Use of Real-World Evidence to Support Regulatory Decision-Making for Medical Devices: Guidance for Industry and Food and Drug Administration Staff

us-fda-guidance-6dcdb9c4e609bdcdcf571ca23e3ac89dUse of Real-World Evidence to Support Regulatory Decision-Making for Medical Devices: Guidance for Industry and Food and Drug Administration StaffGuidance

Abrir fonte

Administrative guidance, not enacted legislation. This captured publisher edition preserves the listed document and attachments; later developments are not independently consolidated.

Docket Number:
FDA-2023-D-4395
Issued by:
Guidance Issuing Office
Center for Devices and Radiological Health
Center for Biologics Evaluation and Research

FDA is issuing this guidance to clarify how FDA evaluates real-world data (RWD) to determine whether they are of sufficient quality for generating real-world evidence (RWE) that can be used in FDA regulatory decision-making for medical devices. This guidance includes expanded and updated recommendations to FDA staff and sponsors considering using RWE to support a regulatory submission for medical devices. This document supersedes the final guidance of the same name, issued August 31, 2017.


Submit Comments

You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))

If unable to submit comments online, please mail written comments to:

Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852

All written comments should be identified with this document's docket number: FDA-2023-D-4395.

Contains Nonbinding Recommendations
Use of Real-World Evidence to
Support Regulatory Decision-Making
for Medical Devices
Guidance for Industry and
Food and Drug Administration Staff
Document issued on December 18, 2025.
The draft of this document was issued on December 19, 2023.
This document supersedes “Use of Real-World Evidence to Support
Regulatory Decision-Making for Medical Devices,” issued on August 31, 2017.
For questions about this document regarding CDRH-regulated devices, contact the Office of
Clinical Evidence and Analysis at CDRHClinicalEvidence@fda.hhs.gov. For questions about
this document regarding CBER-regulated devices, contact the Office of Communication,
Outreach, and Development (OCOD) at 800-835-4709 or 240-402-8010, or by email at
industry.biologics@fda.hhs.gov.
U.S. Department of Health and Human Services
Food and Drug Administration
Center for Devices and Radiological Health
Center for Biologics Evaluation and Research
Contains Nonbinding Recommendations
Preface
Public Comment
You may submit electronic comments and suggestions at any time for Agency consideration to
https://www.regulations.gov. Submit written comments to the Dockets Management Staff, Food
and Drug Administration, 5630 Fishers Lane, Room 1061, (HFA-305), Rockville, MD 20852-
1740. Identify all comments with the docket number FDA-2023-D-4395. Comments may not be
acted upon by the Agency until the document is next revised or updated.
Additional Copies
CDRH
Additional copies are available from the Internet. You may also send an email request to
CDRH-Guidance@fda.hhs.gov to receive a copy of the guidance. Please include the document
number GUI01500012 and complete title of the guidance in the request.
CBER
Additional copies are available from the Office of Communication, Outreach, and Development
(OCOD), Center for Biologics Evaluation and Research (CBER), by calling 800-835-4709 or
240-402-8010, by email, industry.biologics@fda.hhs.gov, or from the Internet at
https://www.fda.gov/vaccines-blood-biologics/guidance-compliance-regulatory-information-
biologics/biologics-guidances.
Contains Nonbinding Recommendations
Table of Contents
I. Introduction ............................................................................................................................ 1
II. Background ............................................................................................................................ 2
III. Scope ...................................................................................................................................... 5
IV. Regulatory Context in Which Use of RWE May be Appropriate ......................................... 6
A. General Considerations for the Use of RWE ................................................................... 6
B. Application of Investigational Device Exemption (IDE) Requirements in 21 CFR 812
to Clinical Studies Using RWD ................................................................................................ 9
C. Application of RWD from Devices Authorized for Emergency Use Under Section 564
of the FD&C Act ..................................................................................................................... 11
V. Assessing Data Relevance and Reliability ........................................................................... 11
A. Relevance ....................................................................................................................... 13
(1) Data Availability ...................................................................................................... 13
(2) Linkages ................................................................................................................... 14
(3) Timeliness ................................................................................................................ 14
(4) Generalizability ........................................................................................................ 15
B. Reliability....................................................................................................................... 15
(1) Data Accrual ............................................................................................................ 15
(2) Data Quality and Integrity ....................................................................................... 17
VI. Considerations for Methodologies for Collection and Analysis of RWD to Generate RWE
19
A. Methods for Study Design Using RWD ........................................................................ 20
B. Defining Study Design Elements ................................................................................... 21
(1) Study Time, Relative to Index Date ......................................................................... 21
(2) Development of Conceptual and Operational Definitions for the Study Population,
Device, Comparator, Outcome, and Covariates................................................................ 23
(3) Appropriate Integration of Data Elements Within Study Design and Analysis ...... 25
VII. Documentation for FDA Review ......................................................................................... 26
A. Regulatory Submission Cover Letters ........................................................................... 26
B. Protocol .......................................................................................................................... 26
C. Report Including the Relevance and Reliability Assessment ........................................ 28
D. Additional Information .................................................................................................. 29
Contains Nonbinding Recommendations
Appendix A. Recommended Relevance and Reliability Elements for Documentation and FDA
Review .......................................................................................................................................... 30
Appendix B. Examples Where RWE is Used ............................................................................... 36
Contains Nonbinding Recommendations
1
Use of Real-World Evidence to
Support Regulatory Decision-Making
for Medical Devices
Guidance for Industry and
Food and Drug Administration Staff
This guidance represents the current thinking of the Food and Drug Administration (FDA or
Agency) on this topic. It does not establish any rights for any person and is not binding on
FDA or the public. You can use an alternative approach if it satisfies the requirements of the
applicable statutes and regulations. To discuss an alternative approach, contact the FDA staff
or Office responsible for this guidance as listed on the title page.
I. Introduction
FDA is issuing this guidance to clarify how FDA evaluates real-world data to determine whether
they are of sufficient quality for generating real-world evidence that can be used in FDA
regulatory decision-making for medical devices. This guidance also provides expanded and
updated recommendations to FDA staff and sponsors considering using real-world evidence to
support a regulatory submission for medical devices.1
Real-world data (RWD) are data relating to patient health status and/or the delivery of health
care routinely collected from a variety of sources.
Examples of RWD sources may include electronic health records (EHRs),2 medical claims data,
data from product and disease registries, and data gathered from other sources (such as digital
health technologies (DHTs)) that can inform on health status. Some RWD sources can also be
used as data collection and analysis infrastructure to support many types of study designs
including randomized and non-randomized controlled trials; single-arm studies with or without
comparison to an objective performance criterion, performance goal, or external control;
observational studies; and hybrid designs which combine elements of multiple study designs.
1 For more information on regulatory submissions for medical devices, see Sections III and VII.
2 For the purposes of this guidance, an “EHR” is an electronic record of health-related information on an individual
that conforms to nationally recognized/utilized interoperability standards and that can be created, managed, and
consulted by authorized clinicians and staff across more than one health care organization. Definition adapted from
“The National Alliance for Health Information Technology Report to the Office of the National Coordinator for
Health Information Technology on Defining Key Health Information Technology Terms April 28, 2008.”
Contains Nonbinding Recommendations
2
Real-world evidence (RWE) is the clinical evidence regarding the usage, and potential benefits
or risks, of a medical product derived from analysis of RWD.
This guidance includes factors that FDA considers important to demonstrate whether the RWD
are relevant and reliable for a particular regulatory decision relating to medical devices, as well
as information on how FDA intends to assess these factors. The recommendations and
considerations in this guidance will apply regardless of the RWD source and encompass
processes for conducting studies to generate RWE. FDA recognizes that there may be other
approaches to address the considerations identified in this document. We encourage sponsors to
discuss their approach with FDA, especially if the approach diverges from the recommendations
in this guidance.3 For example, sponsors may submit a Q-Submission when there are questions
on specific aspects of using RWD in a particular regulatory submission, such as questions
regarding the study protocol (see Section VII.B) and/or the sponsor’s approach to their relevance
and reliability assessment (see Section V).
FDA recognizes and anticipates that the Agency and industry may need up to 60 days to perform
activities to operationalize the recommendations within this guidance. For regulatory
submissions that are currently pending with FDA after publication of the guidance, as well as
those submissions received within 60 days following publication of the guidance, FDA generally
does not anticipate that sponsors will be ready to include the newly recommended information
outlined in the guidance in their submission. FDA, however, intends to review any such
information if submitted at any time.
In general, FDA’s guidance documents do not establish legally enforceable responsibilities.
Instead, guidances describe the Agency’s current thinking on a topic and should be viewed only
as recommendations, unless specific regulatory or statutory requirements are cited. The use of
the word should in Agency guidances means that something is suggested or recommended, but
not required.
II. Background
To protect and promote public health, FDA needs to understand and evaluate the available
evidence related to regulated products. For medical devices, common sources of available
evidence include information from non-clinical and clinical studies4 that are provided to FDA by
a device manufacturer or sponsor of a device in a premarket or postmarket submission. FDA
recognizes that a wealth of clinical data in the form of RWD are collected during the treatment
and management of patients. Although these data typically have different quality controls
compared to data collected within a traditional clinical study, under certain circumstances RWD
may be used to generate RWE to help inform or augment FDA’s understanding of the benefit-
risk profile of devices at various points in their life cycle.
3 See FDA’s guidance “Requests for Feedback and Meetings for Medical Device Submissions: The Q-Submission
Program.”
4 For the purposes of this guidance, we use the term “clinical studies” as a broad term to capture clinical research
regarding the safety or effectiveness of a device, regardless of study design. We use the term “traditional clinical
studies” to refer to clinical studies that do not utilize RWD.
Contains Nonbinding Recommendations
3
Many of the considerations and best practices for generating RWE are derived from the same
principles that govern generation of clinical evidence from traditional clinical studies, which are
generally referred to as good clinical practice (GCP). As with all clinical evidence, FDA’s
assessment of RWE provided in support of a particular regulatory decision will be evaluated as
part of the totality of information available to FDA. Per 21 CFR 860.7(c)(1), “[a]lthough [a]
manufacturer may submit any form of evidence to the Food and Drug Administration in an
attempt to substantiate the safety and effectiveness of a device, the agency relies upon only valid
scientific evidence to determine whether there is reasonable assurance that the device is safe and
effective.” RWE derived from relevant and reliable RWD may constitute valid scientific
evidence,5 depending on the study question, study design, regulatory decision, data source(s), and
design and analysis of the specific dataset derived from RWD source(s), and thus may be used to
support regulatory decisions (see Section V for more information on the relevance and reliability
assessment). The standard of valid scientific evidence will apply, as appropriate, regardless of
the source of clinical evidence provided. Additionally, FDA’s recommendations for clinical
evidence will generally apply to RWE, as appropriate, and sponsors should consider the
recommendations in other FDA guidances regarding design considerations and clinical evidence
generation.6
When appropriate, use of RWD may provide an efficient means of generating the necessary
clinical evidence to support regulatory decisions. Some traditional clinical studies may be narrow
in scope but allow for more control of sources of error and bias.7 In comparison, studies that
leverage RWD may be able to evaluate broader clinical experience to support the research
question. Clinical evidence can be generated from studies using RWD, alone or in combination
with data from traditional clinical studies (see Section VI for additional information on study
design and methodologies). Using appropriate design and methodologies, sponsors can leverage
the strengths of these approaches while minimizing potential weaknesses, ensuring the right
balance of external and internal validity for the study and regulatory purpose.
FDA recognizes that uncertainty regarding the benefits and risks of a device may remain after
completing a study using RWD, just as with studies relying on other sources of clinical data. The
recommendations in this guidance are intended to provide clarity on general considerations
5 Under 21 CFR 860.7(c)(2), “valid scientific evidence” is considered “evidence from well-controlled investigations,
partially controlled studies, studies and objective trials without matched controls, well-documented case histories
conducted by qualified experts, and reports of significant human experience with a marketed device, from which it
can fairly and responsibly be concluded by qualified experts that there is reasonable assurance of the safety and
effectiveness of a device under its conditions of use. The evidence required may vary according to the characteristics
of the device, its conditions of use, the existence and adequacy of warnings and other restrictions, and the extent of
experience with its use.”
6 See FDA’s guidances, including “Design Considerations for Pivotal Clinical Investigations for Medical Devices.”
While this guidance was developed primarily for clinical studies used to support Premarket Approval Applications
(PMAs), the recommendations of this guidance may also be used in designing clinical studies used to support other
types of regulatory submissions.
7 For purposes of this guidance, bias is any systematic error in the design, conduct, analysis, interpretation,
publication, or review of a study and its data that results in a mistaken estimate of a treatment’s effect on disease.
This systematic error results from flaws in the method of selecting study participants, in the procedures for gathering
data, and in the decision of how and whether to publish the results. See JM Last. A dictionary of Epidemiology (3rd
edition). New York: Oxford University Press, 1995, and M Szklo & FJ Nieto. Epidemiology: Beyond the basics.
Gaithersburg, MD: Aspen Publishers, Inc., 2000.
Contains Nonbinding Recommendations
4
related to the use of RWD for regulatory decision-making, consistent with existing approaches to
evaluating uncertainty when making benefit-risk determinations8 and the least burdensome
principles.9 When weighing premarket regulatory decisions, FDA considers the totality of
evidence regarding the probable benefits and probable risks of a device, including the extent of
uncertainty in the benefit-risk information. The appropriate extent of uncertainty regarding a
device’s benefits and risks depends on the type of regulatory decision and its context. Assessing
the relevance and reliability of the RWD, as outlined in this guidance, can help to identify the
degree of uncertainty of the benefits and risks of a device, and—in some instances—can resolve
uncertainty (e.g., as introduced by other sources of evidence), which should be considered during
the benefit-risk determinations for the device for a given regulatory purpose. Similarly, FDA
believes that RWD may provide an efficient means to generate the necessary information to
support a particular regulatory decision, consistent with FDA’s least burdensome principles.
RWD may be used for various purposes in a regulatory submission, including both pre- and post-
market (see Section IV.A). Use of relevant and reliable RWD to generate RWE can benefit
interested parties throughout the device ecosystem, including patients, health care providers,
manufacturers, and FDA. Some potential benefits of using RWD may include:
· Obtaining information from broader clinical experiences than are usually represented in
traditional clinical studies.
· Providing new insights into the performance of a device.
· Improving generalizability.
· Capturing data generated over a longer period of time than would be practical in a
traditional clinical study and potentially allowing for data to be gathered on longer-term
outcomes.
· Evaluating outcomes that may not be feasible in traditional clinical studies.
· Providing information with comparable or even superior characteristics to information
collected and analyzed through a traditional clinical study. In some cases, a traditional
clinical study may be impractical or excessively challenging to conduct (e.g., ethical
issues regarding treatment assignment).
· Facilitating more timely completion of postmarket requirements and commitments.
· Decreasing time to market.
In 2017, FDA issued the guidance, “Use of Real-World Evidence to Support Regulatory
Decision-Making for Medical Devices,” in which we described the RWD relevance and
8 See FDA’s guidances, “Factors to Consider When Making Benefit-Risk Determinations in Medical Device
Premarket Approval and De Novo Classifications;” “Consideration of Uncertainty in Making Benefit-Risk
Determinations in Medical Device Premarket Approvals, De Novo Classifications, and Humanitarian Device
Exemptions;” “Benefit-Risk Factors to Consider When Determining Substantial Equivalence in Premarket
Notifications (510(k)) with Different Technological Characteristics;” and “Factors to Consider Regarding Benefit-
Risk in Medical Device Product Availability, Compliance, and Enforcement Decisions.”
9 FDA has defined “least burdensome” to be the minimum amount of information necessary to adequately address a
relevant regulatory question or issue through the most efficient manner at the right time. The least burdensome
definition and principles do not change the applicable statutory and regulatory standards, such as the device approval
or clearance standards, or the applicable requirements, including premarket submission content requirements and the
requirement for valid scientific evidence. For more information on FDA’s least burdensome provisions, see FDA’s
guidance “The Least Burdensome Provisions: Concept and Principles.”
Contains Nonbinding Recommendations
5
reliability factors that FDA assesses to determine if RWD are sufficient for generating RWE.
Subsequently, on December 29, 2022, the Food and Drug Omnibus Reform Act of 2022
(“FDORA”) was signed into law as part of the Consolidated Appropriations Act, 2023, Pub. L.
No. 117-328. Section 3629 of FDORA “Facilitating the Use of Real World Evidence” directs
FDA to issue or revise existing guidance on considerations for the use of RWD and RWE to
support regulatory decision-making. FDA issued the 2023 draft guidance to propose revisions to
the 2017 guidance, “Use of Real-World Evidence to Support Regulatory Decision-Making for
Medical Devices,” to satisfy the requirement under section 3629(a)(2). This guidance also fulfills
a commitment in Section V.F of the Medical Device User Fee Amendments Performance Goals
and Procedures, Fiscal Years 2023 Through 2027 (MDUFA V).10 This guidance is intended to
provide expanded and updated recommendations to industry and FDA staff for conducting an
assessment of the relevance and reliability of RWD to generate RWE for regulatory decision-
making.
III. Scope
This guidance is applicable for the use of RWE to support regulatory submissions for medical
devices.11
The topics covered within this guidance are framed specifically for the use of RWD and RWE in
regulatory submissions for medical devices (e.g., Investigational Device Exemption (IDE),
premarket notification under section 510(k) of the Federal Food, Drug, and Cosmetic (FD&C)
Act (510(k)), PMA, Humanitarian Device Exemption (HDE), De Novo classification request,
submissions related to post-approval study requirements or postmarket surveillance (in response
to an order issued under section 522 of the FD&C Act (522 order)), Clinical Laboratory
Improvement Amendments (CLIA) Waiver by Applications (CW), Dual De Novo/510(k) and
CW Submissions (Duals)). The considerations included in this guidance may be applicable to
supporting uses of RWD across the medical device total product life cycle (TPLC).
This guidance does not address the use of non-clinical data, adverse event reports, secondary use
of data originally generated from a traditional clinical study, or systematic literature reviews, nor
does it address all possible study design/conduct or analytical methodologies. While it does
describe the factors that FDA considers when evaluating relevance and reliability of RWD, it
does not provide a specific set of criteria or other scoring tools for determining the suitability of
any specific RWD source for generating RWE for a particular regulatory decision.
This guidance should not be construed to alter or change in any way the existing evidentiary
standards applicable to FDA’s regulatory decision-making. Rather, this guidance describes the
circumstances under which clinical evidence generated from RWD may be used to support a
10 For more information, see “MDUFA PERFORMANCE GOALS AND PROCEDURES,
FISCAL YEARS 2023 THROUGH 2027.” The MDUFA V commitment letter directed FDA to provide more clarity
about “Fit-for-Purpose of RWD.” This guidance frames “Fit-for-Purpose” with respect to relevance and reliability,
as that better reflects the intent of the phrase “Fit-for-Purpose.”
11 This guidance does not apply to drugs and biological products. For information on the RWE program for drugs
and biological products, see the FDA webpage, “Center for Biologics Evaluation and Research & Center for Drug
Evaluation and Research Real-World Evidence.”
Contains Nonbinding Recommendations
6
variety of FDA decisions based on the existing evidentiary standards. While FDA encourages the
use of relevant and reliable data to generate clinical evidence, including RWE, this guidance
neither mandates use of RWD and RWE nor restricts other means of providing evidence to
support regulatory decision-making. This guidance does not affect any federal, state, or local
laws or regulations, or foreign laws or regulations that may be applicable to the use or collection
of RWD, or that provide protections for human subjects (including informed consent
requirements) or patient privacy. This guidance should be used to complement, but not
supersede, other device-specific and GCP guidance documents.
IV. Regulatory Context in Which Use of RWE May be
Appropriate
A. General Considerations for the Use of RWE
In general, FDA considers the use of RWD to be appropriate to generate RWE when the RWD
are relevant to and reliable for informing or supporting a particular regulatory decision. It is
important to understand the strengths and limitations of the underlying RWD and how these
qualities impact their relevance and reliability. Similarly, a central part of FDA’s evaluation of
RWE is considering the specific regulatory decision in the context of FDA’s existing evidentiary
standards.
FDA recognizes that RWE can be generated from a variety of RWD sources that are primarily
intended for a purpose other than research. For example, administrative claims data12 are
typically collected for purposes of billing or payment for medical care. Disease-specific RWD
sources may be useful for tracking progression or outcomes of diseases, such as specific rare or
poorly understood diseases. Treatment-specific RWD sources may have several purposes,
including for assessment and tracking of overall outcomes, providing assessment of hospital
operations, informing performance improvement initiatives, or providing risk prediction and
benchmarking data for specific therapies. The suitability of the RWD source may be determined
by the factors outlined in Section V and the availability of sufficient data to address the study
question of interest.
FDA does not endorse one type of RWD over another. Sponsors should select the appropriate
RWD sources based on their suitability to address the specific study questions. Data sources that
may be considered RWD sources, if they include data that are routinely collected, include the
following:13
12 For the purposes of this guidance, “administrative claims data” means claims data that arise from a person’s use of
the health care system and reimbursement of health care providers for that care. Definition adapted from Strom et
al., Textbook of Pharmacoepidemiology 6th ed. 2022, p. 137.
13 Published literature may contain clinical data from a variety of sources, including traditional clinical study data
and RWD. If a sponsor is proposing to use RWD from published literature to support regulatory decision-making,
sponsors should specify the RWD source type (e.g., if a journal article presents a retrospective analysis of EHRs, the
RWD source should be specified as EHR).
Contains Nonbinding Recommendations
7
· Registries14 (e.g., medical device registry);15
· EHRs;
· Administrative claims data;
· Chargemaster and/or billing data;
· Patient-generated data16 (e.g., patient-reported outcomes or performance outcomes) that
are created, reported, or gathered by patients, including in-home use, potentially using
wearables and other DHTs;17
· Device-generated data (e.g., implantable devices, physiological monitoring devices);
· Public health surveillance data (e.g., COVID-1918 case surveillance);
· Clinically annotated biobanks; and
· Medical device data repositories (e.g., imaging, electrocardiography databases).
Some purposes for which use of RWD may potentially be applicable in a regulatory submission
include the following:
· To generate hypotheses to be tested in a clinical study;
· As a historical control, an informative prior in a Bayesian analysis of a clinical trial,19
or as one source of data in a hierarchical model or a hybrid data synthesis;
· As a concurrent control group or as a mechanism for collecting data to support
marketing authorization when a registry, EHR, claims data, or some other systematic
data collection mechanism exists;
14 For the purposes of this guidance, “registry” means an organized system that uses observational study methods to
collect uniform data (clinical and other) to evaluate specified outcomes for a population defined by a particular
disease, condition, or exposure, and that serves one or more stated scientific, clinical, or policy purposes. Definition
adapted from the Agency for Healthcare Research and Quality’s (AHRQ’s) “Registries for Evaluating Patient
Outcomes: A User’s Guide: 4th Edition.” doi: 10.23970/AHRQEPCREGISTRIES4.
15 For the purposes of this guidance, “medical device registry” means an organized system with a primary aim to
increase the knowledge on medical devices contributing to improve the quality of patient care that continuously
collects relevant data, evaluates meaningful outcomes, and comprehensively covers the population defined by
exposure to particular device(s) at a reasonably generalizable scale (e.g., international, national, regional, and health
system). Definition is cited from the IMDRF document “Principles of International System of Registries Linked to
Other Data Sources and Tools.”
16 The use of the term “patient-generated data” is consistent with the use of this term in the “Framework for FDA’s
Real-World Evidence Program” document. Patient-generated data includes patient-generated health data. For the
purposes of this guidance, “patient-generated health data” means health-related data created, recorded, or gathered
by or from patients, family members, or other caregivers to help address a health concern. Definition adapted from
HealthIT.gov’s website Patient-Generated Health Data.
17 For the purposes of this guidance, “digital health technology” means a system that uses computing platforms,
connectivity, software, and/or sensors for health care and related uses. If you are considering using RWD collected
with DHTs, please also see FDA’s guidance “Digital Health Technologies for Remote Data Acquisition in Clinical
Investigations.”
18 In 2019, an outbreak of respiratory disease caused by a novel coronavirus began. The virus has been named
“SARS-CoV-2,” and the disease it causes has been named “Coronavirus Disease 2019” (COVID-19). On January
31, 2020, the Secretary of Health and Human Services (HHS) issued a declaration of a Public Health Emergency
(PHE) related to COVID-19 in accordance with section 319 of the Public Health Service Act (PHS Act) (hereinafter
referred to as “section 319 PHE declaration”) and mobilized the Operating Divisions of HHS. In addition, on March
13, 2020, the President declared a national emergency in response to COVID-19. The section 319 PHE declaration
related to COVID-19 expired on May 11, 2023.
19 For more information on Bayesian trials, see FDA’s guidance “Guidance for the Use of Bayesian Statistics in
Medical Device Clinical Trials.”
Contains Nonbinding Recommendations
8
· As a mechanism for training and/or ongoing monitoring of an artificial intelligence
(AI)-enabled medical device throughout the device’s TPLC;20
· To generate evidence to identify, demonstrate, or support the clinical validity of a
biomarker or clinical outcome assessment;
· To develop algorithms used in a medical device;
· To generate (primary) clinical evidence to support marketing authorization (e.g.,
510(k), PMA, HDE, or De Novo request);
· To generate evidence directly by the subject device to provide new information on
safety or effectiveness;
· To generate evidence to support a determination on whether the subject device meets
the statutory criteria for a CLIA waiver21 (e.g., CW and Duals22);
· To generate evidence to support the interpretability of the primary clinical evidence
(e.g., to demonstrate that the study population for an investigation conducted outside
the United States (OUS) is representative of the U.S. population,23 or to provide
context for an adverse event observed in the clinical study);
· To generate evidence to support a petition for reclassification of a medical device
under section 513(e) or (f)(3) of the FD&C Act;
· To generate evidence for expanding the labeling of a device to include additional
indications for use or to update the labeling to include new information on safety and
effectiveness;24
· To develop a performance goal (PG) or Objective Performance Criterion (OPC) using
appropriate statistical methods, such as a patient-level meta-analysis from a
sufficiently relevant and reliable RWD source. As device technology evolves over
time, an OPC or PG could also be updated using RWD;25
· To generate evidence for postmarket surveillance. Through ongoing surveillance,
signals are at times identified that suggest there may be a safety issue with a medical
device. RWE may be generated using RWD to refine these signals for purposes of
informing appropriate corrective actions and communication;26
20 For additional information and resources on AI, please see the following FDA webpages: Guidances with Digital
Health Content and Artificial Intelligence and Machine Learning in Software as a Medical Device.
21 For more information on CLIA waiver by applications, see FDA’s guidance “Recommendations for Clinical
Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of In
Vitro Diagnostic Devices.”
22 For more information on the Dual 510(k) and CLIA Waiver by Application pathway, see FDA’s guidance
“Recommendations for Dual 510(k) and CLIA Waiver by Application Studies.” Dual De Novo classification
requests and CLIA Waiver by Application for certain devices (see section 3301 of FDORA) should also consult this
guidance.
23 A study to support a premarket submission in the United States should be relevant to understanding the safety and
effectiveness of the device when used in patients in the United States with regard to subject demographics, standard
of care, and practice of medicine. This is important for studies conducted both in and outside of the United States.
For more information, see FDA’s guidance “Design Considerations for Pivotal Clinical Investigations for Medical
Devices.” See also 21 CFR 812.28 for additional information on acceptance of data from clinical investigations
conducted outside the United States to support an IDE or a device marketing application or submission.
24 See FDA’s guidance “General/Specific Intended Use.”
25 See FDA’s guidance “Design Considerations for Pivotal Clinical Investigations for Medical Devices.”
26 See FDA’s guidance “Public Notification of Emerging Postmarket Medical Device Signals (“Emerging
Signals”).”
Contains Nonbinding Recommendations
9
· To conduct post-approval studies that are imposed as a condition of device approval, as
well as to potentially preclude the need for or address postmarket surveillance
requirements (i.e., 522 orders27 and special controls28); and
· To provide postmarket data in lieu of some premarket data, consistent with FDA’s
policy on balancing premarket and postmarket data collection.29
B. Application of Investigational Device Exemption (IDE)
Requirements in 21 CFR 812 to Clinical Studies Using RWD
An approved IDE permits a device to be shipped lawfully for the purpose of conducting
investigations of the device without complying with certain other requirements of the FD&C Act
that would apply to devices in commercial distribution. The purpose of this investigational
exemption, as set forth in 21 CFR 812.1, “is to encourage, to the extent consistent with the
protection of public health and safety and with ethical standards, the discovery and development
of useful devices intended for human use, and to that end to maintain optimum freedom for
scientific investigators in their pursuit of this purpose.” The IDE regulations apply to all clinical
investigations of devices to determine safety and effectiveness, with certain limited exceptions.30
In many cases, an approved IDE is required before initiating an investigation.31
Whether the collection of RWD for a legally marketed device requires an IDE depends on the
particular facts of the situation. Specifically, if the device is being used in the normal course of
medical practice, an IDE would likely not be required. FDA recognizes that in clinical practice
this could include use of a legally marketed device for uncleared or unapproved uses, where the
device is being administered or prescribed under the authority of a health care practitioner within
a legitimate practitioner-patient relationship. If data collection does not impact how the device is
administered, and the administration and follow-up is within the normal course of medical care,
an IDE would likely not be required. For example, analyses of extant RWD (i.e., RWD already
collected) involving the use in medical care of a device that was not within the cleared or
approved indications for use would generally not be subject to IDE regulations. However, similar
to traditional clinical studies, if data are being gathered to determine the safety and effectiveness
of the device, and the process for gathering the data would influence treatment decisions, such
administration of the device would likely not be within the normal course of medical practice
and an IDE may be required. It is important to note that certain follow-up activities performed
for the purpose of research may also warrant submission of an IDE. The following represent
hypothetical examples to illustrate when an IDE application32 to FDA may or may not be
required for a study with RWD.
27 See FDA’s guidance “Postmarket Surveillance Under Section 522 of the Federal Food, Drug, and Cosmetic Act.”
28 See 21 CFR 860.3 for the definition of “special controls.”
29 See FDA’s guidance “Balancing Premarket and Postmarket Data Collection for Devices Subject to Premarket
Approval.”
30 See 21 CFR 812.2(a).
31 See 21 CFR 812.2 generally, for information on the scope of studies subject to the requirements of Part 812, and
see 21 CFR 812.3(h) for the definition of “investigation.”
32 See 21 CFR 812.20.
Contains Nonbinding Recommendations
10
Example 1 - IDE application to FDA required
A sponsor plans to conduct a prospective clinical study for a new use of a legally marketed
device that presents a potential for serious risk to the health, safety, or welfare of the patient. The
hybrid study protocol prespecifies enrollment criteria, investigational device implantation, and
follow-up. To collect safety and effectiveness data, the sponsor will use protocolized data
collection for short-term outcomes, and leverage administrative claims data to monitor longer-
term outcomes including re-admittance or additional treatments, and discharge of study
participants. Administrative claims data are RWD that are routinely collected during the normal
course of clinical care and not originally intended to answer a specific research question. This
study would require an IDE application to be submitted to FDA, because 1) the device is being
investigated for a new use, and enrollment, implantation of the device, and follow-up are being
directed by a research protocol and are not within the normal course of clinical care; and 2) the
study was determined to present a potential for serious risk to the health, safety, or welfare of the
patient, and therefore the device would be considered a significant risk device33 under the IDE
regulations.34
Example 2 - IDE application to FDA not required
The sponsor of a marketed implant plans to study the clinical use patterns of their device,
including off-label uses, to better understand the associated safety and effectiveness profile. To
collect the safety and effectiveness data, the sponsor will leverage extant data generated at least
one year prior to the study initiation from a hospital’s EHR system to assess the preoperative
patient characteristics, operative notes, and postoperative outcomes. The EHR data used for this
study are a type of RWD that are routinely collected during the normal course of clinical care
and were not originally intended to answer a specific research question. This study does not
require an IDE application to FDA, because 1) the use of the device and follow-up for each
patient was determined as part of routine clinical care; and 2) the data used in the study are
routinely collected during clinical care and the process for collecting the data did not influence
treatment or care decisions.
Example 3 - IDE application to FDA not required
The sponsor of a marketed non-implanted device plans to study the clinical use patterns of their
device to better understand the associated safety and effectiveness profile. This is a prospective
study with a research protocol that states that physicians are to treat patients per the normal
course of medical care. The protocol does not specify how patients should be treated or followed
up. Routinely collected data from use of the device are recorded in the hospital’s EHR system,
and for this study those data generated as a result of use of the device in a procedure that is not
within its labeled use are duplicated into a study-specific module. The module includes data on
patient demographics, lab results, adverse events, hospitalizations post-procedure, and
reinterventions. This study does not require an IDE application to FDA, because 1) the use of the
device for each patient was determined as part of routine clinical care; and 2) the data used in the
study are routinely collected during clinical care and the process of collecting the data did not
influence treatment or care decisions.
33 See 21 CFR 812.3(m) for the definition of “significant risk device.”
34 See 21 CFR 812.20(a)(1) for additional information on the requirement for a sponsor to submit an IDE to use a
significant risk device in an investigation.
Contains Nonbinding Recommendations
11
If an IDE application is determined to be required, FDA intends to work with the IDE sponsor to
develop the least burdensome approach to facilitate the efficient generation of RWE. Note that
regardless of the applicability of 21 CFR Part 812, FDA regulations at 21 CFR Part 56
(Institutional Review Board (IRB) review), Part 50 (Protection of Human Subjects), and Part 54
(Financial Disclosure) may apply to RWE generation activities, as may other federal, state, and
local laws regarding human subject protections. Should a sponsor or IRB be unclear regarding
the applicability of any of these regulations for a particular RWD collection activity or use, the
sponsor or IRB can contact FDA at CDRHClinicalEvidence@fda.hhs.gov.
C. Application of RWD from Devices Authorized for
Emergency Use Under Section 564 of the FD&C Act
Section 564 of the FD&C Act provides that FDA may, after the HHS Secretary has made a
declaration of emergency or threat justifying authorization of emergency use (an “EUA
declaration”), authorize the emergency use of an unapproved product35 or an unapproved use of
an approved product for certain emergency circumstances.
The routine clinical use of a device authorized under an EUA, when used within the scope of its
authorization, is not considered to be a clinical investigation (see section 564(k) of the FD&C
Act and Section IV.B for more information on the application of IDE requirements in 21 CFR
Part 812 to the collection of RWD). Clinical data routinely collected from the use of a device
authorized under an EUA may be considered RWD and may be used to support regulatory
decision-making, if determined to be relevant and reliable for the particular study question and
regulatory purpose. Generally, the recommendations in this guidance may apply to RWD from
devices authorized under an EUA. Additionally, Appendix B includes an example of RWD from
a device authorized under an EUA that was used in a subsequent premarket submission.
Device use pursuant to an EUA may lead to additional sources and novel uses of RWD to
support FDA decision-making. We encourage sponsors to consider the recommendations in this
guidance for devices authorized under an EUA (e.g., devices authorized under an EUA during
the COVID-19 pandemic).
V. Assessing Data Relevance and Reliability
To determine the potential suitability of RWD to generate RWE for regulatory decision-making,
the sponsor should conduct a relevance and reliability assessment (see Section VII.C), and when
a sponsor submits such an assessment, FDA intends to review and evaluate that assessment. The
sponsor’s assessment should address relevance and reliability of the RWD source(s) (including
data elements), as well as the study design and analytic components of the study. Sponsors
should understand the strengths and limitations of generating evidence from RWD to address a
specific regulatory purpose and discuss these limitations in their submission. A submission
relying on RWD should describe the relevance and reliability factors listed in Section V.A and
35 Under sections 564(a)(2)(A) and 564(a)(4)(D) of the FD&C Act, an unapproved product is one that “is not
approved, licensed, or cleared for commercial distribution under section 505, 510(k), 512, or 515 of [the FD&C] Act
or section 351 of the [PHS] Act or conditionally approved under section 571 of [the FD&C] Act.”
Contains Nonbinding Recommendations
12
V.B, as applicable.
Whether RWD are sufficiently relevant and reliable for use will, in part, depend on the particular
regulatory decision for which the data have been provided. As part of the relevance and
reliability assessment of an RWD dataset, if multiple RWD sources are used, the submission
should address how each RWD source contributes to the relevance and reliability of the final
dataset.
The data should be accurate, as complete as possible, and of adequate data quality to credibly
address the question at hand. Conducting a clinical investigation in accordance with GCP
provides assurance that the data and results from the clinical investigation are credible and
accurate and that the rights, safety, and well-being of subjects are protected.36 Best practices are
generally well-established for incorporating GCP into the design and execution of traditional
clinical studies. RWD have certain characteristics, including that these data are typically
collected outside of a research setting, such that additional precautions should be considered for
RWD studies to ensure they are similarly credible and accurate, and that appropriate patient
protections are in place. Sponsors should ensure that a submission containing RWD
demonstrates that sufficient precautions were considered during study conduct and that
appropriate patient protections are in place. Sponsors should demonstrate the RWD are credible
and accurate through the relevance and reliability evaluation.
Additionally, sponsors should ensure that RWD were collected using good data management
practices and are sufficiently robust. Sponsors should also consider data related to appropriate
demographic characteristics and relevant covariates of the intended use population.37
The submission should demonstrate that studies using RWD were designed to mitigate potential
bias. FDA recommends that a study protocol and analysis plan be created prior to analyzing
RWD, regardless of whether the RWD are extant or if they are to be collected in the future. An
existing RWD source may have some inherent sources of bias that could limit the relevance or
reliability for drawing causal inferences between medical device exposures38 and outcomes.
To help ensure the relevance and reliability of the source data, FDA recommends sponsors
consider the factors contained in this section. Appendix A includes an example of elements that
FDA recommends sponsors document and have available for inspection, as well as
recommended elements for sponsors to include in the appropriate regulatory submission for FDA
review.
36 See, for example, 21 CFR 812.28(a)(1), which defines good clinical practice in the context of device
investigations conducted outside the United States and references the credibility and accuracy of clinical
investigation data and results.
37 For example, see the following FDA guidances: “Guidance for the Use of Bayesian Statistics in Medical Device
Clinical Trials” and “Evaluation of Sex-Specific Data in Medical Device Clinical Studies.”
38 The exposure is the variable or data element whose causal effect is estimated in a study. For many studies
supporting a regulatory submission, the exposure is the medical device. For some studies, especially those providing
supportive clinical evidence, another exposure may be assessed to provide context for the use, safety, or
effectiveness of the medical device within clinical practice.
Contains Nonbinding Recommendations
13
A. Relevance
Relevance includes consideration of availability, timeliness, and generalizability of the RWD.
When needed information is not available in one data source, sponsors may choose to provide
linkage to other data source(s). Important relevance factors that FDA intends to consider in
determining whether RWD are suitable for generating RWE for regulatory use include the
following:
(1) Data Availability
The RWD should contain sufficient detail to capture the information needed to evaluate the
question being addressed in the target population. If the sponsor determines that the RWD source
is insufficient on its own, the sponsor should consider whether supplemental data sources are
available and sufficient to provide any missing information necessary to address the study
question.
Relevant considerations should include whether the RWD contain information on the following:
· Use of the device or other exposure in the study population. If the RWD source
captures the device identifier (DI)39 portion of the unique device identifier (UDI),40
this should be documented to support robustness of the device identification and
minimize exposure misclassification. If the DI is not captured within the particular
RWD source, alternative methodologies for device identification may be considered
for device identification (e.g., other structured data or clinical notes within the same
RWD source);
· Outcome(s) of interest in the study population;
· Covariates that may impact the exposure or outcomes of interest (e.g., RWD source
contains signs, symptoms, treatments, procedures, diagnoses, patient and family
history, pre-existing conditions, labs, demographics, and results which may be used to
construct covariates that are relevant to the study question); and
For example, the minimum set of data fields in a registry may be
insufficient for a specific study question and additional data fields may
be needed for the registry to be relevant. The registry should retain
information documenting the start or stop of collection during the study
time frame for data fields related to the specific study question.
· Longitudinality, including longevity (the length of time that data for an individual is
captured within the RWD source) and continuity of care.
Information across the continuum of care41 (i.e., data observability)
may aid the assessment of the likelihood that all exposures and
outcomes of interest will be captured for regulatory decision-making.
39 The device identifier is a mandatory, fixed portion of a UDI that identifies the specific version or model of a
device and the labeler of that device. 21 CFR 830.3. For more information regarding UDI, see FDA’s webpage,
“UDI Rule, Guidances, Training, and Other Resources.”
40 For class I devices, the Universal Product Code (UPC) may serve as the UDI (21 CFR 801.40(d)). In these
instances, the UPC should be included.
41 For the purposes of this guidance, “continuum of care” refers to the extent of the individual’s pertinent health
data, which is captured across settings/environments of care that is represented in the RWD source.
Contains Nonbinding Recommendations
14
For example, tertiary care hospitalization data may not have adequate
data availability to study outcomes that are likely to be diagnosed in an
emergency for all patients, because patients are likely to go to a nearby
hospital in emergencies but may travel to another location for a
specialty device procedure.
(2) Linkages
Sponsors should assess whether and how data from different sources can be obtained and
integrated given the potential for heterogeneity in target population characteristics, clinical
practices, and coding across data sources. A description of this assessment should be provided in
the regulatory submission for FDA review.
Any linkages performed within and across RWD sources should use a predefined linkage
methodology42 that is scientifically valid, protects the privacy of individuals (e.g., privacy
preserving record linkage or PPRL) whose data will be used, supports interoperability, and
accounts for differences in coding and reporting across sources. The following considerations
should be assessed by the sponsor and addressed in the submission, as appropriate:
· Adequacy of line-level linkages (i.e., that the same individuals are being matched),
including predefined rules to check for logical consistency and value ranges to
confirm that data were retrieved accurately from a linked data source; and
· Application of strategies to correct for redundant data, to resolve any inconsistencies,
and assess the potential for missing data.
Because patients typically visit multiple health care sites, especially in
geographically contiguous areas, the inclusion of de-identified data
from many sites creates the possibility that there will be multiple
records from different health care sites for a single individual (e.g.,
including when new technology such as patient-generated data start to
be included in the health care record). This can result in overcounts of
a particular data measure. Appropriate documentation, including
justification for how data (e.g., covariates, outcomes) inconsistences
were resolved, should be provided. Alternatively, if some site records
are not available, this can result in a collection of histories that reflect
only a fraction of the patient’s total health care history.
(3) Timeliness
As with traditional clinical studies, the time between data collection and release for research
should be reasonable and the RWD considered for the study should reflect the current clinical
environment (e.g., RWD from before a major change in clinical practice may not be timely).
Sponsors should consider changes in clinical practice and guidelines over time (e.g., criteria for
disease diagnosis, cancer staging), characteristics of a condition (e.g., prevalent strain of a
pathogen), and health status of the population. If data are being collected within the RWD source
42 See “An Overview of Record Linkage Methods - Linking Data for Health Services Research: A Framework and
Instructional Guide - NCBI Bookshelf.” doi: 10.1.1.79.1519.
Contains Nonbinding Recommendations
15
during the study time frame, then the sponsors should specify within the protocol the timing for
RWD updates and should define the reporting schedule in the regulatory submission for FDA
review. Additionally, the sponsor should provide a rationale for the reporting schedule and
whether it is appropriate for the regulatory purpose. When a sponsor provides such information
to FDA, FDA intends to use that information to determine whether the reporting schedule is
appropriate for the regulatory submission.
(4) Generalizability
Once a study question is defined, the study sample that is selected should be adequately
representative of the patients in the RWD source that are reflective of the proposed intended use
population. Identifying the study sample should take into consideration how the application of
inclusion/exclusion criteria will affect the overall generalizability of the RWE to the U.S.
intended use population (see Section VI). If upon quantitative assessment, the study sample is
shown to not be representative of the target population, then analyses should be conducted to
evaluate generalizability of the study findings.
B. Reliability
Reliability includes consideration of accrual, quality, and integrity of RWD. The sponsor should
attempt to obtain all information and associated documentation related to processes, procedures,
and methods around data accrual (see Section V.B.1) and data quality and integrity (see Section
V.B.2), as described below. If the sponsor is not able to obtain this information and associated
documentation, the sponsor should identify the entity(ies) which do have access/permission to
the respective processes, procedures, and methods. The sponsor should then determine the level
of access which could be shared with FDA, as well as the potential for third parties to provide
the needed documentation directly to FDA. The availability of the respective processes,
procedures, and methods should be described in the regulatory submission for FDA review. FDA
recognizes that some data sources will not allow sponsors to obtain all information and
associated documentation. Although FDA does not discourage use of these data sources, we note
that uncertainty may arise if FDA does not have access to all of the necessary processes,
procedures, and methods. FDA notes that the sponsor should conduct the reliability assessment,
with all information that could be obtained, and provide a rationale for why the RWD is reliable
for the regulatory purpose.
Important reliability factors that FDA considers in determining whether the RWD are suitable for
generating RWE for regulatory use include the following:
(1) Data Accrual
To ensure the reliability of the RWD source, data should be collected and processed in a
consistent and methodical manner. The manner of collection may differ for newly developed
RWD sources which are actively collecting data (e.g., data dictionary to provide a common
definitional framework in a registry), using nationally or internationally recognized coding
systems (e.g., International Classification of Diseases, Tenth Revision, Clinical Modification
(ICD-10-CM), Logical Observation Identifiers, Names, and Codes (LOINC), UDI, Current
Contains Nonbinding Recommendations
16
Procedural Terminology (CPT) in EHR or Claims) or custom-designed structured data capture
(e.g., data capture within a device), or using unstructured data capture (e.g., narrative portion of
clinical notes). Any of these approaches may be able to demonstrate sufficient reliability to
support regulatory decision-making. Factors FDA intends to consider in making this
determination include:
· Adequacy of information and descriptors about data sources provided in the
regulatory submission for FDA review, which should include information on:
· Data types;
· Health care settings/environment(s);
· Purpose of data collection;
· How data were obtained at point of data capture;
· How data are accessed by study team and sponsor;
· Any data transformations, including any modifications made for privacy
protection;
· Full data dictionary or common data capture form, if applicable;
· Device information, including types of identifiers (e.g., DI) and indication for
use;
· Completeness of fields that would typically be completed for all individuals
and needed for most study questions (e.g., age, sex, DI);
· Time frame (including common temporal framework for collection of data)
and latency of the data (including the timeliness of data entry, transmission,
and availability);
· Version control of the RWD source(s); and
· Key technical and privacy-related information. Sponsors should document
routine migration of data between various sources over time (e.g., indicate the
date and time of data transfers, linkages).
· Adequacy of information about data accrual methods and procedures provided in the
regulatory submission for FDA review, which should include information on:
· Data collection procedures;
· Use of common data capture forms;
· Common definitional frameworks;
· Data cleaning and cross-referencing procedures;
· The sources and technical methods used for data element capture (e.g., chart
abstraction, point of care entry, EHR integration, UDI capture, data records
from the device, and linkage to administrative claims data); and
· Methods for data retrieval and processes to minimize missing data extraction,
implausible values, and data quality checks in data captured at the point of
care (e.g., during clinical practice for manual or automated health care data
collection processes) to ensure accuracy and completeness of core data fields.
Contains Nonbinding Recommendations
17
(2) Data Quality and Integrity43
When considering RWD sources for regulatory purposes, sponsors should consider the methods
and systems used to help ensure sufficient data quality, including any data quality assurance
plans and procedures developed for the RWD source itself. Because evaluation of RWD sources
may not always permit specific line-item source verification, important factors sponsors should
consider and address in a submission include:
· Quality control processes;
· Regardless of the original purpose for collecting the RWD, procedures for
data collection and quality assurance should be put into place during the data
source design and development stages to optimize the reliability, quality, and
usefulness of the data, as appropriate. These procedures should be described
in the regulatory submission for FDA review.
· Where appropriate, processes should include site and data monitoring, data
quality audit programs, and evaluation of ongoing training programs for data
collection.
· Records regarding the assessment of adherence to the RWD source’s
established data quality assurance and quality control policies and procedures
should be retained.
· Assessment of completeness, accuracy, and consistency across sites and over time;
· Data should be captured in a manner designed to minimize missingness.
Missingness and out of range values should be assessed for each data element.
The amount of missingness per individual (across data elements) should also
be assessed, if possible. The impact of missingness should be considered and
thresholds for unacceptable levels of missingness should be pre-determined.
Additionally, quantitative assessment of the potential bias associated with
high missingness should be performed and included in the interpretation of the
study.
· Data should be reflective of the real-world patient experience (e.g.,
interactions with health care, disease trajectory, outcomes) with the condition
of interest.
· Consistency of data capture should be used across sites and over time.44 If any
changes are needed (e.g., where diagnostic criteria, definitions, or clinical
practice change over several years), then sponsors should document those
changes and assess their impact on the study question and provide summary
information in the regulatory submission for FDA review.
· Auditing rules, methods, and the mitigation strategies used to reduce errors
should be documented.45
43 For more information on registry design and execution to better ensure data quality, sponsors can consult
published literature such as AHRQ’s “Registries for Evaluating Patient Outcomes: A User’s Guide: 4th Edition,”
Section 1 Chapter 3, Registry Design and Section 3 Chapter 11, Obtaining Data and Quality Assurance. doi:
10.23970/AHRQEPCREGISTRIES4.
44 More information on the Patient-Centered Outcomes Research Institute (PCORI) Conduct of Registry Studies is
available in the following document “Standards in the Conduct of Registry Studies for Patient-Centered Outcomes
Research.”
45 For more information on documentation for FDA review, see Section VII and Appendix A.
Contains Nonbinding Recommendations
18
· Study sample size46 should be justified to address the study question and
regulatory purpose. It should be determined based on adequate statistical
power or precision to detect a clinically meaningful difference.
· If extant RWD are used, the number of individuals may not be
adequate to meet the target sample size.
· If the number of available RWD individuals is not large
enough to meet the target sample size; sponsors should
consider the use of multiple existing RWD sources.
· If the sample size could be expected to increase in the near
future (e.g., device is new to market), sponsors should consider
accruing more data.
· Sponsors should account for planned statistical analysis within the
study size calculations (e.g., 1:1 matching of propensity scores in a
study population where 10% of individuals receive the device would
remove approximately 89% of individuals with comparator from the
analysis).
· Establishment and adherence to data collection, recording, and source verification
procedures;
· As with all clinical evidence generation, data provenance and traceability are
important. Sponsors should plan and document all aspects of data extraction,
aggregation, curation, storage, and availability for research, as described
below.
· Sponsors should ensure any automated electronic transmission of data fields
to a repository (e.g., registry or data warehouse) occurs in a secure, consistent,
and reproducible fashion, including applicable processes.
· Adherence to source verification procedures and data collection and recording
procedures should be documented for completeness and consistency.
· Data checks and procedures should be prespecified to help address identified
errors (e.g., in coding or interpretation of the source documentation or
transformation).
· Sponsors should describe the mitigations used to address audit findings,
including data corrections.
· Sponsors should identify the source document(s) and first instance47 of data
available to sponsor. Sponsors should generate data quality documentation
from the first instance through RWD dataset(s) used to address the study
question.
46 For the purposes of this guidance, sample size is considered part of completeness/accuracy of the RWD to address
the study question and is considered a factor of reliability under data quality and integrity. Outside of this guidance,
sample size may be considered a factor of relevance. Regardless, sample size is generally included in the relevance
and reliability assessment.
47 For the purposes of this guidance, the “first instance” is considered to be the data as initially available to the
sponsor. For example, if the raw data from an EHR are available, then the first instance is at the point of capture for
the data element. However, if the sponsor only has access to curated data or a specific dataset, then the “first
instance” is the initially ingested data by the sponsor, even though it is not the original data collected. FDA
acknowledges that in some cases the “first instance” dataset may be the same as the curated dataset.
Contains Nonbinding Recommendations
19
· If using a common data model, sponsors should ensure documentation of the
transformation of data from the original source to the common data model is
retained.
· Data audit trail, including assessment of discrepancies, should be included, as
applicable.
For extant data sources, the sponsor may have access to the
initial capture of data (e.g., direct access to EHR data as
obtained via data entry) or may only have access to a partially
curated data source (e.g., administrative claims data,
aggregated EHR). We recommend sponsors maintain
information on the data audit trail from the first instance of data
available to the sponsor through all aspects of data analysis.
Further, sponsors should obtain as much information as
possible about the audit trail from the data holder if the first
instance is not at the point of data entry.
· As with all clinical evidence generation, FDA recommends that the sponsor
have access to the RWD dataset from the first instance used for the analyses
throughout the regulatory decision-making process. When the RWD source is
not owned by the sponsor, where participant-level data is accessible, the
sponsor should attempt to obtain participant-level data for each participant. If
not available, the sponsor should define the entity(ies) which do have
access/permission for data entry, quality assurance, storage, aggregation or
other linkage, and assessment of traceability from data entry to dataset, as
applicable. Sponsors should consider the level of access which could be
shared with FDA and the potential for third parties to provide participant-level
data directly to FDA. The availability of data should be described in the
regulatory submission for FDA review. FDA recognizes that some data
sources either provide only aggregate-level outputs or will not allow sponsors
to access the participant-level data. FDA does not discourage use of these data
sources, and a sponsor’s inability to obtain participant-level data does not
preclude FDA’s evaluation. Sponsors should describe how lack of participant-
level data does or does not affect the RWE.
· Adequate patient protections (e.g., methods to protect the privacy of individuals’
health data and adherence to applicable privacy and ethics standards) established in
advance of executing the study protocol; and
· Prior demonstration of RWE generation from the data source.
Sponsors should provide documentation (including the relevance and
reliability assessment) of any previous use of the same RWD source for a
similar target population and/or peer-reviewed literature of RWE generation
from the data source, as applicable.
VI. Considerations for Methodologies for Collection and
Analysis of RWD to Generate RWE
A study using relevant and reliable RWD in a well-designed and rigorously analyzed manner
Contains Nonbinding Recommendations
20
may be less burdensome than a traditional clinical study. Just as traditional clinical studies
should be carefully designed, studies using RWD also should undergo careful assessment before
embarking on and during development of the study to ensure that the data are relevant and
reliable for the study question and regulatory purpose. FDA recognizes that some regulatory
decisions may not be adequately supported using RWE, for various reasons, and we therefore
recommend that sponsors consider methodologies to address factors that can impact
interpretability of a study using RWD.
Scientifically sound clinical study planning in advance of statistically valid analyses is important
regardless of whether a study uses a traditional clinical study approach, uses only RWD, or
incorporates a hybrid design. Further, just like for traditional clinical studies and in addition to
the study design and analysis considerations described in Sections VI.A and VI.B, FDA
recommends that a sponsor document their decisions and the associated rationale for the
following:
· Whether to include randomization, concurrent, or historical controls;
· The choice of performance goals and objective performance criteria;
· Type I and type II error control;
· Data gathering or dependence on extant data;
· Bias mitigation strategies;
· Precision of outcome measures and other data elements, as applicable; and
· All other known factors pertinent to interpretation of the study results (e.g.,
generalizability of the RWE findings to the intended use population).
Although many of the considerations in this section for data collection and analysis are not novel
in the context of clinical evidence generation, there may be unique aspects of these
considerations for studies using RWD. Additionally, the information presented in this section is
intended to augment, not replace, information in other FDA guidances on the design of clinical
studies for regulatory decision-making. The information in this section is intended to clarify
implementation of these concepts and practices when using RWD. In particular, the information
below is intended to complement the recommendations for clinical studies in the FDA guidance
“Design Considerations for Pivotal Clinical Investigations for Medical Devices.”
A. Methods for Study Design Using RWD
Generally, FDA does not endorse a specific type of study design for clinical studies, regardless
of whether it is a traditional clinical study or uses RWD. As with all clinical evidence generation,
choosing the appropriate design for studies using RWD depends on the study question, device,
outcome, key covariates, and the specific study objectives or hypotheses. Additionally, sponsors
should consider the regulatory purpose of the generated clinical evidence. FDA recognizes that
multiple types of study designs may also be useful to generate RWE. These study designs may
include:
· Single-arm studies with comparisons to external controls, objective performance
criteria, or performance goals, in whole or part;
· Non-interventional studies (observational studies) (e.g., comparative cohort
Contains Nonbinding Recommendations
21
studies, case-control studies, self-controlled studies, and descriptive studies); and
· Randomized controlled trials using RWD to supplement one or more study arms.
B. Defining Study Design Elements
For studies using RWD, as with all clinical studies, after determining the overarching study
design, the study time frame and collection of data elements should be defined, followed by a
system to capture specific data elements (e.g., in a case report form). Additional data capture
requirements may necessitate justification or adjudication, especially for study endpoints. FDA
recommends clearly defining the individual data elements derived from the RWD source to
develop study-specific RWD. Similarly, FDA recommends that sponsors show that the data
elements, as defined and applied within the study design, are relevant and reliable for the
particular regulatory decision. For analysis of RWD and interpretation of RWE, sponsors should
have a study design that describes the study time frame, the predefined set of data elements, and
a systematic consideration that the proposed data elements are all necessary for inclusion and
represent all the key data elements.
(1) Study Time, Relative to Index Date
In traditional clinical studies, a participant is often enrolled into the study, has a baseline visit,
and then first uses the device or has a procedure on the “index date” (see Figure 1). After that,
the participant is usually followed for a period of time until a final visit. Data elements are
collected at each visit, although different information may be gathered at each visit, and
additional data elements may be collected outside of clinical care (e.g., via patient-generated data
such as a participant diary or wearable). The participant continues to be followed through a last
study visit. If a visit or other data collection is missed, then the participant may be contacted or
additional questions may be asked at the next visit to gather key information.
Figure 1. Traditional clinical studies - example of study time frame relative to index date
In clinical studies using RWD where data are collected after the study is designed, it is possible
that a similar visit structure and data collection will be available (e.g., within a registry).
However, follow-up visits may not occur on a set schedule or more patient-generated data may
be collected (e.g., via a patient diary or wearable). For extant data such as EHR or administrative
claims, baseline and follow-up data may not always be collected on a set schedule; rather data
collection coincides with clinical care over a period of time (see Figure 2). Individuals may also
enter or exit the source database as their life situation changes (e.g., move out of a geographic
area or a change in health insurance). Thus, continuity of care is an important consideration for
studies using RWD.
A visual depiction, such as that exemplified in Figure 2, may be helpful in identifying the timing
Contains Nonbinding Recommendations
22
of collection for each data element relative to the index date, which will help to identify potential
bias. Data elements that may impact the initial device use should be collected before or at the
time of initiation. Outcomes of device use occur after use of the device is initiated. Additionally,
the index date for the use of the comparator to the device would occur at a similar point in the
progression of disease. As with a traditional clinical study, discussion of these ideas with the
study team or with FDA may be aided by the visual depiction of when each data element will be
assessed.
Figure 2. Clinical study using RWD - example of study time frame relative to index date
Follow-up in a study using RWD typically extends from the index date of device use until either
the end of the pre-planned follow-up time or the last time identified within the RWD source.
FDA considers the study end date to be the last date that participant follow-up could occur. This
date is set on a day when data checks/audits can ensure that the underlying data are of sufficient
quality for use in research. Any data in the RWD source indicating that a participant had
subsequent care is no longer included in the study (i.e., study participation is censored on this
date). Thus, similar to the first site being ready for enrollment in traditional clinical studies, FDA
anticipates that the study time frame will be defined to begin on the earliest date that the first
data element could be collected and extend through the latest date that the last data element could
be collected.
In addition to the study design elements discussed above, any change in the standard of care,
availability of the device or other treatments, or other relevant factors (e.g., change in hospital
care due to a PHE) should be included on the graphical depiction. This additional information
may aid in systematically capturing these time-dependent data elements and provide support for
their inclusion as covariates in analyses or consideration of sensitivity analysis (e.g., assessing
whether a change in ICD-CM coding from a prior edition or a major change in clinical practice
affects study results).
The amount of observed time for the individuals within the study population should be long
enough to adequately measure all data elements in a study – from the beginning of the baseline
period through the end of the follow-up needed to assess the outcome(s) of interest – in a sample
large enough to provide adequate statistical power to detect the minimal clinically important
difference in the primary outcome.
Contains Nonbinding Recommendations
23
(2) Development of Conceptual and Operational Definitions for
the Study Population, Device, Comparator, Outcome, and
Covariates
As with any clinical study, all data elements should be defined before the start of a study using
RWD and should address the specific study question when valid and appropriate analytical
methods are applied (i.e., the data are amenable to sound clinical and statistical analysis). A
“conceptual definition” that describes the construct or feature of each data element in general or
quantitative terms should be generated using a shorthand name or notation. This conceptual
definition should reflect the current medical and scientific thinking regarding the variable of
interest, such as: (1) clinical criteria to define a condition for population selection or as an
outcome of interest or a covariate; or (2) measurement of the device or procedure to define an
exposure of interest. For example, a conceptual definition might be “acute myocardial infarction
(AMI)” or “AMI evidenced by increased troponin.” For a traditional clinical study, the sponsor
defines the collection and timing of each data element, whether at a visit or between visits, and
usually has the ability to contact the participant to limit missing data or to solicit additional
information if a visit was missed. In a study using RWD, the data elements may be collected in a
similar fashion (e.g., registry) or need to be defined from clinical care visits (e.g., EHR or
administrative claims data) or some other algorithm (e.g., combining unstructured EHR and
patient-reported data).
An “operational definition” describing all of the components needed to identify complete and
accurate data elements from the data source should also be generated. While an operational
definition would typically be generated in a case report form in a traditional clinical study,
operational definitions in a study using RWD frequently include combining structured codes or
unstructured notes (e.g., clinical notes) in an algorithm to identify presence of the data element.
FDA considers the operational definition to include three components, as applicable:
· Time frame over which assessment occurs;
· Specific codes/component(s) assessed (e.g., via code lists); and
· Algorithm for combining the components (leading to positive identification or lack of
identification, including assumptions underlying how event and no event occurrence
are defined). If machine learning is used to define criteria, sponsors should provide a
full description of data management practices including the specifications of the
model/algorithm (e.g., training, tuning, and testing), data collection and the data
attributable to the proposed intended use population (e.g., with respect to race,
ethnicity, disease severity, sex, age, socioeconomic characteristics). Sponsors should
also provide information that demonstrates the machine learning approach is
appropriate for defining the criteria, such as verification and validation of the
machine learning approach.
The availability of different data types in studies using RWD may make it possible to establish
operational definitions that are different from those typically used in traditional clinical studies.
These definitions may or may not be appropriate in the context of the study question being
addressed depending on the study question and regulatory purpose. It is important to consider
whether the operational definition will capture the intended concept for each data element and
Contains Nonbinding Recommendations
24
FDA notes that small differences in the choice of operational definition in a specific data source
(e.g., requiring two diagnoses rather than one diagnosis of AMI in the example above) may have
a large impact on study results (e.g., considerably decrease the identification of the disease or
condition under study). These data elements should be determined before conducting study
analysis(es), with the exception of processes for developing operational definitions from
conceptual definitions.
As with traditional clinical studies, in choosing existing operational definitions or developing
new ones, sponsors should maximize identification of those who have the condition and to
minimize incorrectly identifying those without the condition as having it (i.e., minimize
misclassification). FDA considers minimizing misclassification to be a critical part of the process
of defining an operational definition. FDA recommends reviewing previous studies using the
RWD source, including published literature, and gathering expert opinion when developing
operational definitions. For some data elements, a rationale for the operational definition based
on previous studies or expert opinion may be sufficient. Some data elements may warrant more
scrutiny to ensure that the interpretation of study results is not substantially impacted by their
misclassification or missingness. Further exploration is recommended for data elements that are
not aligned with expectation.
In some cases, it may be appropriate to conduct a validation study in which quantitative
measurements of the operational definition are compared to a “ground truth” reference standard.
This may result in updating the operational definition to ensure that these critical data elements
are accurately identified. When conducting a validation study, a protocol should be developed
before initiating the data collection and analysis specific to the validation. The protocol generally
contains the plan to compare the operational definition in the RWD (e.g., administrative claims
data) with the “ground truth” in the reference standard (e.g., validating that administrative billing
diagnosis accurately represents a point-of-care diagnosis by comparing an operational definition
in administrative claims against an EHR) and prespecification of the acceptance criterion for
each validation measure (e.g., sensitivity, specificity, or positive and negative predictive values)
that is of interest.
In some RWD sources, the data elements that would be preferred for traditional clinical studies
may not be available to the sponsor. However, a proxy for this missing information could be
developed based on the information that is collected in the RWD source. Proxies can be
developed for a wide range of uses, including identifying study participants (i.e., applying
inclusion/exclusion criteria) as well as certain study endpoints. While it may be possible to use
the proxy, sponsors should determine the suitability of a proxy by considering whether the proxy
is clinically relevant and may call for additional data gathering or conducting validation of the
proposed operational definition for the data element, if using the proxy adds too much
uncertainty to the study interpretation. FDA encourages use of measures that participants or
practicing clinicians deem meaningful as potential data elements for studies using RWD.
Development of endpoints or potential consideration of proxy outcomes may be warranted to
address some study questions. Additionally, development of proxies for key covariates may also
be appropriate to address some study questions.
The conceptual and operational definitions should be included in the study protocol (see Section
Contains Nonbinding Recommendations
25
VII.B)
(3) Appropriate Integration of Data Elements Within Study
Design and Analysis
Once the device and outcome are determined per the study question, variables associated with
the device or outcome (i.e., confounders, mediators, effect modifiers, and additional risk factors)
should be addressed within the study design (as both conceptual and operational definitions) and
analysis to minimize bias and uncertainty. Confounders (unobserved and observed) impact both
the device and outcome, and may need to be conditioned upon to accurately measure the causal
effect of the device on the outcome. Mediators, or intermediate variables, are impacted by the
device and subsequently impact the outcome, and should be carefully considered, as conditioning
on such variables may dilute the total effect of the device on the outcome or introduce bias.
Effect modifiers or effect modification refers to heterogeneity of effect of the device on the
outcome by strata of another variable (i.e., modifier), and reporting stratum-specific results may
be appropriate. As with all clinical studies, prespecified subgroup analyses for sex, age, and other
groups are recommended because there may be differential effects of the device on the outcome
by subgroups.48
Based on the prespecified research question(s) identified, the study protocol should concisely
identify which data elements fall into each of the types of variables (e.g., exposure/device;
outcome(s); covariates: confounders, mediators, effect modifiers, risk factors). An effective way
to identify which data elements fall into each of these types of variables is to generate and
analyze causal diagrams. Causal diagrams and their subsequent assessment may provide a
rationale for the design and analysis choices, and therefore should also be provided in the study
protocol. One example of a type of causal diagram is a directed acyclic graph (DAG), as depicted
in Figure 3. Additionally, causal diagrams may provide a resource to aid discussions for selecting
an appropriate study design amongst the study team or with FDA. Covariates affecting or
affected by the exposure or outcome should be noted within the causal diagram, irrespective of
availability within RWD, and assessed for potential relationships among other variables. Causal
diagrams illustrate the relationships between covariates, including the direct causal path from the
exposure to the outcome, indirect causal paths through mediators, and non-causal (backdoor)
paths through confounders. Causal diagrams can also be used to identify a minimally sufficient
set of confounders for which to control so that all non-causal paths between the exposure and
outcome are blocked, without conditioning on variables that are influenced by both the exposure
and outcome (which would introduce collider-stratification bias).49 Thus, potentially only a
subset of the covariates initially identified may be needed to control for confounding.50
48 See FDA’s guidances, including, “Evaluation of Sex-Specific Data in Medical Device Clinical Studies.”
49 For further information on collider-stratification bias, sponsors can consult published literature such as: Weiskopf
NG et al., Healthcare utilization is a collider: an introduction to collider bias in EHR data reuse, Journal of the
American Medical Informatics Association 2023, 30(5): 971-977. doi: 10.1093/jamia/ocad013
50 For further information on the construction and logic of DAGs, sponsors can consult published literature such as:
Merchant AT and Pitiphat W; Directed acyclic graphs (DAGs): an aid to assess confounding in dental research,
Community Dentistry and Oral Epidemiology 2002, 30: 399–404. doi: 10.1034/j.1600-0528.2002.00008.x;
Akinkugbe AA et al., Directed Acyclic Graphs for Oral Disease Research, Journal of Dental Research 2016, 95(8):
853-859. doi:10.1177/0022034516639920.
Contains Nonbinding Recommendations
26
Figure 3. Example of directed acyclic graph to identify potential data elements and assess
which are key for the study question
VII. Documentation for FDA Review
This section describes the documentation recommended to support the use of the RWD for
generating RWE for regulatory purposes and applies to device regulatory submissions submitted
to CDRH and CBER, including but not limited to, pre-submissions, 510(k)s, PMAs, Biologics
License Applications (BLAs), HDEs, De Novos, IDEs, post-approval study PMA supplements,
submissions in response to a 522 order, CLIA Waiver by Applications, and Duals.
A. Regulatory Submission Cover Letters
FDA recommends sponsors identify RWD and RWE as part of the regulatory submission cover
letter to help facilitate review and internal tracking. Specifically, FDA recommends sponsors
include the following in the cover letter for each submission that includes RWD:
· Purpose of using RWE to support the submission (see list of examples in Section
IV.A);
· Study design (i.e., type of study) using RWD to generate RWE (see list of examples
in Section VI.A);
· RWD source(s) used to generate RWE (see list of examples in Section IV.A); and
· Specific RWD source(s) and version, including the following information, if
applicable:
· Data source name;
· Data source provider;
· Version number; and
· Date of extraction and date range of data extracted.
B. Protocol
As with traditional clinical studies, sponsors should include the protocol(s) for any RWD-based
studies as part of the regulatory submission to FDA. FDA recommends that sponsors finalize the
protocol and analysis plan prior to reviewing the outcome data of a study and before performing
Contains Nonbinding Recommendations
27
prespecified analyses. Sponsors should indicate in the regulatory submission whether or not the
protocol and analysis plan were finalized prior to the analyses. While the protocol is being
developed, sponsors should not have access to the outcome measure results. Beyond protocol
development, it is also important to consider when individuals should not have access to
outcomes in the dataset used for the study (e.g., generating summary scores, propensity score
modeling, determining level of data missingness). Any revisions to the protocol should be dated
and time-stamped, and the rationale for each change should be provided.
Sponsors should implement the relevance and reliability concepts in this guidance when
developing their study protocol. Similar to protocols submitted with traditional clinical studies,
sponsors should provide the following information, as appropriate, and avoiding redundancies, in
the protocol for the study generating the RWE, when the RWD or RWE is included in the
regulatory submission:
· Study synopsis;
· Background and study purpose, including:
· Explanation of how the source data is or is not representative of the general
disease/population with the condition, including sufficient previous research
to interpret study results within the context of the target population, disease
trajectory, and current clinical care;
· Description of the device included in the study, including the DI portion of the
UDI, if available. For devices excepted from the UDI requirements, sponsors
should include the version(s) of the device; and
· Reports of prior use of the device relevant to the study;
· Study aims and objectives;
· Study design, including study period (see Section VI.A);
· Study design diagrams are suggested to clarify (1) potential study entry dates within
study period and (2) assessment of all other data elements in relation to cohort entry
or index date;
· Data source(s), including a description of how the setting/environment(s) of data
capture provided adequate continuity of care (see Section V);
· Identification of any common data model structure used for housing the RWD
source or for transformed study-specific RWD, if applicable; and
· Identification of linkage methodology from original RWD sources and
associated attributes that should be considered;
· Data elements (conceptual and operational definitions for all), including:
· Determination of initiation, continuance, and discontinuation of device
exposure, if applicable (see Section V.A);
· Study population, including inclusion and exclusion criteria, and
generalizability to the intended use population;
· Device and comparator, if included in the study;
· Outcome and endpoints; and
· Covariates;
Contains Nonbinding Recommendations
28
· Causal diagram;
· Statistical/data analysis plan;
· Data management and quality control plans (see Section V.B);
· Sample size and statistical power;
· Description of human subject protections, as appropriate, including informed consent,
IRB determination, deidentification plan (e.g., to remove participant identifiers from
patient-generated data or device-generated data), and data confidentiality plans;
· Plans for adverse event reporting;51
· Milestones and timeline;
· Auditing and monitoring plans, as applicable;
· If validation or adjudication of data element(s) were conducted (see Section VI.B),
sponsors should include the study plan and results for validation or adjudication; and
· A copy of the data dictionary used, if one was used or developed.
C. Report Including the Relevance and Reliability
Assessment
As with traditional clinical studies, sponsors should submit a study report for any RWD-based
studies with the study results, discussion, and conclusion as part of the regulatory submission to
FDA. If sponsors include RWE in regulatory submissions, they should include how the RWE
supports the purpose of the submission and their relevance and reliability assessment of the
RWD to generate RWE with the following:
· An assessment of the key relevance and reliability factors for the study using RWD
(see Sections V and VI), which may include:
· Data availability, linkages, timeliness, and timing of data availability,
generalizability of data (see Section V.A);
· Data accrual, quality, and integrity, date of data extraction (see Section V.B);
· Completeness and accuracy of study sample reflecting the target population,
study design and planning (see Section V.B.2);
· Date of data extraction (see Section V.B); and
· Study purpose, specific data elements, and assessment of confounding (see
Section VI.B).
In addition to the relevance and reliability assessment of the RWD, we recommend that the
sponsor provide the following contextual information, including:
· All elements from the protocol, updated to reflect how the study was conducted;
· Justification that any changes or modifications to the protocol did not affect the validity
of the resulting RWE;
51 See 21 CFR Part 803. For additional information, please refer to FDA’s webpage Medical Device Reporting
(MDR): How to Report Medical Device Problems.
Contains Nonbinding Recommendations
29
· Discussion of the occurrence and rationale for any protocol deviations;
· A description of how the RWE is being used within the context of all the clinical
evidence submitted to FDA; and
· If unique considerations exist for the specific RWD source, sponsors should describe
these considerations and how they impact the overall assessment of the data.
D. Additional Information
As with traditional clinical studies, sponsors should also provide the following information in
regulatory submissions that include RWE:
· ClinicalTrials.gov National Clinical Trial (NCT) Number,52 if applicable;
· Informed consent and IRB documentation, as applicable;
· Initial and continuing IRB review and approval; and
· Initial and approved changes to informed consent;
· List of investigational sites, if any, including mailing address, contact information,
and investigator name; and
· Case Report Form templates, as developed by the sponsor, if applicable (e.g., these
templates may be helpful if EHR or claims data are not mapped to the dataset from
the source).
52 ClinicalTrials.gov assigns a unique NCT number to each clinical study registered on their webpage. See
https://clinicaltrials.gov/ for more information.
Contains Nonbinding Recommendations
30
Appendix A. Recommended Relevance and Reliability
Elements for Documentation and FDA Review
The following is an example of recommended relevance and reliability elements, and is intended
to assist sponsors and FDA in determining the relevance and reliability of RWD for a regulatory
decision. The tables below summarize the recommended elements identified throughout the
guidance for sponsors to document and provide in FDA submissions, and the recommended
locations for where to include this information. The tables are not intended to serve as either a
mandatory or exclusive
checklist. Rather, the tables provide a simplified summary of the key elements that sponsors
should use to assess relevance and reliability. Depending on the regulatory submission,
additional considerations may be applicable as noted in Section V, even if not specified below.
Table 1- Recommended RWD Relevance Elements for Submission of RWE
Item (Linked to Section V) Information
for Sponsors
to Document
(e.g., to
make
available for
inspection)
Information
for Sponsors
to Provide to
FDA in
Submission
Recommended
Location in FDA
Submission
Determine if the RWD contains sufficient
detail to capture data elements and address
the study question
x (detailed) x (rationale) Protocol
(rationale for
study question
and data element
definitions)
Assess longitudinality of data source x Protocol
Assess continuity of care in data source x Protocol and
report
Ensure reasonable time between data
collection and release for research
x Protocol and
report
Consider changes in clinical
practice/guidelines over time
x Protocol
Specify timing for RWD updates x Protocol and
report
Assess whether and how data from different
sources can be obtained and integrated,
given the potential for heterogeneity in
population characteristics, clinical
practices, and coding across data sources
x
Protocol
If applicable, use of a predefined linkage
methodology that is scientifically valid and
accounts for differences in coding and
reporting across sources
x (detailed) x (high level) Protocol
Contains Nonbinding Recommendations
31
Item (Linked to Section V) Information
for Sponsors
to Document
(e.g., to
make
available for
inspection)
Information
for Sponsors
to Provide to
FDA in
Submission
Recommended
Location in FDA
Submission
Assess adequacy of line-level linkages x Report
Apply strategies to correct for redundant
data, to resolve any inconsistencies, and
assess the potential for missing data
x
Demonstrate interoperability of the linked
data systems
x
Demonstrate appropriate patient privacy
protections are applied (e.g., PPRL) with
linked data
x Protocol and
report
Justify how the study sample selected is
adequately representative of the patients in
the RWD source that are reflective of the
proposed intended use population.
Identifying the study sample should take
into consideration how the application of
inclusion/exclusion criteria will affect the
overall generalizability of the RWE to the
U.S. intended use population.
x Protocol and
report
Contains Nonbinding Recommendations
32
Table 2- Recommended RWD Reliability Elements for Submission of RWE
Item (Linked to
Section V)
Information for
Sponsors to
Document (e.g., to
make available for
inspection)
Information for
Sponsors to
Provide to FDA in
Submission
Recommended
Location in FDA
Submission
Provide all information
and associated
documentation related
to processes,
procedures, and
methods around data
accrual, data quality,
and integrity; if unable
to do so, describe level
of access obtained, and
consider providing
access for FDA
x (detailed) x (high level) Protocol
Provide information
and descriptors about
data source(s)
x Protocol
If applicable, describe
defined processes
around data monitoring
and programs for
quality audit and on-
going training of data
collection
x
Document routine
migration of data from
various sources over
time
x
Describe sources and
technical methods used
for data element
capture
x
Describe methods for
data retrieval and
processes to minimize
missing data
extraction, implausible
values, and data quality
checks
x
Contains Nonbinding Recommendations
33
Item (Linked to
Section V)
Information for
Sponsors to
Document (e.g., to
make available for
inspection)
Information for
Sponsors to
Provide to FDA in
Submission
Recommended
Location in FDA
Submission
If not data owner,
describe level of
access, attempt to gain
participant-level data,
if accessible, and
consider providing
access for FDA
x Protocol
Describe the quality of
the data captured
x (detailed) x (high level) Report
Plan and document the
process of extraction,
aggregation, curation,
storage, and
availability of data for
research
x
Generate data flow and
supporting data quality
documentation from
the first instance
through RWD
dataset(s) used to
address the study
question
x
Define and follow
procedures for quality
control, including data
collection and quality
assurance
x
Provide assessment of
completeness,
accuracy, and
consistency across sites
and over time
x Report
Contains Nonbinding Recommendations
34
Item (Linked to
Section V)
Information for
Sponsors to
Document (e.g., to
make available for
inspection)
Information for
Sponsors to
Provide to FDA in
Submission
Recommended
Location in FDA
Submission
Assess consistency of
data capture across
sites and over time; if
any changes are needed
(e.g., diagnostic
criteria, definitions, or
clinical practice change
over several years),
then document those
changes and assess
their impact on the
study results
x (detailed) x (high level) Report
Assess missingness
and out of range values
for each data element
x Protocol and report
Describe how data
elements captured and
included in the study
are reflective of the
real-world patient
experience (e.g.,
interactions with health
care, disease trajectory,
outcomes) with the
condition of interest
x Protocol and report
Document the auditing
rules and methods used
and the mitigation
strategies used to
reduce errors
x
Justify how study
sample size addresses
the study question and
regulatory purpose
with adequate
statistical power or
precision for planned
analyses
x Protocol and report
Contains Nonbinding Recommendations
35
Item (Linked to
Section V)
Information for
Sponsors to
Document (e.g., to
make available for
inspection)
Information for
Sponsors to
Provide to FDA in
Submission
Recommended
Location in FDA
Submission
Document adherence
to source verification
procedures and data
collection and
recording procedures
for completeness and
consistency
x
Prespecify data checks
and procedures to help
address identified
errors
x
Describe mitigations to
address audit findings,
including data
corrections
x
Provide documentation
of any previous RWD
source relevance and
reliability assessment
for a similar target
population and/or peer-
reviewed literature of
RWE generation from
data source, as
applicable
x Protocol
Ensure adequate
patient protections
(e.g., methods to
protect the privacy of
individuals’ health data
and adherence to
applicable privacy and
ethics standards)
established in advance
of executing the study
protocol
x Protocol and report
Contains Nonbinding Recommendations
36
Appendix B. Examples Where RWE is Used
Most of the following examples are generalized from actual uses of RWD in support of FDA
regulatory decision-making. These examples do not represent a comprehensive list of all
potential uses or sources of RWD but do describe some situations where RWE might be used to
support regulatory decision-making. These examples also include high-level discussion of some
considerations for relevance and reliability, rather than the more detailed information that should
be provided in submissions, as recommended in Sections V-VII, and Appendix A. For additional
examples of RWE used in regulatory decisions, see the following FDA document: Examples of
Real-World Evidence (RWE) Used in Medical Device Regulatory Decisions. Should a sponsor
wish to use RWD to generate RWE for similar regulatory purposes as described below, the
acceptability of the use of RWD would depend on the specifics of the submission, such as the
device, indications, and results of analyses.
Example 1: New approval of PMA
An implanted device, available OUS, used RWE as the primary clinical evidence to support the
original PMA submitted to FDA. RWD from an OUS registry in one country, which included
hundreds of patients with more than two years of follow-up, was compared against performance
goals (PG) for device safety and effectiveness. In this example, an IDE application was not
required for the clinical study because it was conducted OUS. An IDE application was also not
required because the RWD comprised routinely collected extant data from an OUS registry. The
PGs were derived from a prospectively defined meta-analysis of published literature and data
from a U.S. registry of a comparator device legally marketed in the U.S. The study evaluated
routinely collected functional outcome data and patient reported outcome measures. The safety
assessment also included a comparison of the serious device-related adverse event rates for the
subject device extracted from the OUS registry to rates in the same literature studies used to
derive the PG. These analyses served as the primary basis supporting approval of the PMA.
Important considerations from FDA’s review, supporting the relevance and reliability of the
RWD in this submission, included, in part: 1) devices of interest were clearly identified in the
OUS registry, 2) data from the OUS registry were generalizable to U.S. patients, and 3) the OUS
registry had adequate data capture methods to ensure completeness, accuracy, and consistency
across sites of key variables, including covariates of interest and outcome measures.
Example 2: Expanded indications for use
An implanted device was authorized under a PMA and was subsequently available in the U.S.
for several years for one indication for use. The sponsor of the original PMA submitted a PMA
panel-track supplement seeking to expand the indication for this device, which included a
systematic literature review as well as RWD. The systematic literature review served as the
primary clinical evidence to support safety and effectiveness of the device for the new use. The
RWD comprised information from the sponsor’s patient database linked with Medicare claims,
and were provided as supplemental clinical evidence to support the safety of the device with the
proposed indication. An IDE application was not required for the clinical study using RWD
because the clinical data for the subject device were obtained from extant data that had been
generated as part of routine clinical care. The RWE characterized the safety profile in patients
with the specific condition (i.e., new indication) implanted with the device. A comprehensive
review was conducted to identify validated diagnostic (i.e., ICD-CM) codes associated with
safety outcomes, and used to assess these outcomes among patients receiving the implant and
Contains Nonbinding Recommendations
37
diagnosed with the specific condition compared to patients implanted with the device that did not
have the condition. The sponsor’s patient database and Medicare claims were linked using
probabilistic methods. Important considerations from FDA’s review, supporting the relevance
and reliability of the RWD in this submission, included, in part: 1) the devices were clearly
identified in the sponsor’s patient database, 2) Medicare claims captured the relevant outcomes
of interest in the study population, 3) the linkage methods were considered adequate, and 4) the
data accrual and quality of both RWD sources were considered acceptable. Although the safety
information from the RWE indicated a higher adverse event rate for the subject device in patients
with the specific condition compared to patients without the condition, this was expected for the
respective patient population, and was similar to results from a separate systematic literature
review provided by the sponsor. The RWE allowed for an assessment of device safety in a much
larger patient population than the published randomized control trials in the device space and
allowed for estimation of rare safety events. The RWE was part of the totality of clinical
evidence to support that the probable benefits of the device outweigh the probable risks, to
support approval of the submission.
Example 3: Submissions in response to a 522 order
The manufacturer of a class II device was issued a 522 order to address public health questions
on the safety of the device. The device met the statutory criteria for issuance of a 522 order
because its failure would be reasonably likely to have serious adverse health consequences, and
because it was also expected to have significant use in pediatric populations. To meet the 522
postmarket surveillance study requirement, the manufacturer discussed plans with FDA, and
designed a single-arm, prospective study using non-RWD to assess safety outcomes at 12 months
and using RWD to support the continued assessment of the software for its intended use. The
routinely collected device-generated data from patient use of the software were automatically
uploaded via a mobile application into the manufacturer’s database, and were considered RWD.
The device-generated data and method of collection was consistent with the device’s intended
use. There was also monthly follow-up with participants via emailed survey as part of the study
to gather adverse event information. This non-RWD from the follow-up were then linked to the
routinely collected device-generated data to determine if the safety events were associated with
device use. The outcomes were compared to a historical control. The device-generated data were
considered sufficiently relevant and reliable to fulfill the postmarket requirement. Important
considerations from FDA’s review, supporting the relevance and reliability of the RWD in this
submission, included, in part, 1) outcomes of interest were adequately identified from the device-
generated data, 2) the data from the RWD source were considered representative of the intended
use population, including pediatric patients, and 3) adequate data capture methods were used to
ensure completeness, accuracy, and consistency for the manufacturer’s database. At the end of
the study, the 522 postmarket surveillance study met its endpoints and there were no
unanticipated adverse device events. This study addressed the public health questions from the
522 order, using RWD from the device combined with information from separate study-defined
follow-up information from surveys, to support overall generalizability to the intended use
population, and to provide specific adverse event information beyond what was provided in the
premarket clinical study.
Example 4: Control group
A sponsor submitted a PMA panel-track supplement seeking approval for an indication
Contains Nonbinding Recommendations
38
expansion, which relied on a comparison of participants from a single-arm traditional clinical
study to an external control group from a U.S. registry. The devices from the control group were
identified through the use of UDIs. The registry did not routinely collect one key covariate for
the control group; therefore, the participating sites separately gathered and added this
information to the registry from extant EHRs as this covariate was routinely collected and
documented in the EHR. Both the single-arm study and the registry used a common data capture
form with the same definitions for key variables (covariates and outcomes of interest). The
sponsor proposed a proxy outcome from the registry, which was also a secondary endpoint in the
traditional clinical study, because the primary endpoint typically used for traditional clinical
studies in the device space could not be assessed in the registry. This proxy outcome from the
registry was considered clinically acceptable for this particular study by the sponsor and FDA.
To account for potential differences between the device and control groups, the sponsor
performed a propensity-score adjusted analysis using pre-specified variables. The propensity
score results were reviewed and considered acceptable by FDA before the sponsor performed the
outcome analysis. Important considerations from FDA’s review, supporting the relevance and
reliability of the RWD in this submission, included, in part: 1) use of the control device was
definitively captured through UDIs in the registry, 2) the registry captured the outcomes and
covariates of interest in the relevant patient population, and 3) acceptable data accrual methods
were used, such as use of a common data capture form in the registry and adequate data
abstraction from extant EHRs. Although there were missing data, sensitivity analyses
demonstrated that the extent of missing data did not change the conclusions of the study. This
RWE was part of the totality of clinical evidence to support that the probable benefits of the
device outweigh the probable risks, and served as the primary basis for supporting approval of
the PMA supplement.
Example 5: Granting of De Novo
A sponsor submitted data from two clinical studies to support a De Novo request for a device.
One was a prospective, repeated measures, single-arm study, which served as the primary
clinical evidence to support safety and effectiveness for the De Novo request. In addition, the
sponsor provided a clinical study using RWD from EHRs from over 500 patients, as
supplemental clinical evidence. The overall objective of this study using RWD was to confirm
the performance of the subject device, and to further investigate trends in treatment outcomes for
patients with different severities of the disease. The sponsor used this study to provide additional,
powered, single-measure analyses evaluating the primary effectiveness endpoint in the cohort
from the single-arm study, as well as in additional subgroup cohorts not studied in the
prospective study. This evidence helped address uncertainty that remained related to the
prospective, single-arm study. Important considerations from FDA’s review, supporting the
relevance and reliability of the RWD in this submission, included, in part: 1) the appropriate
length of follow-up for the endpoints of interest was available in the RWD source and 2) the data
abstraction methods included prespecified, standardized definitions for key variables (including
covariates, device and outcome measures) and were considered acceptable. The RWE generated
from these data demonstrated consistent outcome measures with the prospective single arm
study. Overall, the clinical evidence from the clinical studies and RWE supported granting of the
De Novo request for this device.
Contains Nonbinding Recommendations
39
Example 6: RWE obtained from use of EUA device
In response to the COVID-19 outbreak, FDA authorized the emergency use of certain devices
under section 564 of the FD&C Act. A sponsor of a test initially authorized for emergency use
utilized RWD from a clinically annotated biobank, where samples were routinely collected, to
support traditional premarket clearance and the expansion of the indications for use under a Dual
510(k) and CW. The RWD from the clinically annotated biobank were provided in addition to
data from a prospective clinical study. Important considerations from FDA’s review, supporting
the relevance and reliability of the RWD in this submission, included, in part, 1) appropriate
outcome data elements were captured and consistently recorded using established systems
including LOINC or Systematized Nomenclature of Medicine Clinical Terms (SNOMED CT)
for completeness and consistency; 2) the study population derived from the clinically annotated
biobank was adequately representative of the intended use population and the samples from the
biobank were appropriately selected. In addition, the results (e.g., usability and performance)
could be reasonably generalized to support use of the device in this patient population. The
sponsor and FDA also agreed that the candidate testing needed to be conducted in accordance
with the authorized instructions for use. In addition, any samples that were excluded from the
study were appropriately justified. The RWE in this submission allowed for an appropriate
assessment of clinical claims made by the sponsor in a larger patient population than the
prospective clinical study and complemented its findings.
Contains Nonbinding Recommendations
40
Guidance History* Date Description
Level 1 Final
Guidance
December 2025 See Notice of Availability for
more information.** This
guidance supersedes the final
guidance titled “Use of Real-
World Evidence to Support
Regulatory Decision-Making for
Medical Devices” and published
August 2017.
Reissued as Level 1
Draft Guidance
December 2023 See Notice of Availability for
more information.**
*This table was implemented, beginning December 2025, and previous guidance history may not
be captured in totality.
**The Notice of Availability is accessible via the Search for FDA Guidance Documents
webpage.