Evaluation of Internal Standard Responses During Chromatographic Bioanalysis: Questions and Answers

us-fda-guidance-6578abce29366dab592f554c9fa66341Evaluation of Internal Standard Responses During Chromatographic Bioanalysis: Questions and AnswersGuidance

Abrir fonte

Administrative guidance, not enacted legislation. This captured publisher edition preserves the listed document and attachments; later developments are not independently consolidated.

Docket Number:
FDA-2009-D-0008
Issued by:
Guidance Issuing Office
Center for Drug Evaluation and Research

This guidance provides recommendations to sponsors, applicants, and contract research organizations regarding internal standard (IS) response variability in chromatographic analytical data submitted in investigational new drug applications, new drug applications, abbreviated new drug applications, biologics license applications, and supplements. Chromatographic analytical methods are commonly used to quantitate analyte concentrations in samples from nonclinical and clinical studies to support regulatory submissions. Depending upon its source, IS response variability may impact the accuracy of analyte concentration measurements. This question and answer (Q&A) document provides the Agency’s current thinking on IS response variability and its potential impact on the accuracy of analyte concentration measurements. This Q&A also suggests approaches to determine whether observed IS response variability is likely to impact the accuracy of the data such that further investigation into the root cause(s) is warranted.


Submit Comments

You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))

If unable to submit comments online, please mail written comments to:

Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852

All written comments should be identified with this document's docket number: FDA-2009-D-0008.

Evaluation of Internal
Standard Responses During
Chromatographic
Bioanalysis: Questions and
Answers
Guidance for Industry
U.S. Department of Health and Human Services
Food and Drug Administration
Center for Drug Evaluation and Research (CDER)
September 2019
Biopharmaceutics
Evaluation of Internal
Standard Responses During
Chromatographic
Bioanalysis: Questions and
Answers
Guidance for Industry
Additional copies are available from:
Office of Communications, Division of Drug Information
Center for Drug Evaluation and Research
Food and Drug Administration
10001 New Hampshire Ave., Hillandale Bldg., 4 th Floor
Silver Spring, MD 20993-0002
Phone: 855-543-3784 or 301-796-3400; Fax: 301-431-6353
Email: druginfo@fda.hhs.gov
https://www.fda.gov/Drugs/GuidanceComplianceRegulatoryInformation/Guidances/default.htm
U.S. Department of Health and Human Services
Food and Drug Administration
Center for Drug Evaluation and Research (CDER)
September 2019
Biopharmaceutics
Contains Nonbinding Recommendations
i
TABLE OF CONTENTS
I. 	INTRODUCTION............................................................................................................. 1
II. 	QUESTIONS AND ANSWERS ....................................................................................... 2
Contains Nonbinding Recommendations
1
Evaluation of Internal Standard Responses during
Chromatographic Bioanalysis: Questions and Answers
Guidance for Industry1
This guidance represents the current thinking of the Food and Drug Administration (FDA or Agency) on
this topic. It does not establish any rights for any person and is not binding on FDA or the public. You can
use an alternative approach if it satisfies the requirements of the applicable statutes and regulations. To
discuss an alternative approach, contact the FDA office responsible for this guidance as listed on the title
page.
I. 	INTRODUCTION
This guidance provides recommendations to sponsors, applicants, and contract research
organizations regarding internal standard (IS) response 2 variability in chromatographic
analytical data submitted in investigational new drug applications, new drug applications,
abbreviated new drug applications, biologics license applications, and supplements.
Chromatographic analytical methods are commonly used to quantitate analyte concentrations in
samples from nonclinical and clinical studies to support regulatory submissions. Depending upon
its source, IS response variability may impact the accuracy of analyte concentration
measurements. This question and answer (Q&A) document provides the Agency’s current
thinking on IS response variability and its potential impact on the accuracy of analyte
concentration measurements. This Q&A also suggests approaches to determine whether
observed IS response variability is likely to impact the accuracy of the data such that further
investigation into the root cause(s) is warranted.
In general, FDA’s guidance documents do not establish legally enforceable responsibilities.
Instead, guidances describe the Agency’s current thinking on a topic and should be viewed only
as recommendations, unless specific regulatory or statutory requirements are cited. The use of
the word should in Agency guidances means that something is suggested or recommended, but
not required.
1 This guidance has been prepared by the Office of Study Integrity and Surveillance, the Office of Generic Drugs,
and the Office of Clinical Pharmacology in the Center for Drug Evaluation and Research at the Food and Drug
Administration. You may submit comments on this guidance at any time. Submit comments to Docket No. FDA-
2017-D-6821 (available at https://www.regulations.gov/docket?D=FDA-2017-D-6821). See the instructions in that
docket for submitting comments on this and other Level 2 guidances.
2 An IS response is a measurement of the signal from the IS, which is typically generated during liquid
chromatography-mass spectrometry bioanalysis.
Contains Nonbinding Recommendations
2
II. 	QUESTIONS AND ANSWERS
Q1: What are internal standards in analytical chemistry, and why are they used in
bioanalysis?
A1: In analytical chemistry, ISs are usually a structural analog or stable isotope of an analyte of
interest and are commonly used in chromatographic analytical methods to correct for variability
in sample processing and analysis. In chromatographic bioanalysis, an IS is added to all samples,
including calibration standards (Cals), quality controls (QCs), and subject samples prior to
extraction. The specific IS selected for a method should have similar physicochemical properties
to the analyte of interest so that it behaves in a similar fashion to the analyte and reflects any
changes to the analyte measurement that may occur during sample processing and analysis.
Q2: What are sources of IS response variability?
A2: By design, the intention is to add the same amount of IS to all samples. However, variability
in IS responses among samples is occasionally observed, even among samples analyzed in the
same run. Potential sources of IS response variability include human errors made during sample
preparation or processing, instrumental issues that may occur during analysis, and matrix effects.
Q3: When is IS response variability not likely to impact the accuracy of the data?
A3: IS response variability is not likely to impact the accuracy of the data when the range of IS
responses for subject samples is similar to the range of IS responses for Cals/QCs in the same
analytical run. For example, variability in the IS responses for subject samples is not likely to
impact the accuracy of the data under the following circumstances:
Example 1: IS responses for subject samples are similar to IS responses observed for Cals/QCs
in the same analytical run (see Figure 1).
Contains Nonbinding Recommendations
3
Figure 1. Plot of IS responses in an analytical run, exemplifying IS responses for subject
samples that are similar to IS responses observed for Cals/QCs
Example 2: IS responses for subject samples demonstrate a gradual drift or repeating pattern,
but the range of IS responses for subject samples is similar to the range of IS responses for
Cals/QCs in the same run (see Figure 2).
Figure 2. Plot of IS responses in an analytical run, exemplifying IS responses for subject
samples demonstrating a gradual drift or repeating pattern similar to Cals/QCs
1.E+07
2.E+07
3.E+07
4.E+07
5.E+07
0 	4 	8 	12 	16 	20 	24 	28 	32 	36 	40 	44 	48 	52
IS Peak Area Counts
Injection Number
Cals
QCs
Blanks
Subject 1
Subject 2
1.E+07
2.E+07
3.E+07
4.E+07
5.E+07
0 	4 	8 	12 	16 	20 	24 	28 	32 	36 	40 	44 	48 	52
IS peak Area Counts
Injection Number
Cals
QCs
Blanks
Subject 1
Subject 2
Contains Nonbinding Recommendations
4
Q4: When can IS response variability impact the accuracy of the data?
A4: IS response variability may impact the accuracy of the data when the range of the IS
responses for subject samples is different than the range of IS responses for Cals/QCs in the
same run. For example, IS response variability for subject samples may impact the accuracy of
the data under the following circumstances:
Example 1: IS responses for one or more subject samples are substantially different from the IS
responses for the majority of the other subject samples, and the IS responses for Cals/QCs do not
demonstrate a similar variability pattern (see Figure 3).
Figure 3. Plot of IS responses in an analytical run, exemplifying an IS response for a
subject sample that is substantially different from the responses for the majority of the
other subject samples and Cals/QCs
Example 2: There is a gradual increase or decrease extended beyond the range of the IS
responses for subject samples, and there are no IS responses for QCs interspersed among the
impacted segment of the run (see Figure 4).
1.E+07
2.E+07
3.E+07
4.E+07
5.E+07
6.E+07
7.E+07
8.E+07
9.E+07
0 	4 	8 	12 	16 	20 	24 	28 	32 	36 	40 	44 	48 	52
IS Peak Area Counts
Injection Number
Cals
QCs
Blanks
Subject 1
Subject 2
Contains Nonbinding Recommendations
5
Figure 4. Plot of IS responses in an analytical run, exemplifying a gradual decrease in IS
responses for subject 2 samples without IS responses for QCs interspersed among the
impacted segment of the run
Example 3: IS responses for subject samples are consistently lower or higher than IS responses
for Cals/QCs (see Figure 5).
Figure 5. Plot of IS responses in an analytical run, exemplifying IS responses for subject 2
samples that are consistently higher than IS responses for Cals/QCs
1.E+07
2.E+07
3.E+07
4.E+07
5.E+07
0 	4 	8 	12 	16 	20 	24 	28 	32 	36 	40 	44 	48 	52
IS Peak Area
Injection Number
Cals
QCs
Blanks
Subject 1
Subject 2
1.E+07
2.E+07
3.E+07
4.E+07
5.E+07
6.E+07
7.E+07
8.E+07
0 	4 	8 	12 	16 	20 	24 	28 	32 	36 	40 	44 	48 	52
IS Peak Area Counts
Injection Number
Cals
QCs
Blanks
Subject 1
Subject 2
Contains Nonbinding Recommendations
6
Reviewing the scatter plot of IS responses for all samples in the run might help identify
variability trends that may or may not impact the accuracy of the data.
Q5: How can sponsors address concerns about IS response variability?
A5: When IS responses for subject samples are outside of the ranges observed for Cals/QCs that
meet the run acceptance criteria, the subject sample concentration measurements might not be
accurate. Therefore, sponsors should determine whether this observed difference impacts the
accuracy of the data for subject samples.
For subject samples with IS responses significantly different from the responses for Cals/QCs (as
pre-defined in a standard operating procedure (SOP)) in a run (see Figure 3), pre-established
criteria for repeat analysis should be followed. The repeat analysis criteria should be justified and
the protocol for the number of replicates reanalyzed for each sample and final values reported
should be established in an SOP before initiation of the study.
For runs where a gradual drift or pattern in IS responses is observed between subject samples and
there are no IS responses for Cals/QCs within the impacted segment of the run, subject samples
displaying the aberrant IS responses (see Figure 4) should be reanalyzed. If the IS responses for
subject samples in the repeat analysis are similar to those of Cals/QCs, and if the analyte
concentrations in a repeat run are within 20% of the original assay results of the individual
samples, generally no further investigation into the cause of the initial variability is necessary.
Sample values to be reported should follow a pre-established procedure. However, if the IS
responses for subject samples in the repeat analysis are similar to those of Cals/QCs, but the
analyte concentrations in the repeat analysis are not within 20% of the original assay results of
the individual samples, then sponsors should report analyte concentrations from the repeat
analysis.
When IS responses for subject samples are consistently lower or higher than IS responses for
Cals/QCs (see Figure 5), a subset of subject samples should be reanalyzed. If IS responses of
subject samples in the repeat analysis are again not within the range for those of Cals/QCs (i.e.,
differences in IS responses in the original run are reproducible), and repeat analyte
concentrations are still comparable to the original values, additional data might be needed to
demonstrate that the differences in sample matrices (i.e., matrices for subject samples vs.
matrices for Cals/QCs) do not impact the accuracy of the analyte concentration measurements.
The approach(es) for investigation might include, but are not limited to, an analysis of QCs
prepared in pre-dose matrix of individual impacted subjects, serial dilution of impacted subject
samples using the matrices used to prepare Cals/QCs, or any other scientifically justified
approach.