 22
7
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
“document review,” “technical review,” and “internal technical discussions.” Doc. 5203-
6 at 36-37. These are the kinds of information gathering and synthesizing for which
MacLean can rely on assistants. See Kim, 2023 WL 3674673, at *1 (“[A]n expert may rely
on . . . reliable data collected by others[.]” (citation modified)).
Plaintiffs contend MacLean’s assistants impermissibly engaged in “core analytical
work” by “organizing, synthesizing, and interpreting the scientific literature, internal
documents, and testing data.” Doc. 5203 at 7. But Rule 703 instructs that “[a]n expert
may base an opinion on facts or data in the case that the expert has been made aware of[.]”
See Fed. R. Evid. 703. It follows that an expert can be made aware of facts or data through
the work of assistants. See Walsh v. Reliance Tr. Co., No. CV-19-03178-PHX-ROS, 2023
WL 1966921, at *11 (D. Ariz. Feb. 13, 2023) (“Defendants fail to explain why it was
improper for Wert to rely on the work of associates in crafting his opinions. Given the
requirements of Rule 703, Wert was free to rely on reliable facts or data provided to him
by others.”).
Plaintiffs challenge the roles of certain assistants. They argue that because Dr.
Kreder, as project manager, “assigned tasks, managed the workflow, and oversaw the
team,” then MacLean cannot explain his methodology. Doc. 5203 at 5. As project
manager, Kreder was responsible for “the division of labor” and “handle[d] the work flow.”
Doc. 5203-5 at 35. MacLean explained Kreder helped write the draft report, assisted in
document review, reviewed depositions, and read expert reports. Doc. 5757-1 at 12, 14,
23-24, 26-27. But MacLean also conducted these tasks. Id. at 12, 23-24, 26. He testified
that instructions for assistants came from him, not Kreder. Doc. 5203-5 at 9-10 (“[Q.] And
was [Dr. Kreder] also responsible for providing instructions in the review? A. No. Those
instruction would have come from me.”). That Kreder assisted in report preparation as the
project manager does not detract from MacLean’s methodology.
Plaintiffs also challenge the role of Dr. Favreau. They cite In re Zantac (Ranitidine)
Prods. Liab. Litig., 644 F. Supp. 3d 1075 (S.D. Fla. 2022), for the proposition that the
Court should exclude opinions where the “testifying expert delegated methodological
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 7 of 22
8
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
decisions to the unarticulated ‘judgment and discretion of individual [Ph.D.] analysts.’”
Doc. 5203 at 7 (quoting In re Zantac, 644 F. Supp. 3d at 1125-26). They refer to MacLean
delegating literature review to Favreau but not knowing what keywords she used to conduct
her review. Doc. 5203-5 at 29. Keyword searching is not the kind of methodological
decision contemplated by In re Zantac, which scrutinized the use of judgment in a
laboratory technique. See 644 F. Supp. 3d at 1125 (finding an unreliable methodology
where “[a]nalysts were allowed to use their judgment when manually integrating the results
of the chromatography that the laboratory conducted for th[e] litigation”).
Plaintiffs next argue that Dr. Holyoak, who drafted sections of the report addressing
FDA 510(k) and ISO standards, impermissibly exercised judgment outside of MacLean’s
expertise. Plaintiffs contend – without explanation – that Holyoak has expertise in these
areas, “in stark contrast to Dr. MacLean[.]” Docs. 5203 at 8, 5203-5 at 19-20. But as noted
by Defendants, MacLean testified that he and Holyoak “have similar knowledge” in those
areas. Docs. 5203-5 at 22, 5757 at 8 n.9. A review of the doctors’ resumes confirms this.
See Doc. 5757-2 at 5, 12. MacLean and Holyoak hold Ph.D.s in Materials Science and
Biomedical Engineering, respectively. Doc. 5757-2 at 6, 12. Although Holyoak
specializes in regulatory compliance and has experience applying ISO standards (id. at 12),
MacLean has experience in product compliance assessments and in assessing risk during
the product qualification stage (id. at 5-6). He is “well versed in product recall
investigations” across numerous federal agencies and is “familiar with testing and material
standards[.]” Id. at 5. MacLean also testified that Holyoak’s role in the report was limited.
Doc. 5203-5 at 21 (“I would give [Holyoak] some sort of starting point. And he might say,
. . . ‘Here’s a comment for you to think about. Let’s discuss.’ I mean, that’s how the QC
process works.”). The Court is satisfied Holyoak’s contributions to the report were not
beyond MacLean’s ken. See Wolkowitz, 2008 WL 1885770, at *4.
Plaintiffs next challenge the reliability of MacLean’s literature and document
reviews. They argue “there is simply no way for the Court to determine whether the
literature cited in Dr. MacLean’s report was gathered and evaluated in a systematic,
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 8 of 22
9
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
scientifically valid manner” (Doc. 5203 at 8), presumably given his reliance on assistants
to conduct portions of the literature review. But MacLean testified that he reviewed every
source cited in the report. Doc. 5203-5 at 8 (“[E]very single citation that’s in the report I
would have reviewed.”). He also testified that his assistants shared his methodology for
the document review process. Id. at 7 (“Q. So what was the methodology used by whoever
went through these documents to decide what documents to use in the expert report? A.
It’s the same process. You review, synthesize. Q. What were the instructions given by you?
A. Just that, review documents, synthesize them. We had lots of team discussions on what
information people were finding, data they were finding.”).
Plaintiffs argue MacLean’s document review is methodologically flawed because
neither he nor his assistants “had access to (and thus could not search) the two-million
some documents produced in discovery,” and that he instead reviewed “a carefully curated
subset of documents.” Doc. 5203 at 8. But Plaintiffs – who themselves provided their
experts with selected Bard documents – do not get to decide what documents Defendants’
experts review. Rule 702 does not require an expert to review every document produced
through discovery. See Walsh, 2023 WL 1966921, at *11 (“Defendants complain Wert’s
opinions are not based on ‘sufficient facts or data’ because he did not review ‘thousands of
pages’ they believe are relevant. . . . Defendants do not cite any authority establishing such
a rule[.]”)
Plaintiffs contend “MacLean has no idea who reviewed what [during document
review], what standards guided their review, or what criteria governed the selection of
documents to include or exclude from the report.” Doc. 5203 at 8. As discussed, MacLean
may rely on data gathered by his assistants – he may employ assistants to perform
document review, a task well within his ken. See Kim, 2023 WL 3674673, at *1 (“[A]n
expert may rely on . . . reliable data collected by others[.]” (citation modified)); Wolkowitz,
2008 WL 1885770, at *4 (“An expert may use assistants in performing his work, so long
as those assistants do not exercise professional judgment that is beyond the expert’s ken.”
(citation modified)).
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 9 of 22
10
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
Beyond challenging MacLean’s literature and document reviews, Plaintiffs contend
the level of assistance in MacLean’s report renders his non-testing opinions without any
reliable methodology. Doc. 5203 at 8-9. They argue that “[l]aboratory testing, material
analyses, and data interpretation were, once again, all performed by others. Even the
process of ‘building out the structure’ of the report was largely handled by helpers.” Id. at
8. But Rule 702 does not place parameters on how an assistant can help with an expert’s
report, so long as they “do not exercise professional judgment that is beyond the expert’s
ken” and the expert substantially participates. Wolkowitz, 2008 WL 1885770, at *4;
Arrowood, 2021 WL 5989746, at *2. Both conditions are met here.
2. MacLean’s Methodology.
Of the six pages Plaintiffs dedicate to arguing MacLean’s non-testing opinions lack
methodology for his use of assistants, they spend only one paragraph contending that his
methodology fails because it is not sufficiently described in his report. Doc. 5203 at 9.
Plaintiffs then commit all 13 pages of their reply to expanding on and particularizing this
one-paragraph argument. Doc. 6024. The Court disagrees with the argument. Defendants
have shown by a preponderance of the evidence that MacLean’s non-testing opinions are
supported by a reliable methodology.
Contrary to Plaintiffs’ contention that “[n]o methodology whatsoever is disclosed
in [MacLean’s] report” (id. at 3), the Methodology Employed section describes MacLean’s
methodology. Doc. 5203-1 at 23. He reached his non-testing opinions by “review[ing]
relevant production documents provided by counsel, deposition testimony[,] and scientific
and engineering literature.” Id. He reviewed “design history files, Bard internal
performance testing data, manufacturing records and protocols, regulatory submissions,
post market surveillance data, industry standards, regulatory guidance, clinical data, and
studies conducted on relevant materials and devices,” as well as Plaintiffs’ experts’ reports.
Id. MacLean provides a list of materials he considered. Id. (citing Appendix C). His
review of these sources for his non-testing opinions was “based on [his] education,
technical training, and experience[.]” Id.
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 10 of 22
11
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
The Court acknowledges MacLean’s methodology section is short on particularity,
but it is not “no methodology whatsoever” as Plaintiffs contend. Further, in identifying
MacLean’s methodology, the Court may consider his report in its entirety, including his
qualifications and experience. See In re Bard IVC Filters Prods. Liab. Litig., No. MDL
15-02641-PHX DGC, 2017 WL 6523833, at *5 (D. Ariz. Dec. 21, 2017) (“[A]lthough Dr.
Parisian’s description of her methodology could be clearer, the Court concludes from her
report as a whole that her methodology is sufficient to support [her] opinions[.]”).
MacLean has actively practiced in the field of polymer science and engineering for
30 years. Doc. 5203-1 at 18. As a Principal Engineer and the Practice Director of the
Polymers and Chemistry Practice at Exponent – a technical consulting firm engaged in “the
analysis and prevention” of engineering and scientific failures – MacLean “lead[s] a
technical staff of approximately 60 practicing scientist and engineers[.]” Id. at 17. He is a
Senior Member of the Society of Plastics Engineers (SPE), a member of SPE’s Medical
Plastics division, and an “active, voting member of the ASTM International D20
Committee which develops and publishes consensus-based standards for polymeric
materials.” Id. MacLean has engaged in scientific research that “reviewed and synthesized
hundreds of journal articles, authoritative texts and other technical treatises to guide [his]
understanding of the physico-chemical interactions between the human body and
implantable devices made from polymeric materials.” Id. at 21. He has published and
presented his research. Id. at 22. He employed this “review and synthesizing” process
when developing the opinions at issue here. See Doc. 5203-5 at 6-7.
The Court is satisfied MacLean applied a sufficiently reliable methodology to his
non-testing opinions. See Smilovits v. First Solar, Inc., No. CV12-00555-PHX-DGC, 2019
WL 6875492, at *4 (D. Ariz. Dec. 17, 2019) (“Under Rule 702 and Daubert, expert
testimony ‘is reliable if the knowledge underlying it has a reliable basis in the knowledge
and experience of the relevant discipline.’” (quoting Primiano v. Cook, 598 F.3d 558, 565
(9th Cir. 2010)).
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 11 of 22
12
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
Plaintiffs argue MacLean’s report outline is evidence of the absence of a
methodology. Doc. 6024 at 3. They argue that “[b]efore reviewing a single scientific
source, Dr. MacLean created an outline of his Opinions[.]” Id. This allegation is not
supported by the testimony Plaintiffs cite. MacLean testified he worked from an outline
as “a starting point for the expert report” and that it preceded document review, but he
never stated the outline contained pre-determined opinions. Doc. 6024-1 at 29-30.
MacLean testified about “building out” the report and how it was a “living document” (id.
at 32, 38-39, 40-41), but this testimony does not support a finding that he reached his
opinions before completing his research. In fact, MacLean clarified that the building-out
process was an objective inquiry. Id. at 29 (“There was no bias in terms of we’re looking
for specific citations because we wanted to, you know, have some specific objective on
something in the outline. I mean, that’s not how it works.”). The Court is not persuaded
that MacLean’s use of a report outline shows an absence of methodology.
Plaintiffs contend MacLean’s experience cannot serve as methodology, relying on
Farmers Ins. Co. of Ariz. v. DNS Auto Glass Shop LLC, 2024 WL 1256042 (D. Ariz. Mar.
25, 2024). Doc. 6024 at 4. But in Farmers, the expert rendered opinions that were “only
based off of [his] beliefs.” 2024 WL 1256042, at *9. The expert based other opinions on
his “analysis,” which he never described. Id. (“Mr. Hart’s report never describes this
analysis, what it entailed, or what it found. He states that he analyzed ‘the local physical
market’ in the region and relied on his ‘background and many years of experience in the
industry,’ but he never describes the local market or what it disclosed, or what he learned
about prices from his years on the industry. One reads his report in vain for any indication
of the facts and data on which he relied or the methodology used in his analysis.” (citation
omitted)). This Court excluded the expert’s opinions in Farmers because he relied on his
experience with no explanation of how it helped him reach his opinions and with no facts
or data to support the opinions. That is not the case here. MacLean describes his
methodology in his report and cites extensively to the facts and data on which he relies.
See Doc. 5203-1. His experience only supplements his described methodology.
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 12 of 22
13
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
Plaintiffs argue the testimony in MacLean’s errata (Doc. 6024-1) should not be
considered because his errata were disclosed after Plaintiffs’ motion to exclude was filed
and “exceed the scope of permissible changes to deposition testimony under Federal Rule
of Civil Procedure 30(e).” Doc. 6024 at 6. Yet, Plaintiffs cite MacLean’s errata repeatedly
to support their own arguments. See, e.g., Doc. 6024 at 3 (citing Doc. 6024-1 at 27-32, 38-
39, 40-41), 4 (citing Doc. 6024-1 at 6-7, 10, 13, 19-21, 22-24). While the Court has cited
a few statements from MacLean’s errata in support of its findings, the Court would reach
the same conclusion even if it disregarded the errata.
Plaintiffs lastly argue that Exponent’s quality management process, which MacLean
references when explaining the mandated peer review of the literature and documents he
relies upon (see, e.g., Doc. 5757-1 at 3) is not “proof” of reliability. Doc. 6024 at 11. True,
but it is a factor the Court may consider in evaluating the reliability of MacLean’s opinions,
and it supports the reliability of those opinions.
Rule 702 rejects a bright-line approach in determining the reliability of an expert’s
opinions. See Fed. R. Evid. 702 advisory committee’s note to 2000 amendment (“[N]ot
only must each stage of the expert’s testimony be reliable, but each stage must be evaluated
practically and flexibly without bright-line exclusionary (or inclusionary) rules.”; “[T]he
trial judge must have considerable leeway in deciding in a particular case how to go about
determining whether particular expert testimony is reliable.” (citations omitted)). The
Court takes that approach here and finds MacLean’s non-testing opinions are supported by
a reliable methodology.
B. Safer Alternative Design Opinions.
Plaintiffs argue MacLean’s opinions on safer alternative designs lack a reliable
methodology or support from facts or data. Doc. 5203 at 10. They claim the opinions “are
unsupported by any independent analysis, discernible methodology, and/or factual basis,” and
that “such support does not exist anywhere in Dr. MacLean’s report.” Id. Defendants
respond by walking through the detailed analysis conducted by MacLean on Plaintiffs’
alternative design theories, noting his extensive citations to literature, documents, and data.
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 13 of 22
14
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
Doc. 5757 at 10-14. Plaintiffs’ reply brief says nothing in response – it presents no further
argument on this issue. Doc. 6024.
MacLean cites support for his rebuttal opinions on Plaintiffs’ safer alternative
designs, including literature and internal documents. See Doc. 5203-3 at 73-93. He
explains that “[t]o gain a more reliable understanding of Plaintiffs’ proposed alternatives
and evaluate their claims, [he] . . . collected associated literature from both produced
documents and the public domain where possible.” Id. at 76. He categorized Plaintiffs’
experts’ proposed technologies in two appendices (I and J) (id. at 75), which he cites
throughout his safer-alternative-design section. See, e.g., id. at 77 nn. 668-69, 82 nn. 678-
81. He explains in detail why he finds Plaintiffs’ proposed alternative designs
unpersuasive, including that some are based merely on bench tests, some are addressed
only in academic literature (which often never results in actual products, as he explains),
none has been approved by the FDA for use in IPCs, and many do not satisfy the
requirements Plaintiffs’ experts identify for use in actual devices. Id. at 73-93. The Court
will not take time to recount the full detail of MacLean’s report, but it clearly is supported
by sufficient facts and data.
For methodology, MacLean reviewed and analyzed various sources, including the
“cited literature and produced documents” from Plaintiffs’ expert reports. Doc. 5203-1 at
23. His review and analysis was informed by his “education, technical training, and
experience.” Id.; see Smilovits, 2019 WL 6875492, at *4 (“Under Rule 702 and Daubert,
expert testimony ‘is reliable if the knowledge underlying it has a reliable basis in the
knowledge and experience of the relevant discipline.’” (quoting Primiano, 598 F.3d at
565)).
Plaintiffs challenge only one specific safer alternative design opinion. MacLean
opines that “‘Plaintiffs’ experts frequently misrepresent the clinical benefits associated
with’ their proposed safer alternative technologies.” Doc. 5203 at 10; see Doc. 5203-3 at
83.1 Plaintiffs argue MacLean cites “no scientific evidence contradicting Plaintiffs’
1 Throughout their motion, Plaintiffs’ record cites to MacLean’s report consistently fail to
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 14 of 22
15
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
experts’ opinions, no refutation of the existing body of data on the issue, and no
methodology at all.” Doc. 5203 at 10. But MacLean provides scientific evidence through
his scientific testing of competitor catheters. He opines that “[Plaintiffs’ expert] presents
BioFlo as a non-defective, mechanically superior alternative, even though testing indicates
that BioFlo catheters use more barium sulfate than Bard’s ChronoFlex and silicone
catheters (see Section 6.5).” Doc. 5203-3 at 86 (emphasis omitted). Section 6.5 of
MacLean’s report describes the results of his scientific testing of competitor catheters. Id.
at 38-39 (see Section 6.5.2 and Table 13). MacLean also opines that the Rhapsody
catheters discussed in Plaintiffs’ expert Dr. Ratner’s report “were not marketed as ‘fracture-
resistant’ as stated by the Plaintiff’s experts, but as, ‘flexible, non-compressible, and
reinforced silicone.’” Id. at 84-85 (citations omitted).
MacLean further supports his opinion that Plaintiffs’ experts misrepresent the
clinical benefits of alternative technologies by demonstrating their absence of clinical
support. See id. at 87 (“I was unable to find substantial information from the manufacturer
on the Orca IPC[.]”; “Dr. Ratner does not cite any documents when stating [his opinion.]”;
“Dr. Ratner and Dr. Sheikhi provide no additional documentation or citations that
demonstrate that the Orca IPC would offer any clinical benefit or meet applicable device
standards.”; “[D]ue to the limited information available, I cannot reasonably assess that the
Orca IPC is a relevant alternative.”). The Court finds MacLean’s methodology sufficiently
reliable under Rule 702(c) & (d).
Plaintiffs next challenge MacLean’s literature review. They argue MacLean does
not address literature “central to Plaintiffs’ experts’ [safer alternative design] opinions”
such as the Suleman or Piran study, among others. Docs. 5203 at 10-11, 6024 at 3.
Plaintiffs may cross-examine MacLean on his failure to address contrary authority, but the
Court does not find this a basis for the exclusion of his opinions.
Plaintiffs contend that MacLean criticizes Plaintiffs’ experts for relying on the Iida
study yet “never evaluated its data, addressed its findings, or applied any methodology to
direct the Court to the relevant material. The Court has tried to find the relevant parts of
the record in preparing this order.
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 15 of 22
16
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
assess its significance,” and apparently “parrots Dr. Thisted’s conclusion about Iida.”
Doc. 5203 at 11-12. This is incorrect. MacLean independently reviewed the Iida study
and explained why its results offer little support for Plaintiffs’ alternative design. See
Doc. 5203-3 at 90 (“The clinical evidence presented by Iida et al. is of limited value in
determining the alleged benefits of reinforced catheters compared to the material
technologies used in Bard’s IPCs. Catheter fracture occurred for just one of the 308
patients with IPC’s assessed in the study, corresponding to a 0.3% failure rate. The authors
did not state which catheter type fractured, meaning it is unknown whether the fractured
catheter was from an Orphis IPC, a Bard IPC, or from a completely different brand of IPC
(31 IPCs used in the study were not identified).” (footnote omitted)).
MacLean does not parrot Thisted’s conclusion about the Iida study, but permissibly
relies on Thisted’s conclusion to supplement his own opinion. Id. (“Thus, in addition to
the limitations discussed by Dr. Thisted regarding the overall structure of the cited study,
the results themselves do not support any conclusions regarding the risk of fracture in IPCs
with different catheter designs.” (emphasis omitted and added) (citation omitted)); In re
Bard IVC Filters Prods. Liab. Litig., No. MDL 15-02641-PHX DGC, 2018 WL 495188,
at *3 (D. Ariz. Jan. 22, 2018) (explaining experts may not “simply parrot the opinions of
other experts, . . . but they can rely on opinions stated by other experts”). Plaintiffs’
criticisms of MacLean’s review of the Iida study are appropriate for cross-examination but
not a basis for exclusion of his opinions.
Plaintiffs argue MacLean “does not engage with the data [in his Appendices I and
J], evaluate the underlying methodologies, or assess the findings of the cited studies or
evidence supporting these alternative designs.” Doc. 5203 at 12. The Court disagrees.
MacLean discusses the articles and studies contained in the appendices throughout his
report. See, e.g., Doc. 5203-3 at 77 nn. 668-69, 82 nn. 678-81. And even if MacLean did
not review every source referenced in Appendices I and J, that would not render his
opinions inadmissible. See In re Bard IVC Filters Prods. Liab. Litig., No. MDL 15-02641-
PHX DGC, 2017 WL 11696720, at *3 (D. Ariz. Dec. 22, 2017) (“The parties disagree on
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 16 of 22
17
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
the extent to which the doctors reviewed the factual material on which they rely. . . . .
[E]ven if the doctors did not review every relevant Bard document . . . and relied to some
extent on Dr. Kessler’s extensive chronology and summary, the Court cannot conclude that
this renders their opinions inadmissible.”).
Plaintiffs argue, “[a]t a bare minimum, that MacLean should be precluded from
commenting on scientific literature or commercial products never discussed in his report”
and “from offering opinions about scientific literature or commercial products that are
merely inventoried in Appendices I and J but are not otherwise discussed in his report.”
Doc. 5203 at 11, 13. The Court instructed the parties in Case Management Order 9 that
their experts’ testimony would be confined to their report and depositions. Doc. 114 ¶ II.
F. Plaintiffs are free to object to MacLean’s testimony if they believe he is exceeding the
scope of his report and deposition testimony.
C. 510k Clearence Opinions.
Plaintiffs argue MacLean renders opinions about the 510(k) process and industry
standards that “suggest[] that those regulatory and industry mechanisms establish the
‘safety and effectiveness’ of Defendants’ ports.” Doc. 5203 at 13. Defendants contend
MacLean only describes “substantial equivalence” under the 510(k) process and the
industry standards used to demonstrate the comparative “safety and effectiveness” of a
device submitted to the 510(k) process. Doc. 5757 at 14 (citing Docs. 5203-1 at 51, 5203-
3 at 42).
Plaintiffs are correct that “[t]he 510(k) process focuses on device equivalence, not
device safety.” In re Bard IVC, 289 F. Supp. 3d 1045, 1047-48 (D. Ariz. 2018). It does
not determine a device is “safe and effective,” only “that the device is substantially
equivalent to the predicate device.” In re Bard IVC Filters Prods. Liab. Litig., No. MDL
15-02641-PHX DGC, 2017 WL 5625547, at *7 (D. Ariz. Nov. 22, 2017), aff’d sub nom.
In re Bard IVC Filters Prod. Liab. Litig., 969 F.3d 1067 (9th Cir. 2020) (citation omitted).
While the FDA may find “the device ‘is as safe and effective’ as the predicate device,” this
is a comparative exercise, not a safety determination. Id. “The assumption is that the
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 17 of 22
18
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
predicate device is safe and effective enough to be on the market, and that the proposed
device, if sufficiently similar, must be so as well.” Id.
MacLean opines that “a substantial[ly] equivalent device [submitted to the 510(k)
process] can have different technological characteristics but can be accompanied by
appropriate clinical or scientific data that demonstrate the device is as safe and effective as
a legally marketed device.” See Doc. 5203-1 at 51 (emphasis added). This accurately
describes the 510(k) process.
MacLean further opines that “FDA requires that submissions for 510(k) clearance
of new IPC devices demonstrate substantial equivalence to an existing IPC product on the
market to demonstrate safety and effectiveness.” Doc. 5203-3 at 42. This opinion appears
to overstate the effect of FDA 510(k) clearance. It implies that substantial equivalence
shows a submitted device is safe and effective. Instead, the comparative exercise under
510(k) may only demonstrate the device is “as safe and effective” as the predicate device.
Defendants should ensure MacLean discusses the 510(k) process in comparative terms
only. Plaintiffs may object if he does not.
Plaintiffs separately argue that under Federal Rule of Civil Procedure 403,
“[a]llowing Dr. MacLean to recast regulatory clearance and industry standards as proof of
safety and efficacy also would create a substantial danger of unfair prejudice, confuse the
issues and mislead the jury.” Doc. 5203 at 13-14 (citation omitted). MacLean’s opinions,
when cabined as instructed by the Court, will not assert that clearing the 510(k) process is
tantamount to an FDA determination of safety and efficacy.
D. Opinions on Dr. El-Ghannam’s SEM Analyses.
MacLean provides rebuttal opinions critiquing the SEM analyses of Plaintiffs’
expert, Dr. El-Ghannam. Plaintiffs contend MacLean’s rebuttal opinions are unreliable
because they lack a factual foundation under Rule 702(b). Doc. 5203 at 14.
Through his SEM analyses, El-Ghannam applied metal coating to sample catheters
to increase their conductivity. Id. MacLean criticizes El-Ghannam for applying too thick
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 18 of 22
19
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
a metal layer, which MacLean opines caused the cracks and defects El-Ghannam attributes
to flaws in the catheter surface. Id.
Plaintiffs contend MacLean “analyzed too few samples and disregarded the visible
fine features and surface degradation in Dr. El-Ghannam’s images,” which Plaintiffs claim,
“by [MacLean’s] own cited principles, would not be discernible if the metal coating were
excessively thick.” Id. Plaintiffs quote MacLean’s report, arguing “Exponent concedes
that it only reviewed ‘[s]ome’ of Dr. El-Ghannam’s catheter samples.” Id. at 15. The
citation provided by Plaintiffs does not direct the Court to the quoted language, but the
Court considers the following language from MacLean’s report: “Some of Dr. El-
Ghannam’s SEM catheter samples were analyzed using SEM after he completed his own
analysis.” Doc. 5203-4 at 2 (emphasis added). From his analysis of “some” of El-
Ghannam’s catheter samples, MacLean concluded the metal coating applied by El-
Ghannam was too thick. Id.
Defendants direct the Court to testimony from Dr. Mkhoyan, who worked with
MacLean to review the El-Ghannam SEM analysis and who testified that he “carefully
reviewed every single image from Dr. El-Ghannam’s report[.]” Doc. 5757 at 16 (quoting
Doc. 5754-2 at 31). The Court will not assume MacLean reviewed the same materials as
Mkhoyan to find MacLean’s independent opinions have a sufficient factual foundation.
While MacLean’s report does not explain how many catheter samples were tested
in his analysis, there is no specific number that renders his analysis unreliable. Appendix
E of his report provides numerous SEM images depicting the results of El-Ghannam’s
catheter samples tested by MacLean (Doc. 5205-1). These images show MacLean tested
numerous catheters from El-Ghannam’s sample set. The Court finds this portion of
MacLean’s report to be sufficiently supported by facts and data and a sufficiently reliable
method of analysis to be presented to the jury. Plaintiffs dispute the evidentiary strength
of the SEM images contained in Exhibit E. See Doc. 5203 at 16 (“[N]one of those images
contain measurements of the supposed extra-thick metal coating, and the vast majority are
not taken from the orientation necessary to even roughly estimate thickness, whether of the
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 19 of 22
20
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
coating or the catheter surface.”). Plaintiffs may raise this contention on cross-
examination.
The Court also agrees with Defendants’ argument that “[i]t is the proper role of
rebuttal experts to critique plaintiffs’ expert’s methodologies and point out potential flaws
in the plaintiff’s experts’ reports.” Doc. 5757 at 17 (quoting Aviva Sports, Inc. v. Fingerhut
Direct Mktg., Inc., 829 F. Supp. 2d 802, 834 (D. Minn. 2011)). Calling into question the
accuracy of El-Ghannam’s findings for even some sample catheters properly “point[s] out
the potential flaws” in his report. Aviva Sports, Inc., 829 F. Supp. 2d at 834.
The Court reaches the same conclusion on Plaintiffs’ contention that MacLean’s
analysis “rests on only three measurements taken from a single location on a single
catheter.” Doc. 5203 at 15. Plaintiffs cite figure 58 from MacLean’s report that depicts
three measurements recorded from a single catheter. See Doc. 5203-4 at 3 (figure 58). But
this figure is demonstrative, not exhaustive. It depicts one of the sample catheters tested
by MacLean. It does not purport to contain all tested samples. Nor does MacLean state he
measured only one of the tested catheters or that the average thickness he recorded
(135.6nm) was calculated from a single catheter. That MacLean prefaced his analysis of
the tested sample catheters by stating he analyzed “[s]ome of Dr. El-Ghannam’s SEM
catheter samples” (Doc. 5203-4 at 2) suggests the contrary – that MacLean took
measurements from a plurality of sample catheters.
Plaintiffs also suggest that “based on the arrows drawn on his figure” (figure 58),
MacLean’s measurements of the catheter’s metal coating may have measured the “double-
sided copper tape used to mount the sample,” not the catheter, rendering his opinion on the
metal coating’s thickness “misleading and methodologically flawed.” Doc. 5203 at 15.
Defendants contend MacLean “specifically addressed both the catheters and the copper
tape they are mounted to” and “performed EDS to confirm that the cracks were in the
overly thick gold coating applied to the samples.” Doc. 5757 at 16 n.15 (citing Doc. 5203-
4 at 3-4); see also Doc. 5203-4 at 5 (figure 60 depicting MacLean’s EDS analysis). The
Court will not rely on Plaintiffs’ hypothesized suggestion that MacLean improperly
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 20 of 22
21
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
measured the tape rather than the sample catheter. In his report, MacLean distinguishes
between the catheter and the tape. See Doc. 5203-4 at 4 (“As can be seen in the figure, the
same cracking morphology that is present on the catheter surface is also present in the
background of the SEM images on the copper tape that the catheter are mounted to.”).
MacLean’s EDS analysis also demonstrates he identified compositional differences in the
catheter’s polymer surface and its metal coating (id. at 5), indicating he measured the
catheter and not the tape. The Court will not construe the placement of an instructive
arrow on a demonstrative figure as somehow undermining MacLean’s methodology.
Plaintiffs contend MacLean’s “central claim is that all of Dr. El-Ghannam’s SEM
images are unreliable because of excessive metal coating,” and that to support such a claim,
he “was required to measure a sufficiently large number of samples to determine whether
overly thick coatings actually occurred with any frequency.” Doc. 5203 at 16. Plaintiffs
never cite where MacLean’s report makes such a claim.
Plaintiffs next contend MacLean “completely fails to address evidence of surface
degradation plainly visible in many of Dr. El-Ghannam’s SEM images,” which they claim
demonstrates defects in the catheter samples and “bacterial colonies” on their surface. Id.
They argue this demonstrates an “absence of methodology altogether.” Id. at 17. The
Court disagrees. As acknowledged by Plaintiffs, MacLean addresses El-Ghannam’s
opinions on the presence of surface degradation in the sample catheters. See Doc. 5203-4
at 3-7 (MacLean responding to El-Ghannam’s observations of (1) “a cracked morphology
on the surface of the pristine polymer catheters,” and (2) “polymer . . . flaking off the
surface of the explanted SEM samples”). MacLean does not respond to every type of
surface degradation discussed by El-Ghannam (Doc. 5203 at 16) or his allegations of
biofilm being present (Doc. 5205-5 at 7-8), but the Court does not find this to be a basis
for exclusion. Plaintiffs may cross-examine MacLean on his failure to address certain
opinions of El-Ghannam.
Plaintiffs also contend MacLean’s opinion that El-Ghannam applied too thick a
metal coating to his sample catheters – averaging 135.6 nm of the tested sample catheters
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 21 of 22
22
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
(Doc. 5203-4 at 2) – is undermined by his report stating that “when a gold coat is applied
at more than 10-25 nm, detailed features of the surfaces are lost” (id. at 6), and yet “fine
surface details” were “clearly visible in many of Dr. El-Ghannam’s SEM images[.]” Doc.
5203 at 17. This alleged inconsistency may be raised on cross examination. It is not a
basis for exclusion of MacLean’s otherwise sufficiently supported opinions on El-
Ghannam’s SEM analyses.
IT IS ORDERED that Defendants’ motion to exclude the opinions of Dr. Steven
MacLean (Doc. 5203) is denied.
Dated this 24th day of February, 2026.
Case 2:23-md-03081-DGC Document 7222 Filed 02/24/26 Page 22 of 22

Continue sua pesquisa no ChatGPT ou Claude